Thermo 410i Instruction Manual

Thermo 410i Instruction Manual

Optical filter co2 analyzer
Table of Contents

Advertisement

Quick Links

Model 410i
Instruction Manual
Optical Filter CO
Analyzer
2
Part number 102011-00
8Apr2005

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 410i and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Thermo 410i

  • Page 1 Model 410i Instruction Manual Optical Filter CO Analyzer Part number 102011-00 8Apr2005...
  • Page 2 © 2005 Thermo Electron Corporation. All rights reserved. “Analyze. Detect. Measure. Control” is a trademark of Thermo Electron Corporation. All other product names and logos are the property of their respective owner. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult...
  • Page 3: Table Of Contents

    Set Custom Ranges ............3-16 Averaging Time..............3-17 Calibration Factors Menu ............. 3-18 Background Correction ..........3-19 Span Coefficients............3-21 Calibration Menu ..............3-22 Calibrate CO Background ..........3-23 Calibrate CO Coefficient..........3-24 Zero/Span Check Menu............. 3-24 Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 4 Service Menu ..............3-101 Single/Dual/Auto Select ........... 3-102 Pressure Calibration ............3-103 Flow Calibration .............. 3-106 Initial S/R Ratio............... 3-108 Multi-point Calibration ........... 3-109 Preamp Board Calibration..........3-112 Temperature Calibration..........3-113 Analog Output Calibration ..........3-113 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 5 Cleaning the Outside Case ............5-2 IR Source Replacement ............5-2 Fan Filter Inspection and Cleaning .........5-2 Leak Test and Pump Check Out..........5-3 External Leaks ..............5-3 Leaks Across the Optional Zero/Span and Sample Solenoid Valves...................5-4 Pump Rebuilding..............5-4 Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 6 Ambient Temperature Calibration ........7-31 I/O Expansion Board (Optional) Replacement ..... 7-32 Digital Output Board Replacement........7-34 Motherboard Replacement............ 7-35 Measurement Interface Board Replacement ......7-36 Front Panel Board Replacement..........7-37 LCD Module Replacement ...........7-38 Service Locations..............7-39 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 7 Ethernet Connection............8-11 External Accessory Connector ..........8-11 Chapter 9 Optional Equipment ................9-1 Internal Zero/Span and Sample Valves........9-1 Teflon Particulate Filter ............9-1 I/O Expansion Board Assembly..........9-1 Terminal Block and Cable Kits ..........9-1 Mounting Options..............9-2 Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 8 Format Specifier for Binary Responses ....... B-41 Format Specifier for Front-Panel Layout ......B-41 Appendix C MODBUS Protocol ................C-1 Serial Communication Parameters ..........C-2 TCP Communication Parameters ...........C-2 Application Data Unit Definition ...........C-2 Function Codes...............C-3 MODBUS Commands Supported..........C-8 viii Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 9 • Chapter 9 “Optional Equipment” describes the optional equipment that can be used with this analyzer. • Appendix A “Warranty” is a copy of the warranty statement. Thermo Electron Corporation Model 48i Instruction Manual...
  • Page 10 The hazard or unsafe practice could result in minor to moderate personal injury if the warning is ignored. Equipment Damage The hazard or unsafe practice could result in property damage if the warning is ignored. Model 48i Instruction Manual Thermo Electron Corporation...
  • Page 11 The service procedures in this manual are restricted to qualified service personnel only. The Model 410i is supplied with a three-wire grounding cord. Under no circumstances should this grounding system be defeated. CAUTION If the LCD panel breaks, do not to let the liquid crystal contact your skin or clothes.
  • Page 12 Preface Model 48i Instruction Manual Thermo Electron Corporation...
  • Page 13: Introduction

    The Model 410i Analyzer measures CO concentration using advanced NDIR technology. The Model 410i combines proven detection technology, easy to use menu-driven software, and advanced diagnostics to offer unsurpassed flexibility and reliability. The Model 410i has the following features: • 320 x 240 graphics display •...
  • Page 14: Principle Of Operation

    The Model 410i operates on the principle that carbon dioxide (CO absorbs infrared radiation at a wavelength of 4.26 microns. The sample is drawn into the Model 410i through the sample bulkhead, as shown in Figure 1-1. The sample flows through the optical bench.
  • Page 15: Specifications

    Introduction Specifications Figure 1-1. Model 410i Flow Schematic Specifications Table 1-1. Model 410i Specifications Preset ranges 0-200, 500, 1000, 2000, 5000, and 10000 ppm 0-500, 1000, 2000, 5000, 10000, 20000 mg/m Custom ranges 0-200 to 10000 ppm ppm 0-500 to 20000 mg/m Zero roise 0.5 ppm RMS (60 second averaging time)
  • Page 16 1 RS-232 or RS-485 with two connectors, baud rate 1200–115200, data bits, parity, and stop bits, protocols: C-Link, MODBUS, and streaming data (all user selectable) Ethernet connection RJ45 connector for 10Mbs Ethernet connection, static or dynamic TCP/IP addressing 410i Instruction Manual Thermo Electron Corporation...
  • Page 17: Installation

    Equipment Damage Do not attempt to lift the instrument by the cover or other external fittings. Unpacking and The Model 410i is shipped complete in one container. If there is obvious Inspection damage to the shipping container when you receive the instrument, notify the carrier immediately and hold for inspection.
  • Page 18 Units without Optional I/O Board Figure 2-1. Remove the Packing Material 4. Remove the three shipping screws (Figure 2-2). Shipping Screws (3) Figure 2-2. Remove the Shipping Screws 5. Check for possible damage during shipment. Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 19: Setup Procedure

    6. Connect a source of CO -free air or nitrogen to the PURGE port. WARNING The Model 410i is supplied with a three-wire grounding cord. Under no circumstances should this grounding system be defeated. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 20 Installation Setup Procedure Figure 2-3. Model 410i Rear Panel Vent to Exhaust Line at Atmospheric Pressure Input Gas Line Instrument Bulkhead Figure 2-4. Atmospheric Dump Bypass Plumbing Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 21: Startup

    4. Set instrument parameters such as operating ranges and averaging times to appropriate settings. For more information about instrument parameters, see the “Operation” chapter. 5. Before beginning the actual monitoring, perform a multipoint calibration as described in the “Calibration” chapter. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 22 Installation Startup Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 23: Operation

    • “Service Menu” page 3-101 describes service related menu items. • “Password” page 3-120 describes how to enter/change a password, lock and unlock the analyzer. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 24: Display

    Figure 3-1. 410i Front Panel Display CAUTION If the LCD panel breaks, do not let the liquid crystal contact your skin or clothes. If the liquid crystal contacts your skin or clothes, wash it off immediately using soap and water.
  • Page 25: Pushbuttons

    ) move the cursor up, down, left, and right or = Left, Right change values and states in specific screens. = Enter is used to select a menu item, accept/set/save a change, and/or toggle on/off functions. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 26: Soft Keys

    DIAGS ALARM Software Overview The Model 410i is based on menu-driven software as illustrated by the flowchart in Figure 3-3. The Power-Up screen, shown at the top of the flowchart, is displayed each time the instrument is turned on. This screen is displayed while the instrument is warming up and performing self-checks.
  • Page 28: Power-Up Screen

    Operation Software Overview Power-Up Screen The Power-Up screen is displayed on power up of the Model 410i. The Self-Test is displayed while internal components are warming up and diagnostic checks are performed. Run Screen The Run screen displays the CO concentrations.
  • Page 29: Main Menu

    Use select a submenu. XXLO WCONCENTRATIONXXLO 15.80 PPM SAMPLE 12:34 ALARMx MAIN MENU: >RANGE AVERAGING TIME CALIBRATION FACTORS CALIBRATION INSTRUMENT CONTROLS DIAGNOSTICS ALARMS RANGE DIAGS ALARM SERVICE PASSWORD Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 30: Range Menu

