Commissioning
8
3
Commissioning
3.1
Scope of delivery
Thermostat box OxiTop
Special wrench
Operating manual
3.2
Power supply
®
The OxiTop
Box is supplied with power via the power pack included in the
scope of delivery.
3.3
Initial commissioning
1
Position the instrument with sufficient space for ventilation.
2
Connect the cold device plug to the instrument.
3
Connect the power plug to a power outlet.
4
Switch on the instrument (switch on the rear side)
5
For preparation of the sample bottles and execution of the measure-
ment refer to the operating manual of the OxiTop
3.4
Setting the temperature
The instrument is set to 20 °C ± 1 °C.
For a different incubation temperature change the desired value at the tem-
perature regulator.
1
Lift off the closing cap (interior).
2
Modify the desired value with the aid of the special wrench (scope of
delivery).
3
After a sufficient delay for the temperature adjustment of the sample
liquid in the bottles, check the new temperature value (e.g. 25 °C) and
correct it as necessary.
4
Close the closing cap (interior) again.
®
Box (/115)
®
Box
OxiTop
®
measuring heads.
ba31112e04