Step 2 - Agilent Technologies 85052A Repair Manual

3.5 mm and type-n slotless contact
Table of Contents

Advertisement

Slotless Contact Repair
If one or more of the inner contact's ngers are broken, you must

Step 2

remove that nger before you remove the entire inner contact:
a. Under magni cation, look down inside the inner contact and
locate the broken nger or ngers.
The ngers may have already fallen out. If so, continue with
step 3.
b. If you can see the broken ngers inside the inner contact, turn the
device upside down and gently tap on it (see Figure 2-8).
Using this gentle tapping, try to force the broken ngers to drop
out or at least move forward far enough so that you can remove
them with the tweezers (see Figure 2-9). Do not, under any
circumstances, use anything other then gentle tapping.
c. After you remove the broken inner contact ngers and can
carefully insert the inner contact removal tool, go to step 3.
Figure 2-6. Inserting the Removal Tool
Figure 2-7. Moving a Damaged Finger
Repairing A Slotless Contact
2-5