    ALARMx RANGE: >GAS UNITS RANGE SET CUSTOM RANGES RANGE DIAGS ALARM XLOW RANGE CONCENTRATION 15.8 PPM SAMPLE 12:34 ALARMx RANGE: >GAS UNITS LO RANGE HI RANGE 10000 SET CUSTOM RANGES RANGE DIAGS ALARM Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 31: Single Range Mode

    None None Ground 16, 18, 19, 35, 37 Signal Ground Note All channels are user definable. If any customization has been made to the analog output configuration, the default selections may not apply. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 32: Dual Range Mode

    Figure 3-5. See Table 3-3 for channels and pin connections. Dual range mode may be selected from the “Single/Dual/Auto Select” in the “Service Menu,” later in this chapter. Figure 3-5. Pin-Out of Rear Panel Connector in Dual Range Mode 3-10 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 33: Autorange Mode

    For example, the low CO range is set to 0–200 ppm and the high CO range is set to 0–1,000 ppm. Thermo Electron Corporation Model 410i Instruction Manual 3-11...
  • Page 34 Figure 3-7. Refer to Table 3-4 for channels and pin connections. Autorange mode may be selected from the “Single/Dual/Auto Select” in the “Service Menu”, later in this chapter. 3-12 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 35 None Ground 16, 18, 19, 35, 37 Signal Ground Note All channels are user definable. If any customization has been made to the analog output configuration, the default selections may not apply. Thermo Electron Corporation Model 410i Instruction Manual 3-13...
  • Page 36: Gas Units

    XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx GAS UNITS: CURRENTLY: SET TO: MG/M3 ? AND SET RANGES AND CAL DEFAULT SETTINGXXXXXXXXXXX CHANGE VALUE SAVE RANGE DIAGS ALARM 3-14 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 37: Co 2 Range

    Range menu or to return to the Press Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx RANGE: CURRENTLY: 10000 SET TO: 5000 ? CHANGE VALUE SAVE VALUE RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-15...
  • Page 38: Set Custom Ranges

    Range menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx CUSTOM RANGES: >CUSTOM RANGE 1 255.5 CUSTOM RANGE 2 275.0 CUSTOM RANGE 3 525.0 RANGE DIAGS ALARM 3-16 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 39: Averaging Time

    Therefore, the lower the averaging time the faster the front panel display and analog outputs respond to concentration changes. Longer averaging times are typically used to smooth output data. Thermo Electron Corporation Model 410i Instruction Manual 3-17...
  • Page 40: Calibration Factors Menu

    “Calibration Menu” later in this chapter. However, the instrument can also be calibrated manually using the Calibration Factors menu. To manually calibrate the instrument, see “CO Backgrounds” and “CO Span Coefficients” below for more information. 3-18 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 41 The background signal is the combination of electrical noise and scattered light. Before the analyzer sets the CO reading to zero, it stores these value as the CO background correction. Thermo Electron Corporation Model 410i Instruction Manual 3-19...
  • Page 42 Bkg. • to increment or decrement the proposed background value. • to save the new background. Press • to return to the Calibration Factors menu or Press return to the Run screen. 3-20 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 43: Co Span Coefficients

    In the Main Menu, choose Calibration Factors > Hi Coef. • to increment or decrement the coefficient value. • Press to save the new coefficient. • Press to return to the Calibration Factors menu or return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-21...
  • Page 44: Calibration Menu

    • Press to return to the Main Menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx CALIBRATION: >CAL BACKGROUND CAL COEF ZERO/SPAN CHECK RANGE DIAGS ALARM 3-22 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 45: Calibrate Co Background

    Calibration menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx CO2 BACKGROUND: CO2: 15.8 CURRENTLY: 0.0 ? SET CO2 TO ZERO RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-23...
  • Page 46: Calibrate Co Coefficient

    Note Zero and Span Calibration Reset are toggle items that change between yes or no when selected, and present only if auto calibration is installed. • In the Main Menu, choose Calibration > Zero/Span Check. 3-24 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 47 • Press to accept a change. • Press to return to the Zero/Span Check menu or return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-25...
  • Page 48 Zero/Span Check menu or return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx ZERO/SPAN PERIOD: CURRENTLY: 120 HRS SET TO: 200 HRS ? CHANGE VALUE SAVE VALUE RANGE DIAGS ALARM 3-26 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 49 5, 10, 20, 30, 60, 90, 120, 180, 240, and 300 seconds. • In the Main Menu, choose Calibration > Zero/Span Check > Zero/Span Avg Sec. • for a list of choices. • Press to save the averaging time. Thermo Electron Corporation Model 410i Instruction Manual 3-27...
  • Page 50 In the Main Menu, choose Calibration > Zero/Span Check > Zero/Span Ratio. • to increment and decrement the numeric value. • Press to save the ratio value. • Press to return to the Zero/Span Check menu or return to the Run screen. 3-28 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 51: Instrument Controls Menu

    Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx INSTRUMENT CONTROLS: >DATALOGGING SETTINGS COMMUNICATION SETTINGS I/O CONFIGURATION TEMPERATURE COMPENSATION PRESSURE COMPENSATION SCREEN CONTRAST SERVICE MODE RANGE DIAGS ALARM DATE/TIME Thermo Electron Corporation Model 410i Instruction Manual 3-29...
  • Page 52: Datalogging Settings

    In the Main Menu, choose Instrument Controls > Datalogging Settings > Select SREC/LREC. • for a list of choices. • Press to set record format. • Press to return to the Datalogging Settings menu or return to the Run screen. 3-30 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 53 Relative Record Filter The Relative Record Filter screen is used to select the starting record to display the number of records back to view. • to increment or decrement the numeric value. Thermo Electron Corporation Model 410i Instruction Manual 3-31...
  • Page 54 12:34 ALARMx DATE AND TIME: 19 MAR 2005 12:34 DAYS SET CURSOR TO MONTHS ACCEPT AS SHOWN RANGE DIAGS ALARM Record Display The Record Display screen (read only) displays the selected records. 3-32 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 55 Datalogging Settings menu or return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx SREC CHANGE: **WARNING** THIS SELECTION WILL ERASE ALL SAVED DATA FOR THIS RECORD TYPE TO CONTINUE RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-33...
  • Page 56 FIELD 22 NONE FIELD 23 NONE FIELD 24 NONE FIELD 25 NONE FIELD 26 NONE FIELD 27 NONE FIELD 28 NONE FIELD 29 NONE FIELD 30 NONE FIELD 31 NONE FIELD 32 NONE 3-34 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 57 > Select Content > Select Field > Concentrations. • to move the cursor up and down. • Press to select a new choice. • Press to return to the Choose Item Type submenu or return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-35...
  • Page 58 Choose Item Type submenu or return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx OTHER MEASUREMENTS: >NONE <-- INTERNAL TEMP CHAMBER TEMP CHAMBER PRES SAMPLE FLOW INTENSITY RANGE DIAGS ALARM MOTOR SPEED 3-36 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 59 In the Main Menu, choose Instrument Controls > Datalogging Settings > Configure Datalogging. • for a list of choices. • Press to select a choice. • Press to return to the Datalogging Settings menu or return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-37...
  • Page 60 Configure Datalogging menu or return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx SET PERIOD FOR SREC: CURRENTLY: OFF MIN SET TO: 60 MIN ? CHANGE VALUE SAVE RANGE DIAGS ALARM 3-38 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 61 > Configure Datalogging > Data Treatment. • for a list of choices. • Press to set the data treatment. • Press to return to the Configure Datalogging menu or return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-39...
  • Page 62: Communication Settings

