Figure 1: Block Diagram Of The Sampling Head - Tektronix SD-14 User Manual

Sampling head
Hide thumbs Also See for SD-14:
Table of Contents

Advertisement

Introduction
2
Acquisition
System
Strobe
CH 1
Generator
Acquisition
System
CH 2

Figure 1: Block Diagram of the Sampling Head

The strobe sense signal is a part of the strobe signal returned to the instru
ment. For calibration purposes, the instrument monitors the time duration of
the strobe drive/strobe sense loop and adjusts delay inside the instrument
to maintain correct strobe timing.
Strobe Drive
Strobe Sense
To/from
Instrument
Delay
Adjustment
SD 14 User Manual