Inspecting A Damaged Slotless Contact - Agilent Technologies 85052A Repair Manual

3.5 mm and type-n slotless contact
Table of Contents

Advertisement

Inspecting A
Damaged Slotless
Contact
2-2
Repairing A Slotless Contact
Inspect the contact under
Usually you can make the repair without disassembling the device to
which the precision slotless contact is attached.
Inspect the slotless contact to see if any of the following conditions
exist:
The inner contact has one or more ngers bent inward or crushed,
preventing proper contact to the male pin (see Figure 2-1).
If so, go to step 1 of \Removing A Slotless Contact."
The inner contact has one or more ngers broken (see Figure 2-2).
If so, go to step 2 of \Removing A Slotless Contact."
The inner contact is pushed inside the center conductor and will
not make contact with the mating connector's male pin (see
Figure 2-3).
If so, go to step 3 of \Removing A Slotless Contact."
The end of the center conductor is dented or scraped near where it
touches the inner contact (see Figure 2-4).
If so, the slotless contact may be damaged beyond the capabilities
of this repair kit, and the device must be repaired or replaced by
Agilent Technologies.
Figure 2-1. Finger Bent In or Crushed

x10 magni cation to de ne the problem.
Slotless Contact Repair