    Instrument Controls menu or return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx COMMUNICATION SETTINGS: >BAUD RATE INSTRUMENT ID COMMUNICATION PROTOCOL STREAMING DATA CONFIG RS-232/RS-485 SELECTION TCP/IP SETTINGS RANGE DIAGS ALARM 3-40 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 63 ID number if two or more of the same instrument are connected to one computer. Valid Instrument ID numbers are from 0 to 127. The Model 410i has a default Instrument ID of 68. For more information about the Instrument ID, see Appendix B “C-Link Protocol Commands”...
  • Page 64 Communication Settings menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx COMMUNICATION PROTOCOL: CURRENTLY: CLINK SET TO: STREAMING ? CHANGE VALUE SAVE VALUE RANGE DIAGS ALARM 3-42 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 65 The Streaming Data Interval screen is used to adjust the streaming data interval. The following interval times are available: 1, 2, 5, 10, 20, 30, 60, 90, 120, 180, 240, and 300 seconds. Thermo Electron Corporation Model 410i Instruction Manual 3-43...
  • Page 66 Settings > Streaming Data Config > Item 1-18. • to move the cursor up and down. • Press to select a choice. • Press to return to the Streaming Data Config submenu or to return to the Run screen. 3-44 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 67 Choose Item Signal submenu or return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx CONCENTRATIONS: NONE RANGE STATUS LO CO2 HI CO2 RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-45...
  • Page 68 Settings > Streaming Data Config > Select Item > Analog Inputs. • to move the cursor up and down. • Press to select a new choice. • Press to return to the Choose Item Signal submenu or return to the Run screen. 3-46 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 69 RS-232 or RS-485. • to confirm and save the new selection change. • Press to return to the Communication Settings menu or to return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-47...
  • Page 70 Settings > TCP/IP Settings. • to move the cursor up and down. • Press to select a choice. • Press to return to the Communication Settings menu or to return to the Run screen. 3-48 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 71 RANGE DIAGS ALARM IP Address The IP Address is used to edit the IP address. • In the Main Menu, choose Instrument Controls > Communication Settings > TCP/IP Settings > IP Address. Thermo Electron Corporation Model 410i Instruction Manual 3-49...
  • Page 72 • Press to save the new netmask. • Press to return to the TCP/IP Settings menu or to return to the Run screen. 3-50 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 73 MOVE CURSOR CHANGE VALUE SAVE RANGE DIAGS ALARM Host Name The host name screen is used to edit the host name. When DHCP is enabled, this name is reported to the DHCP server. Thermo Electron Corporation Model 410i Instruction Manual 3-51...
  • Page 74: I/O Configuration

    In the Main Menu, choose Instrument Controls > I/O Configuration. • to move the cursor up and down. • Press to select a choice. • Press to return to the Instrument Controls menu or return to the Run screen. 3-52 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 75 Output Relay Settings > 1-10. • to move the cursor up and down. • Press to select a choice. • Press to return to the I/O Configuration menu or return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-53...
  • Page 76 The Instrument State submenu allows the user to select the instrument state that is tied to the selected relay output. A submenu lists signal types of either alarm and non-alarm to choose from. 3-54 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 77 Output Relay Settings > Select Relay > Instrument State > Alarms. • for a list of choices. • Press to save the new selection for the relay. • Press to return to the Instrument State submenu or return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-55...
  • Page 78 Output Relay Settings > Relay 1-10 > Instrument State > Non-Alarm. • for a list of choices. • Press to save the new selection for the relay. • Press to return to the Instrument State submenu or return to the Run screen. 3-56 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 79 Digital Input Settings > 1-16. • to move the cursor up and down. • Press to select a choice. • Press to return to the I/O Configuration menu or return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-57...
  • Page 80 • Press to toggle and set the logic state open or closed. • Press to return to the Digital Input Settings menu or return to the Run screen. 3-58 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 81 CHOOSE ACTION: >NONE <-- CO2 MEASURE MODE ZERO GAS SPAN GAS INITIATE ZERO CHECK INITIATE SPAN CHECK SET BACKGROUND RANGE DIAGS ALARM AUTOCAL TO SPAN SET ANALOG OUT ZERO SET ANALOG OUT FS Thermo Electron Corporation Model 410i Instruction Manual 3-59...
  • Page 82 VOLTAGE CHANNEL 3 VOLTAGE CHANNEL 4 VOLTAGE CHANNEL 5 RANGE DIAGS ALARM VOLTAGE CHANNEL 6 CURRENT CHANNEL 1 CURRENT CHANNEL 2 CURRENT CHANNEL 3 CURRENT CHANNEL 4 CURRENT CHANNEL 5 CURRENT CHANNEL 6 3-60 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 83 Analog Output Configuration menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx SELECT OUTPUT RANGE: SELECTED OUTPUT: V ALL CURRENTLY: 0-10V SET TO: 0-5V CHANGE VALUE SAVE RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-61...
  • Page 84 Range Setting Range Status (NOx) Recommend not to change the setting for this output S/R Ratio LO S/R Ratio HI S/R Ratio Internal Temp User-set alarm min value User-set alarm max value 3-62 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 85 Analog Output Config menu or return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx CHOOSE SIGNAL TYPE: CONCENTRATIONS OTHER MEASUREMENTS ANALOG INPUT RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-63...
  • Page 86 • Press to return to the I/O Configuration menu or return to the Run screen. Note The current outputs are only displayed if the I/O expansion board option is installed. 3-64 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 87 3 characters in length, and defaults to IN1to IN8 (user input channel number). • In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Input Config > Select Channel > Descriptor. Thermo Electron Corporation Model 410i Instruction Manual 3-65...
  • Page 88 In the Main Menu, choose Instrument Controls > I/O Configuration > Analog Input Config > Select Channel > Units. • Press to save the new value. • Press to return to the Analog Input Config menu or return to the Run screen. 3-66 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 89 The Number of Table Points screen allows the user to select how many points are used in the conversion table. The points range from 2 to 10, with a default of 2. Thermo Electron Corporation Model 410i Instruction Manual 3-67...
  • Page 90 Analog Input Config > Table Point 1-10. • to move the cursor up and down. • Press to select a choice. • Press to return to the Analog Input Config menu or return to the Run screen. 3-68 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 91 Table Point submenu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx TABLE POINT 01 VOLTS: CURRENTLY: 5.00 SET TO: 8.00 ? INC/DEC SAVE VALUE RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-69...
  • Page 92: Temperature Compensation

    Model 410i does not require temperature compensation for EPA equivalency. When temperature compensation is on, the display shows the current internal instrument temperature (measured by a thermistor on the Interface board).
  • Page 93: Pressure Compensation

    This empirical data is used to compensate for any change in reaction chamber pressure. This compensation can be used even though the Model 410i does not require pressure compensation for EPA equivalency. When pressure compensation is on, the display represents the current pressure in the fluorescence chamber.
  • Page 94: Screen Contrast

    Instrument Controls menu or return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx SCREEN CONTRAST: CURRENTLY: 50 % SET TO: 60 % ? CHANGE VALUE SAVE VALUE RANGE DIAGS ALARM 3-72 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 95: Service Mode

    Model 410i. For more information about the service mode, see “Service Menu” later in this chapter. Note The service mode should be turned off when finished, as it prevents remote operation.
  • Page 96: Diagnostics Menu

    In the Main Menu, choose Instrument Controls > Diagnostics. • to move the cursor up and down. • Press to select a choice. • Press to return to the Instrument Controls menu or return to the Run screen. 3-74 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 97: Program Version

    • Press to return to the Diagnostics menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx PROGRAM VERSION: PRODUCT: MODEL 410i VERSION: 01.00.00 RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-75...
  • Page 98: Voltages

