Agila CC/Agila CC lift express
Ceiling supply units
WARNING
To properly use this medical device,
read and comply with these instruc-
tions for use.

Advertisement

Table of Contents
loading

Summary of Contents for Dräger Agila CC

  • Page 1 Instructions for use Agila CC/Agila CC lift express Ceiling supply units WARNING To properly use this medical device, read and comply with these instruc- tions for use.
  • Page 2 Letters in illustrations denote elements referred to in the text. Trademarks – DrägerService ® – Agila CC ® are trademarks owned by Dräger. – Incidin ® – Dismozon ® are trademarks of other manufacturers. Instructions for use Agila CC/Agila CC lift express...
  • Page 3 NOTE A NOTE provides additional information intended to avoid inconvenience during operation. Abbreviations and symbols For explanations refer to sections "Abbreviations" and "Symbols" in chapter "Overview". Instructions for use Agila CC/Agila CC lift express...
  • Page 4 Experts Experts are persons who perform repair or complex maintenance work on the product. Experts must have the necessary knowledge and experience with complex maintenance work on the product. Instructions for use Agila CC/Agila CC lift express...
  • Page 5: Table Of Contents

    Fault – Cause – Remedy ....27 Agila CC lift express: Emergency operation of the system ....... 28 Cleaning, disinfection, and sterilization.
  • Page 6 This page intentionally left blank. Instructions for use Agila CC/Agila CC lift express...
  • Page 7: For Your Safety And That Of Your Patients

    The medical device is not approved for use in use before using the devices and connections of areas where combustible or explosive gas the media supply unit. mixtures are likely to occur. Instructions for use Agila CC/Agila CC lift express...
  • Page 8 Before operating the medical device, strictly comply with the instructions for use. Instructions for use Agila CC/Agila CC lift express...
  • Page 9 Observe the EMC information during installation and initial operation. For further information see the EMC information provided on page 35. Mobile high-frequency communication equipment can affect medical electrical equipment. Instructions for use Agila CC/Agila CC lift express...
  • Page 10: Product-Specific Safety Information

    The certifications for terminal units (e.g., gas termi- labels, etc.) to fall down into the operation nal units and power sockets) are market-specific. field and compromise the supply to the patient or the connected devices. Instructions for use Agila CC/Agila CC lift express...
  • Page 11: Application

    Application Application Intended use The Agila CC/Agila CC lift express ceiling supply – Data (such as PC networks). unit is used in rooms serving medical purposes. It is the basis for ergonomic workplace layout in It provides operating rooms, intensive care units, and emer- gency units: –...
  • Page 12 Application Instructions for use Agila CC/Agila CC lift express...
  • Page 13: Overview

    B First support arm C Intermediate bearing D Second support arm E Distance tube F Rotating bearing of the media head/media col- G Media head/media column H Drive unit Movable lift arm J Joint coupling Instructions for use Agila CC/Agila CC lift express...
  • Page 14: Abbreviations

    Observe the individual and total max. xx kg load of the device Warning! Strictly follow these Σ kg max. instructions for use Locking of the lift system of the STOP ceiling supply unit Manufacturer RoHS marking Instructions for use Agila CC/Agila CC lift express...
  • Page 15 A Media head: The rating plate is on the top of the media head. Information label B An information label for maximum load is on the distance tube above the media head/media col- umn. Instructions for use Agila CC/Agila CC lift express...
  • Page 16: Definitions Of The Load Descriptions

    Maximum load on the end of the arm capacity system. Media head/ Maximum load Maximum load that can be put on the Media column media head/media column. Racks/Shelves Payload Maximum load minus the weight of racks/shelves. Instructions for use Agila CC/Agila CC lift express...
  • Page 17: Assembly And Preparation

    Make sure that accessories do not get close to the control panel of the ceiling supply unit. Instructions for use Agila CC/Agila CC lift express...
  • Page 18 Stopping the gas supply CAUTION A Apply slight pressure to release bushing, probe returns to park position. Observe the respective manufacturer's instruc- tions when using the gas terminal units. Taking out the plug connection completely Instructions for use Agila CC/Agila CC lift express...
  • Page 19 1 Press probe (A) until it snaps into place.  The indicator (B) turns green, the anesthetic gas scavenging is functional. Connecting the probe for the AIR motor (optional) 1 Press probe (A) until it snaps into place. Instructions for use Agila CC/Agila CC lift express...
  • Page 20: Operation

    5 seconds have elapsed, the ceiling supply unit can If friction brake of column is not set correctly once more be operated from any control panel. and an anesthesia or ventilation device is con- nected the column may pivot. Instructions for use Agila CC/Agila CC lift express...
  • Page 21: Releasing The Locking Brakes

    Only release the key (B) once the system has brake must be released using the keys. come to a standstill. The arm system is locked The keys are installed in the media head or media again. column. Instructions for use Agila CC/Agila CC lift express...
  • Page 22: Positioning The Ceiling Supply Unit

    Move the supply unit carefully. While positioning, ensure that the supply unit does not collide with other people, supply units or surgi- cal lights, as well as walls and other objects. Instructions for use Agila CC/Agila CC lift express...
  • Page 23 3 Turn the media head/media column (A) to the desired position. 1 Turn the media head/media column (A) at right angles to the second support arm (B). 4 Adjust the media head/media column (A). Instructions for use Agila CC/Agila CC lift express...
  • Page 24 (C). 3 Rotate the first support arm (C) around the ceil- ing bearing (D) until it reaches the desired posi- tion. 4 Turn the second support arm (B) back again. Instructions for use Agila CC/Agila CC lift express...
  • Page 25: Agila Cc Lift Express: Electrically-Operated Height

