Dräger Agila Instructions For Use Manual page 43

Ceiling supply units
Hide thumbs Also See for Agila:
Table of Contents

Advertisement

Functionality test and visual inspection of the whole
supply unit.
Visual inspection and tightness testing of
compressed gas hoses.
Visual inspection of all cables and check of
electrical safety.
For the gas terminal units, the instructions for use
from the respective manufacturer must be
observed.
For Dräger terminal units, exchange of the O-ring
and flat sealing ring wear parts.
General overhaul and replacement of all
compressed gas hoses and cables.
In addition, for the Agila lift version, the following
applies:
Test of the lift system and lubrication of the spindle.
Life time of the bulbs
Indirect ceiling light
Floor light
Instructions for use Agila
Initially 5 years after operating the device for the
first time, then annually.
Service personnel must inspect the device.
Initially 5 years after operating the device for the
first time, then annually.
Service personnel must inspect the device.
Every 5 years by experts.
Every 6 years by experts.
Every 12 years by experts.
Every 5 years by experts.
max. 80000 hours
max. 80000 hours
Maintenance
43

Advertisement

Table of Contents
loading

Table of Contents