    The Motherboard screen (read only) is used to display the current voltage readings on the motherboard. • In the Main Menu, choose Diagnostics > Voltages > Motherboard Voltages. • Press to return to the Voltages menu or to return to the Run screen. 3-76 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 99 I/O Board Voltages The I/O Board screen (read only) is used to display the current voltage readings on the I/O expansion board. This menu is only accessible if the I/O expansion board is installed. Thermo Electron Corporation Model 410i Instruction Manual 3-77...
  • Page 100: Temperatures

    In the Main Menu, choose Diagnostics > Temperatures. • Press to return to the Diagnostics menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx TEMPERATURES: >INTERNAL 34.6 BENCH 49.7 RANGE DIAGS ALARM 3-78 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 101: Pressure

    In the Main Menu, choose Diagnostics > Flow. • Press to return to the Diagnostics menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx FLOW: 1.108 LPM RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-79...
  • Page 102: Sample/Reference Ratio

    The AGC Intensity screen (read only) displays the intensity (in Hertz) of the reference channel Automatic Gain Control (AGC) circuit. The AGC circuit optimizes the noise and resolution levels of the Model 410i. The AGC intensity reading should be about 200,000 Hertz.
  • Page 103: Motor Speed

    100.0 % RANGE DIAGS ALARM Analog Input Readings The Analog Input Readings screen (read only) displays the current user-scaled analog readings. • In the Main Menu, choose Diagnostics > Analog Input Readings. Thermo Electron Corporation Model 410i Instruction Manual 3-81...
  • Page 104: Analog Input Voltages

    ANALOG IN 3 0.00 V ANALOG IN 4 0.00 V ANALOG IN 5 0.00 V ANALOG IN 6 0.00 V ANALOG IN 7 0.00 V RANGE DIAGS ALARM ANALOG IN 8 0.00 V 3-82 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 105: Digital Inputs

    • In the Main Menu, choose Diagnostics > Relay States. • Press to return to the Diagnostics menu or to return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-83...
  • Page 106: Test Analog Outputs

    In the Main Menu, choose Diagnostics > Test Analog Outputs. • to move the cursor up and down. • Press to select output. • Press to return to the Diagnostics menu or to return to the Run screen. 3-84 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 107 Voltage Channel 1-6, or Current Channel 1-6. • to move the cursor up and down. • Press to select a choice. • Press to return to the Diagnostics menu or to return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-85...
  • Page 108: Instrument Configuration

    Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx INSTRUMENT CONFIGURATION: >I/O EXPANSION BOARD ZERO/SPAN VALVES ZERO AIR SCRUBBER WHEEL PURGE DILUTION RATIO AUTO CALIBRATION RANGE DIAGS ALARM 3-86 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 109: Contact Information

    In the Main Menu, choose Alarms. • to move the cursor up and down. • Press to select a choice. • Press to return to the Main Menu or to return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-87...
  • Page 110: Internal Temperature

    In the Main Menu, choose Alarms > Internal Temp. • to move the cursor up and down. • Press to select a choice. • Press to return to the Alarms menu or to return to the Run screen. 3-88 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 111 Internal Temperatures menu or return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx INTERNAL TEMPERATURE: ACTUAL MIN: 30.1 SET MIN TO: 10.0 INC/DEC SAVE VALUE RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-89...
  • Page 112: Bench Temperature

    • to increment or decrement the numeric value. • Press to save set value as actual vlaue. • Press to return to the Bench Temperatures menu or return to the Run screen. 3-90 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 113: Pressure

    Alarms menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx PRESSURE: ACTUAL 746.9 mmHg >MIN 250.0 mmHg 1000.0 mmHg RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-91...
  • Page 114: Sample Flow

    In the Main Menu, choose Alarms > Sample Flow. • to move the cursor up and down. • Press to select a choice. • Press to return to the Alarms menu or to return to the Run screen. 3-92 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 115 Sample Flow menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx SAMPLE FLOW: ACTUAL MIN: 0.350 LPM SET MIN TO: 0.500 LPM ? INC/DEC SAVE VALUE RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-93...
  • Page 116: Bias Voltage

    • Press to save set value as actual vlaue. • Press to return to the Bias Voltage menu or to return to the Run screen. 3-94 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 117: Agc Intensity

    Alarms menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx AGC INTENSITY: ACTUAL 200000 Hz >MIN 150000 Hz 300000 Hz RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-95...
  • Page 118: Motor Speed

    100.0% indicates a problem with either the motor or the power supply. • In the Main Menu, choose Alarms > Motor Speed. • Press to return to the Alarms menu or to return to the Run screen. 3-96 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 119: Zero And Span Check

    Press to return to the Alarms menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx ZERO CHECK: ALARM: RESPONSE: 0.108 >MAX OFFSET: 0.200 RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-97...
  • Page 120: Zero And Span Auto Calibration

    • In the Main Menu, choose Alarms > Zero or Span Autocal. • Press to return to the Alarms menu or to return to the Run screen. 3-98 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 121: Concentration

    In the Main Menu, choose Alarms > Concentration. • to move the cursor up and down. • Press to select a choice. • Press to return to the Alarms menu or to return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-99...
  • Page 122 Concentration menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx CONCENTRATION: ACTUAL MIN: 10000 SET MIN TO: 0100.000 MOVE CURSOR INC/DEC SAVE VALUE RANGE DIAGS ALARM 3-100 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 123: Service Menu

    In the Main Menu, choose Service. • to move the cursor up and down. • Press to select a choice. • Press to return to the Main Menu or to return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-101...
  • Page 124: Single/Dual/Auto Select

    Service menu or to return to the Press Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx SINGLE/DUAL/AUTO RANGE: CURRENTLY: SINGLE SET TO: DUAL ? CHANGE VALUE SAVE VALUE RANGE DIAGS ALARM 3-102 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 125: Pressure Calibration

    Note A vacuum pump must be connected to the pressure sensor before performing the zero calibration. • In the Main Menu, choose Service > Pressure Calibration > Zero. • Press to save the current pressure reading as the zero reading. Thermo Electron Corporation Model 410i Instruction Manual 3-103...
  • Page 126 • to move and change the value. • Press to save set to value as actual value. • Press to return to the Pressure Calibration menu or return to the Run screen. 3-104 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 127 Pressure Calibration menu or return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMX RESTORE DEFAULT CAL: RESTORE ARE YOU SURE YOU WANT TO?x PRESS TO CONFIRM RESTORE RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-105...
  • Page 128: Flow Calibration

    In the Main Menu, choose Service > Flow Calibration > Zero. • Press to save the current flow reading as the zero reading. • Press to return to the Flow Calibration menu or to return to the Run screen. 3-106 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 129 Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMX CALIBRATE FLOW SPAN: CURRENTLY: 0.800 LPM SET TO: 0.750 LPM ? MOVE CURSOR CHANGE VALUE SAVE RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-107...
  • Page 130: Initial S/R Ratio

    • In the Main Menu, choose Service > Initial S/R Ratio. • Use the to move and change the value. • Press to save set to value as actual value. 3-108 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 131: Multi-Point Calibration

    In the Main Menu, choose Service > Multipoint Calibration. • Use the to move the cursor up and down. • Press to select a choice. • Press to return to the Service menu or to return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-109...
  • Page 132 RANGE: 10000 CAL FACTOR 1: 1.0000 CAL GAS 1: 0040.000 MOVE CURSOR CHANGE VALUE SAVE RANGE DIAGS ALARM Coefficients The Coefficients screen allows the user to view and recalculate the calibration coefficients. 3-110 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 133 In the Main Menu, choose Service > Multipoint Cal > Choose Cal Point > Coefficients. • Press to recalculate the coefficients. • Press to return to the Multipoint menu or to return to the Run screen. Thermo Electron Corporation Model 410i Instruction Manual 3-111...
  • Page 134: Preamp Board Calibration

