SECTION 5—OPERATING
INSTRUCTIONS
NEVER smoke in an area where oxygen is being
administered.
NEVER use near any type of flame or flammable/explosive
substances, vapors or atmosphere.
NOTE: For this procedure, refer to FIGURE 5.1 on page 22.
1. Prior to each use, inspect product for visible damage. DO
NOT use if any damage is found.
NOTE: If for any reason, any label becomes illegible or lost, contact
your dealer/home care equipment provider.
2. Be sure regulator is in the "Off" position. Align "Off" with
the indicating pointer.
3. Position the cylinder so the oxygen cylinder valve outlet is
pointing away from the user and any other person(s).
DO NOT direct flow of oxygen at any person, or
flammable material when adjusting the flow dial.
4. Attach a standard adult single lumen oxygen nasal
cannula, no longer than seven feet to the product's outlet
according to the cannula manufacturer's instructions. DO
NOT use pediatric, low flow nasal cannulas or oxygen
masks with this product.
NOTE: A unit without cannula attached may pulse if set at position
1-5.