3
END USER
Sessantaquattro+ multichannel amplifier allows non-invasive recording of biopotentials (sEMG, EEG)
detected by superficial electrodes. The end user must be familiar with the technique and have received
proper training in EMG or EEG detection and interpretation.
3.1
Contraindications
Sessantaquattro+ has no particular contraindications when used jointly with personal computers,
provided that all the electrical devices connected to it comply with the safety rules and standards
concerning grounding and leakage currents.
3.2
Side effects
No significant side effects are known. The materials used for manufacturing all the parts in contact with
the patient are biocompatible. Possible slight cutaneous allergic reactions (e.g., skin reddening) are
reduced to a minimum by reducing the duration of bioelectrical signal acquisitions.
4
SAFETY PRECAUTIONS AND OTHER WARNINGS
The use of the multichannel amplifier Sessantaquattro+ is absolutely forbidden in the following
conditions:
•
While other monitoring devices are in use with the patient
•
While electro surgery equipment, short wave or microwave therapy devices are being used.
•
By mentally impaired people.
•
Whenever the equipment is damaged.
Sessantaquattro+ User Manual v.2.0 – March 2023