    Service menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx PREAMP BOARD CAL: SAMPLE: 127807 REFERENCE: 148256 SET TO: CHANGE VALUE SAVE VALUE RANGE DIAGS ALARM 3-112 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 135: Temperature Calibration

    The analog output calibration is visible only when the instrument is in service mode. For more information on the service mode, see “Service Mode” earlier in the chapter. Thermo Electron Corporation Model 410i Instruction Manual 3-113...
  • Page 136 VOLTAGE CHANNEL 3 VOLTAGE CHANNEL 4 VOLTAGE CHANNEL 5 VOLTAGE CHANNEL 6 CURRENT CHANNEL 1 RANGE DIAGS ALARM CURRENT CHANNEL 2 CURRENT CHANNEL 3 CURRENT CHANNEL 4 CURRENT CHANNEL 5 CURRENT CHANNEL 6 3-114 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 137 Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMX ANALOG OUTPUT CAL: ZERO CONNECT METER TO OUTPUT! SELECTED OUTPUT: SET TO: SAVE VALUE INC/DEC SET OUTPUT TO: 0.0 V RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-115...
  • Page 138: Analog Input Calibration

    For more information on the service mode, see “Service Mode” earlier in the chapter. Note This screen is present only when the I/O expansion board option is installed. WARNING This adjustment should only be performed by an instrument service technician. 3-116 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 139 ANALOG INPUT CAL: >CALIBRATE ZERO CALIBRATE FULL SCALE RANGE DIAGS ALARM Analog Input Calibration Zero The Analog Input Calibrate Zero screen allows the user to calibrate the zero state of the selected analog input. Thermo Electron Corporation Model 410i Instruction Manual 3-117...
  • Page 140 Use the to increment or decrement the numeric value. • Press to save the value. • Press to return to the Analog Input Calibration screen or to return to the Run screen. 3-118 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 141: Display Pixel Test

    Service menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx DISPLAY PIXEL TEST: DURING TEST PRESS TO EXIT, TOGGLEXXXXXX BEGIN TEST GO BACK TO MENU RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-119...
  • Page 142: Restore User Defaults

    In the Main Menu, choose Password. • to move the cursor up and down. • Press to select a choice. • Press to return to the Main Menu or to return to the Run screen. 3-120 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 143: Lock Instrument

    DIAGS ALARM Change Password The Change Password is used to set or change the password used to unlock the instrument’s front panel. • In the Main Menu, choose Password > Change Password. Thermo Electron Corporation Model 410i Instruction Manual 3-121...
  • Page 144: Remove Password

    Password Menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx REMOVE PASSWORD: PRESSING ENTER WILL REMOVE CURRENT PASSWORD AND DISABLE LOCKING REMOVE PASSWORD RANGE DIAGS ALARM 3-122 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 145: Enter Password

    • Press to return to the Password Menu or to return to the Run screen. XXLO WCONCENTRATIONXXLO 15.8 PPM SAMPLE 12:34 ALARMx ENTER PASSWORD: ENTER PASSWORD RANGE DIAGS ALARM Thermo Electron Corporation Model 410i Instruction Manual 3-123...
  • Page 146 Operation Password 3-124 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 147: Calibration

    So cylinder zero air should be scrubbed of the residual CO prior to its use in the Model 410i as a dilution gas or a zero standard. It is also possible to use a zero air generator in place of zero air cylinders.
  • Page 148: Flow Meter(S) And Controller(S)

    Standard Cylinder Mixing Chamber A chamber constructed of glass, Teflon , or other nonreactive material, and ® designed to provide thorough mixing of CO and diluent air for the dilution method is required. Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 149: Output Manifold

    90 minutes. Perform the service checks of the “Preventive Maintenance” chapter. Select the operating range and the averaging time of the Model 410i. Note The averaging time should be less than the zero duration and less than the span duration.
  • Page 150: Calibration

    Use the following procedure to set the CO reading to zero. 1. Allow sufficient time for the Model 410i to warm up and stabilize. 2. Adjust the dilution system of Figure 4-1 so that zero air alone is present in the manifold.
  • Page 151: Span Adjust

    The first line of the display shows the current CO concentration reading. The second line of the display shows the CO range and the third line is where the CO concentration is entered. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 152: Additional Concentration Standards

    1. Periodically challenging the instrument with zero air. The output of the zero air supply should be greater than the flow demand of the instrument. In addition, an atmospheric dump bypass should be utilized Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 153 Periodic Zero and Span Checks to ensure that the zero air gas flow is being delivered at atmospheric pressure. Record the Model 410i's response in percent of scale as A . Compute the zero drift from the following equation: Zero...
  • Page 154: References

    Cal Point 1 • Cal Point 2 • Cal Point 3 The customer defines the ranges and Thermo Electron recommends the following calibration points for each range: • Cal Point 1: 80% of range • Cal Point 2: 50% of range •...
  • Page 155: Default Coefficients

    2. Allow the Model 410i to sample this CO concentration standard until a stable response is obtained.
  • Page 156 Service Mode menu. 15. Select Coefficients and press The instrument will automatically calculate the new coefficients and save the new parameters. 16. Repeat Steps 1 through 13 for Hi Multi-Point Cal. 4-10 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 157 See the “Servicing” chapter for a list of replacement parts and the associated replacement procedures. WARNING If the equipment is operated in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 158: Preventive Maintenance

    After cleaning the optics, the IR light intensities remain below 100,000 Hz It is not necessary to recalibrate the Model 410i after replacing the IR source since the Model 410i is a ratio instrument, and replacing the IR source does not affect the calibration.
  • Page 159: Leak Test And Pump Check Out

    If the pump diaphragm is in good condition and the capillary not blocked, it should take less than one minute from the time the inlet is plugged to the time the reading below 250 mm Hg is obtained. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 160: Leaks Across The Optional Zero/Span And Sample Solenoid Valves

    1. Turn instrument OFF, unplug the power cord, and remove the cover. 2. Loosen the fittings and remove both lines going to the pump. 3. Remove the four screws from the top plate, remove top plate, flapper valve, and the bottom plate (Figure 5-2). Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 161 Teflon® (white) side of the diaphragm is facing up and that the flapper valves cover the holes of the top and bottom plate. 6. Perform the “Leak Test and Pump Checkout” procedure described in this chapter. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 162 Preventive Maintenance Pump Rebuilding Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 163: Chapter 6 Troubleshooting

    The Technical Support Department at Thermo Electron can also be consulted in the event of problems. See “Service Locations” at the end of this chapter for contact information. In any correspondence with the factory, please note both the serial number and program number of the instrument.
  • Page 164 Recalibrate the D/A with a DVM known to be in calibration. Table 6-2. Troubleshooting - Alarm Messages Alarm Message Possible Cause Action Alarm - Internal Temp Check fan operation Replace fan if not operating properly. Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 165 Check Gain adjustment. Defective measurement interface Replace measurement board interface board. Alarm - Motor Speed Defective measurement interface Replace measurement board interface board. Defective chopper motor or cable Check chopper motor cable. Replace chopper motor. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 166 Internal cables not connected Check that all internal cables Status properly are connected properly. Recycle AC power to Alarm - Interface Board is defective instrument. If still alarming, Status change board. Alarm - I/O Exp Status Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 167 F R O N T P A N E L P J 6 P O W E R S W TRANSFORMER 3 P in (9 0 VAC or 2 4 0 VAC OPTIONS) Figure 6-1. Board-Level Connection Diagram - Common Electronics Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 168 B E N C H H E A T E R A N D T H E R M . 4 P in 3 P in P J 5 A C P U M P 3 P in Figure 6-2. Board-Level Connection Diagram - Measurement System Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 169: Connector Pin Descriptions