    The keys for height adjustment are either on the control panel on the media head or media column.  At the top and bottom limit positions, the lifting motor switches off automatically. Instructions for use Agila CC/Agila CC lift express...
  • Page 26 30 seconds. After that, the ceiling supply unit must not be actuated for 5 minutes. NOTE Observing these times prevents damage to the supply unit. Instructions for use Agila CC/Agila CC lift express...
  • Page 27: Troubleshooting

    Insufficient gas supply, hissing Compressed gas is escaping Notify hospital maintenance or sound. from inside the ceiling supply unit. DrägerService to check the gas supply. Connected devices must also be taken into consideration. Instructions for use Agila CC/Agila CC lift express...
  • Page 28: Agila Cc Lift Express: Emergency Operation Of The System

    (hospital maintenance). Motor control faulty. Contact DrägerService. Agila CC lift express: Emergency operation of the system In the event of a power supply failure, the lift system NOTE of the Agila lift express version can also be moved The system moves slowly due to the gear ratio.
  • Page 29: Cleaning, Disinfection, And Sterilization

    The material compatibility of various disinfectants has been tested with Dräger accessories to be Parts must be replaced if there are external reprocessed. signs of wear such as cracks, deformation, discoloration, flaking or anything similar. Instructions for use Agila CC/Agila CC lift express...
  • Page 30: Reprocessing List

    Components which can Recommended Machine Manual Sterilization be reprocessed reprocessing cleaning and Cleaning Disinfecting intervals disinfection Media head/media column Per patient Exterior Exterior Shelf Per patient Exterior Exterior Instructions for use Agila CC/Agila CC lift express...
  • Page 31: Maintenance

    ISO 7396-2: – Leakage test, – test for flow and – pressure drop test. If these tests are not passed, the medical devices must not be put into operation. Instructions for use Agila CC/Agila CC lift express...
  • Page 32 General overhaul and replacement of all com- Every 12 years by experts. pressed gas hoses and cables. In addition, for Agila CC lift express version, the fol- Every 5 years by experts. lowing applies: Test of the lift system and lubrication of the spindle.
  • Page 33: Disposal

    Dräger on the Internet at www.drae- ger.com. Use the Search function with the keyword "WEEE" to find the relevant information. If access to Dräger's website is not possible, contact the local Dräger organization. Instructions for use Agila CC/Agila CC lift express...
  • Page 34: Table Of Toxic Or Hazardous Substances Or Elements

    This period shall not be regarded as a warranty period or as a guaranteed lifetime. Prod- ucts bearing a pollution control logo on the label are recyclable and shall not be disposed of casu- ally. Instructions for use Agila CC/Agila CC lift express...
  • Page 35: Technical Data

    The emissions from this device were tested in the following frequency ranges: Emissions Compliance Radiated emissions Class A, group 1 (150 kHz to 30 MHz) Conducted emissions Class A, group 1 (30 MHz to 1 GHz) Instructions for use Agila CC/Agila CC lift express...
  • Page 36 150 kHz to 80 MHz: 3 V, ISM bands: 6 V (IEC 61000-4-6) Electromagnetic fields in the vicinity of wireless Various frequencies from 385 MHz to 5785 MHz: communication devices 9 V/m to 28 V/m Instructions for use Agila CC/Agila CC lift express...
  • Page 37 To ensure that the functional integrity of this device is achieved, there must be a separation distance of at least 1.0 m (3.3 ft) between this device and wire- less high frequency communication devices. Instructions for use Agila CC/Agila CC lift express...
  • Page 38: Product Characteristics

    ISO 11197 Classification in Class II b Connection data 220 to 240 V accordance with EU Directive 93/42/EEC Annex IX UMDNS code Universal Medical 18-046 Device Nomenclature System – Nomenclature for medical devices Instructions for use Agila CC/Agila CC lift express...
  • Page 39  Length data, see "Maximum load" on page 40 supply unit C Swivel range 330° Length of support arm 1 for double-arm ceiling supply unit D Length of distance tube B Length of arm 2 for double-arm ceiling supply unit Instructions for use Agila CC/Agila CC lift express...
  • Page 40 80 kg 2000 mm 80 kg 80 kg 1) The maximum load for Agila CC/Agila CC lift express pole versions (only 1 central support pole underneath the media head) is 60 kg. Instructions for use Agila CC/Agila CC lift express...
  • Page 41: List Of Accessories

    List of accessories List of accessories WARNING Only use authorized accessories listed in "Accessories for supply units" with the part number G15600. Otherwise, the correct func- tioning of the device may be compromised. Instructions for use Agila CC/Agila CC lift express...
  • Page 42 Directive 93/42/EEC concerning Medical Devices Manufacturer Manufacture site Drägerwerk AG & Co. KGaA Shanghai Draeger Medical Moislinger Allee 53 – 55 Instrument Co., Ltd. D-23542 Lübeck Block 3, No. 229, Hupo Road Germany Shanghai International Medical Zone +49 451 8 82-0 Pudong New District +49 451 8 82-20 80 Post code: 201 321 Shanghai, China...