    +RS485 to Interface Board -RS485 to Interface Board 10-BASE-T Ethernet Output (+) Ethernet Output (-) Ethernet Input (+) Ethernet Input (-) EXPANSION BD +24V +24V Ground Ground Ground +RS485 to Expansion Board -RS485 to Expansion Board Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 170 Analog Voltage Output 1 Analog Voltage Output 3 Ground Analog Voltage Output 5 Ground Ground Power Fail Relay COM Power Fail Relay N.O. Contact Ground TTL Input 3 TTL Input 4 TTL Input 6 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 171 24V IN +24V Ground DIGITAL I/O +24V +24V Ground Ground Ground SPI Reset SPI Input SPI Output SPI Board Select SPI Clock EXT. RS485 -RS485 to Rear Panel +RS485 to Rear Panel Ground Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 172 LD1 – LCD Signal LD6 – LCD Signal LD2 – LCD Signal LD7 – LCD Signal LD3 – LCD Signal LCD Bias Voltage Ground Ground LCD_ONOFF – LCD Signal Keypad Row 2 Input 6-10 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 173 Serial Port 1 CTS (+RS485 IN) RS232/RS485:B P1:B Serial Port 2 RX (-RS485 IN) Serial Port 2 TX (-RS485 OUT) Ground Serial Port 2 RTS (+RS485 OUT) Serial Port 2 CTS (+RS485 IN) AC IN AC-HOT AC-NEUT Thermo Electron Corporation Model 410i Instruction Manual 6-11...
  • Page 174 SPI – Data Out SPI - CLK SPI - CS +15V +18V for IR Source Ground -100V -100V Return PRES MJ10 Pressure Sensor Input Ground +15V -15V FLOW MJ11 Flow Sensor Input Ground 6-12 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 175 Z/S SOL. +24V Zero/Span Solenoid Control SAMPLE SOL. +24V Sample Solenoid Control SPAN1 SOL. PJ10 +24V Converter Solenoid Control SPAN2 SOL. PJ11 +24V Span 2 Solenoid Control AC O2 AC-HOT AC-NEUT AC-Ground Thermo Electron Corporation Model 410i Instruction Manual 6-13...
  • Page 176 Designator Description MOTHER Ground BOARD Ground LCLK – LCD Signal Ground Ground LLP – LCD Signal LFLM – LCD Signal LD4 – LCD Signal LD0 – LCD Signal LD5 – LCD Signal 6-14 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 177 +24V +24V LCD DATA LFLM_5V – LCD Signal LLP_5V – LCD Signal LCLK_5V – LCD Signal LCD_ONOFF_5V – LCD Signal Ground LCD Bias Voltage LD0_5V – LCD Signal LD1_5V – LCD Signal Thermo Electron Corporation Model 410i Instruction Manual 6-15...
  • Page 178 EXPANSION I/O Analog Voltage Input 1 Analog Voltage Input 2 Analog Voltage Input 3 Ground Analog Voltage Input 4 Analog Voltage Input 5 Analog Voltage Input 6 Ground Analog Voltage Input 7 6-16 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 179 MOTHER BD +24V +24V Ground Ground Ground +RS485 to Motherboard -RS485 to Motherboard Table 6-7. Digital Output Board Connector Pin Descriptions Connector Reference Signal Label Designator Description MOTHER BD +24V +24V Ground Thermo Electron Corporation Model 410i Instruction Manual 6-17...
  • Page 180 Solenoid Drive Output 7 Solenoid Drive Output 8 Relay 1 Contact b Relay 2 Contact b Relay 3 Contact b Relay 4 Contact b Relay 5 Contact b Relay 6 Contact b 6-18 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 181 SPI – Data Out SPI - CLK SPI - CS +15V +18V for IR Source -100V -100V Return +18V for IR Source IR Source Return IR Detector Cooler + IR Detector Cooler - IR Detector Thermo Electron Corporation Model 410i Instruction Manual 6-19...
  • Page 182: Service Locations

    IR Detector Service Locations For additional assistance, Environmental Instruments Division has service available from exclusive distributors worldwide. Contact one of the phone numbers below for product support and technical information. 866-282-0430 508-520-0430 6-20 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 183: Servicing

    Servicing Chapter 7 This chapter explains how to replace the Model 410i subassemblies. It assumes that a subassembly has been identified as defective and needs to be replaced. For fault location information refer to the “Preventive Maintenance” chapter and the “Troubleshooting” chapter in this manual.
  • Page 184 7-32 “Digital Output Board Replacement” page 7-34 “Motherboard Replacement” page 7-35 “Measurement Interface Board Replacement” page 7-36 “Front Panel Board Replacement” page 7-37 “LCD Module Replacement” page 7-38 “Service Locations” page 7-39 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 185: Safety Precautions

    The firmware can be updated by the user in the field via the serial port or over the Ethernet. This includes both the main processor firmware and the firmware in all low-level processors. Refer to the i-Port manual for the firmware update procedure. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 186: Replacement Parts List

    Servicing Replacement Parts List Replacement Parts Table 7-1 lists the replacement parts for the Model 410i major List subassemblies. Refer to Figure 7-2 to identify the component location. Table 7-1. Model 410i Replacement Parts Part Number Description 100480-00 Front Panel Pushbutton Board...
  • Page 187: Cable List

    Part Number Description 101556-00 Terminal Block and Cable Kit (DB37) Cable List Table 7-2 describes the Model 410i cables. See the “Troubleshooting” chapter for associated connection diagrams and board connector pin descriptions. Table 7-2. Model 410i Cables Part Number Description...
  • Page 188: Lowering The Partition Panel

    The partition panel of the measurement bench can be lowered to improve Partition Panel access to connectors and components. Refer to the following steps when a procedure requires lowering the partition panel (see Figure 7-3). Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 189 1. Turn instrument OFF and unplug the power cord. 2. If the instrument is mounted in a rack, remove it from the rack. 3. Remove the cover. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 190: Fuse Replacement

    250VAC, 3 Amp, SlowBlow (for 100VAC and 110VAC models) 250VAC, 1.25 Amp, SlowBlow (for 220-240VAC models) 1. Turn instrument OFF and unplug the power cord. 2. Remove fuse drawer, located on the AC power connector. Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 191: Fan Replacement

    2. Remove the fan guard from the fan and remove the filter. 3. Pull the power connectors off the fan. 4. Remove the four fan mounting screws and remove the fan. 5. Install a new fan following the previous steps in reverse. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 192: Ir Source Replacement

    3. Loosen both clamp screws from the IR source mounting posts and remove IR source. 4. Install a new IR source by following the previous steps in reverse. Ensure that the IR source element is evenly spaced between the mounting posts. 7-10 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 193: Filter Wheel Replacement

    Filter Wheel Set Screw Figure 7-5. Replacing the IR Source Filter Wheel Use the following procedure to replace the filter wheel. Replacement Equipment Required: Filter wheel Allen wrench, 5/32-inch and 5/64-inch Philips screwdriver Thermo Electron Corporation Model 410i Instruction Manual 7-11...
  • Page 194 6. After the filter wheel is installed, spin the wheel and observe that it runs true on the motor shaft. 7-12 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 195: Chopper Motor Replacement

    Allen wrenches, 5/32-inch and 5/64-inch Flatblade screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity. A properly grounded antistatic wrist strap must be worn while handling any internal component. Thermo Electron Corporation Model 410i Instruction Manual 7-13...
  • Page 196: Optical Bench Replacement

    2. Disconnect the chopper motor cable from the MOT DRV connector on the measurement interface board, and disconnect the detector cable from the PREAMP cable connector on the measurement interface board. 3. Disconnect the plumbing connections from the optical bench. 7-14 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 197: Optical Switch Replacement

    1. Turn instrument OFF, unplug the power cord, and remove the cover. 2. Remove the optical bench following the “Optical Bench Replacement” procedure in this chapter. Thermo Electron Corporation Model 410i Instruction Manual 7-15...
  • Page 198: Bench Heater Assembly Replacement

    Figure 7-8. Replacing the Optical Switch (Inverted View) Bench Heater Use the following procedure to replace the bench heater assembly Assembly (Figure 7-8). Replacement Equipment Required: Bench heater Heat conductive compound Flatblade screwdriver 7-16 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 199: Detector/Preamplifier Assembly Replacement

    3. Remove the two screws holding the preamplifier cover to the preamplifier assembly and remove the cover. Thermo Electron Corporation Model 410i Instruction Manual 7-17...
  • Page 200: Pump Replacement

    Nut driver Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity. A properly grounded antistatic wrist strap must be worn while handling any internal component. 7-18 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 201: Pressure Transducer Replacement

    Captive Screws (4) Shock Mounts (4) Floor Plate Pump Mounting Screws (2) (Bottom) Figure 7-10. Replacing the Pump Pressure Transducer Use the following procedure to replace the pressure transducer Replacement (Figure 7-11). Thermo Electron Corporation Model 410i Instruction Manual 7-19...
  • Page 202 Figure 7-11. Replacing the Pressure Transducer 5. To install the pressure transducer assembly, follow the previous steps in reverse. 6. Calibrate the pressure transducer. Refer to the “Pressure Transducer Calibration” procedure that follows. 7-20 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 203: Pressure Transducer Calibration

    At the Main Menu, press to scroll to Instrument Controls > press > to scroll to Service Mode > and press The Service Mode screen appears. b. Press to toggle the Service Mode to ON. Thermo Electron Corporation Model 410i Instruction Manual 7-21...
  • Page 204 9. Wait at least 10 seconds for the ambient reading to stabilize, use to enter the known barometric pressure, and press to save the pressure value. 10. Reconnect the instrument tubing to the pressure transducer. 11. Install the cover. 7-22 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 205: Flow Transducer Replacement

    Figure 7-12. Replacing the Flow Transducer 5. Install the new flow transducer following the previous steps in reverse. 6. Calibrate the flow transducer. Refer to the “Flow Transducer Calibration” procedure that follows. Thermo Electron Corporation Model 410i Instruction Manual 7-23...
  • Page 206: Flow Transducer Calibration

    Press to toggle the Service Mode to ON. c. Press > to return to the Main Menu. d. Continue the procedure at Step 2 to access the Flow Sensor Cal screen. 7-24 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 207 The Calibrate Flow Span screen appears. 10. Wait at least 10 seconds for the reading to stabilize, use to enter the flow sensor reading, and press save the value. 11. Install the cover. Thermo Electron Corporation Model 410i Instruction Manual 7-25...
  • Page 208: Capillary Cleaning Or Replacement

    4. Install the capillary by following the previous steps in reverse. Capillary Figure 7-13. Cleaning or Replacing the Capillary Optional Zero/Span Use the following procedure to replace the solenoid. and Sample Solenoid Equipment Required: Valve Replacement Solenoid 7-26 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 209: Analog Output Testing

    Use the following procedure to test the analog outputs. 1. Connect a meter to the channel to be tested. Figure 7-14 shows the analog output pins and Table 7-3 identifies the associated channels. Thermo Electron Corporation Model 410i Instruction Manual 7-27...
  • Page 210 Test Analog Outputs, and press The Test Analog Outputs screen displays. 3. Press to scroll to the channel (Voltage Channel 1-6) corresponding to the rear panel terminal pin where the meter is connected, and press 7-28 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 211: Analog Output Adjustment

    At the Main Menu, press to scroll to Instrument Controls > press > to scroll to Service Mode > and press The Service Mode screen displays. Thermo Electron Corporation Model 410i Instruction Manual 7-29...
  • Page 212 Calibrate Full Scale. The Analog Output Cal: Span screen displays. 8. Use until the meter reads the value shown in the Set Output To: line, then press to save the value. 7-30 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 213: Ambient Temperature Calibration

    If the Service Mode is not displayed on the Main Menu, use the following procedure to display it. a. At the Main Menu, press to scroll to Instrument Controls > press > to scroll to Service Mode > and press Thermo Electron Corporation Model 410i Instruction Manual 7-31...
  • Page 214: I/O Expansion Board (Optional) Replacement

    3. Remove the two standoffs holding the I/O expansion board connector to the rear panel (Figure 7-16). 4. Pop off the board from the mounting studs and remove the board. 5. To install the I/O expansion board, follow previous steps in reverse. 7-32 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 215 Motherboard Support Bracket Mounting Stud Motherboard Digital Output Board I/O Expansion Board (Opt) Figure 7-15. Replacing the I/O Expansion Board (Optional) I/O Expansion Board Motherboard Digital Output Board Figure 7-16. Rear Panel Board Connectors Thermo Electron Corporation Model 410i Instruction Manual 7-33...
  • Page 216: Digital Output Board Replacement

    4. Using the nut driver, remove the two standoffs securing the board to the rear panel (Figure 7-16). 5. Pop off the digital output board from the mounting studs and remove the board. 6. To install the digital output board, follow previous steps in reverse. 7-34 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 217: Motherboard Replacement

    5. Using the nut driver, remove the six standoffs securing the board to the rear panel. 6. Pop off the motherboard from motherboard support bracket, and remove the motherboard. 7. To install the motherboard, follow previous steps in reverse. Thermo Electron Corporation Model 410i Instruction Manual 7-35...
  • Page 218: Measurement Interface Board Replacement

    3. To install the measurement interface board, follow the previous steps in reverse. 4. Re-install the measurement bench. Refer to “Removing the Measurement Bench” in this chapter. Mounting Studs (4) Measurement Interface Board Figure 7-17. Replacing the Measurement Interface Board 7-36 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 219: Front Panel Board Replacement

    LCD Module Retaining Screw LCD Module Screws (2) Front Panel Board Mounting Studs (2) Ribbon Cables (3) Two-Wire Connector Bottom Support Figure 7-18. Replacing the Front Panel Board and the LCD Module Thermo Electron Corporation Model 410i Instruction Manual 7-37...
  • Page 220: Lcd Module Replacement

    2. Remove the two screws from right side of LCD module (viewed front front). Refer to Figure 7-18. 3. Disconnect the ribbon cable and the two-wire connector from the front panel board. 7-38 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 221: Service Locations

    5. Replace the LCD module by following previous steps in reverse. Service Locations For additional assistance, Thermo Electron has service available from exclusive distributors worldwide. Contact one of the phone numbers below for product support and technical information.
  • Page 222 Servicing Service Locations 7-40 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 223 8-5 describes the system boards, assemblies, and connectors. • “I/O Components” page 8-8 describes the input and output communication functions and components. Hardware Model 410i hardware components (Figure 8-1) include: • Optical bench • Band-pass filter • Bench heater board •...
  • Page 224 This assembly is used for maintaining the optical bench at a constant temperature. Chopper Motor The chopper motor spins the filter wheel and chopper disk at a uniform speed. Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 225 The pressure transducer measures the pressure of the sample gas. Capillary The capillary and the pump control the sample gas flow. Pump The pump draws the sample gas through the optical filter bench. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 226 The front panel display, serial and Ethernet data ports, and analog outputs are the means of communicating the results of the above calculations. The front panel display presents the concentrations simultaneously. The display is updated every 1-10 seconds, depending on the averaging time. Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 227 External connectors include: • External Accessory • RS-232/485 Communications (two connectors) • Ethernet Communications • I/O connector with Power Fail Relay, 16 Digital Inputs, and 6 Analog Voltage Outputs. Internal Connectors Internal connectors include: Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 228 The flow transducer output is produced by measuring the pressure difference across a precision orifice. This unit is used for measuring the flow of sample gas in the measurement system. Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 229 Graphics display. It serves Board as central location to tie the three connectors required for the function key panel, the graphics display control lines, and the graphics display backlight Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 230 The optional I/O Expansion board includes six isolated current outputs. These are software configured for any one of the following ranges, while (Optional) maintaining a minimum resolution of 11 bits: • 0-20 mA Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 231 The actual use of these inputs will vary based on analyzer configuration. The digital inputs are TTL level compatible and are pulled up within the analyzer. The active state can be user defined in firmware. Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 232 All responses are terminated with a carriage return (hex 0D) Refer to Table 8-1 for the DB9 connector pin configuration. Table 8-1. RS-232 DB Connector Pin Configurations DB9 Pin Function 8-10 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 233 Any serial port protocols may be accessed over Ethernet in addition to the serial port. External Accessory The external accessory connector is not used in the Model 410i analyzer. Connector This port is used in other models to communicate with smart external devices that may be mounted hundreds of feet from the analyzer using an RS-485 electrical interface.
  • Page 234 System Description I/O Components 8-12 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 235 Teflon Particulate A 5-10 micron pore size, 2-inch diameter Teflon® element is available for the Filter Model 410i. This filter should be installed just prior to the SAMPLE bulkhead. When using a filter, all calibrations and span checks must be performed through the filter.
  • Page 236 EIA rack Mounted in an EIA-style rack, includes mounting slides, and front panel EIA-rack mounting handles. Retrofit rack Mounted in a Thermo non-EIA rack, includes mounting slides, and retrofit front panel rack-mounting handles. Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 237 Optional Equipment Mounting Options Figure 9-1. Rack Mount Option Assembly Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 241 PERSON OR ENTITY OTHER THAN SELLER WITHOUT SELLER'S PRIOR WRITTEN APPROVAL, OR ANY USE OF REPLACEMENT PARTS NOT SUPPLIED BY SELLER, SHALL IMMEDIATELY VOID AND CANCEL ALL WARRANTIES WITH RESPECT TO THE AFFECTED PRODUCTS. Thermo Electron Corporation 410i Instruction Manual...
  • Page 242 EXPRESS OR IMPLIED, ORAL OR WRITTEN, WITH RESPECT TO THE PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. SELLER DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR-FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT. 410i Instruction Manual Thermo Electron Corporation...
  • Page 243 Appendix B This appendix provides a description of the C-Link protocol commands that can be used to remotely control a Model 410i analyzer using a host device such as a PC or a datalogger. C-Link protocol may be used over RS-232, RS-485, or Ethernet.
  • Page 244: I/O Configuration

    (ASCII) character. The command in the example below begins with the ASCII character code 171 decimal, which directs the command to the Model 410i, and is terminated by a carriage return “CR” (ASCII character code 13 decimal). <ASCII 196> T I M E <CR>...
  • Page 245 C-Link Protocol Commands Commands Table B-1 lists the 410i C-Link protocol commands. The interface will respond to the command strings outlined below. Table B-1. C-Link Protocol Commands Command Description Page addr dns Reports/sets dns address B-32 addr gw Reports/sets default gateway address...
  • Page 246 Reports/sets current date B-30 default params Sets parameters to default values B-31 dhcp Reports/sets state of use of DHCP B-33 diag volt iob Reports diagnostic voltage level for I/O expansion board B-16 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 247 Reports coefficients of the curve developed from hi multi-point B-23 calibration high range Reports/selects current CO high range B-27 high ratio Reports sample/reference ratio calculated using the high B-10 avering time high sp conc Reports/sets high span concentration B-24 Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 248 Outputs long records in the format specified in the command B-17 lrec Outputs long records B-17 lrec format Reports/sets output format for long records (ASCII or binary) B-18 lrec layout Reports current layout of lrec data B-19 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 249 Reports current layout of short record data B-19 srec mem size Reports maximum number of short records B-19 srec per Reports/sets short record logging period B-19 stream per Reports/sets current set time interval for streaming data B-21 Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 250 Table B-2. The example below sets the low range averaging time to 120 seconds. Send: set low avg time 8 Receive: set low avg time 8 ok Table B-2. Averaging Times Selection Averaging Time (seconds) 1 seconds Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 251 27.2 °C. The example below shows that temperature compensation is on and that the internal temperature is 27.2 °C. Send: internal temp Receive: internal temp 27.2 deg C motor Thermo Electron Corporation Model 410i Instruction Manual...
  • Page 252 In the example below, the instrument is reporting that the AGC circuit is on, that the instrument is in the span gas mode, and that the CO high concentration alarm is activated. Send: flags Receive: flags 80038000 B-10 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 253 The example below sets the agc intensity alarm maximum value to 20. Send: set alarm agc intensity max 20 Receive: set alarm agc intensity max 20 ok alarm bias voltage min alarm bias voltage max Thermo Electron Corporation Model 410i Instruction Manual B-11...
  • Page 254 The example below reports that the CO concentration minimum is 5.2 ppm. Send: alarm conc min Receive: alarm conc min 5.2 ppm set alarm conc min value set alarm conc max value B-12 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 255 The example below sets the motor speed alarm maximum value to 20 minutes. Send: set alarm motor speed max 20 Receive: set alarm motor speed max 20 ok alarm pressure min Thermo Electron Corporation Model 410i Instruction Manual B-13...
  • Page 256 The example below shows the CO concentration minimum alarm trigger to ceiling, according to Table B-3. Send: alarm trig conc Receive: alarm trig conc 1 set alarm trig conc value B-14 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 257 15, negative 15, positive 5, positive 3.3, positive 18 IR, positive 18 MOT, and VBIAS. Each voltage value is separated by a space. Send: diag volt mib Receive: diag volt mib 24.1 14.9 -14.9 4.9 3.2 17.9 17.9 Thermo Electron Corporation Model 410i Instruction Manual B-15...
  • Page 258 These commands report the list of current selections for long record logging data, short record logging data, streaming data output, or the scratch pad (sp) list. The example below shows the list for streaming data output. B-16 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 259 47.995 smpfl 0.000 pres 0.000 avgt1 10 avgt2 10 co2bkg -0.000 co2coef 1.000 loco2coef 1.000 co2range 10000000.000 loco2range 10000000.000 motor 100.000 lrec srec lrec xxxx yy srec xxxx yy lrec aa:bb oo-pp-qq yy srec aa:bb oo-pp-qq yy Thermo Electron Corporation Model 410i Instruction Manual B-17...
  • Page 260 The example below shows the output format for long records is ASCII with text, according to Table B-4. Send: lrec format Receive: lrec format 01 set erec format format B-18 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 261 2038 recs, 10 blocks lrec per srec per These commands report the long and short records logging period. The example below shows that the short record logging period is 5 minutes. Thermo Electron Corporation Model 410i Instruction Manual B-19...
  • Page 262 Note Issuing these commands will clear all the logging data memory. All the existing records should be retrieved using appropriate commands, if required. Send: set malloc lrec 10 Receive: set malloc lrec 10 ok B-20 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 263 “set copy sp to lrec”, “set copy sp to srec”, or “set copy sp to stream” commands. Send: set sp field 1 34 Receive: set sp field 1 34 ok stream per Thermo Electron Corporation Model 410i Instruction Manual B-21...
  • Page 264 This command will auto-calibrate the CO background. The example below shows a successful auto-calibration of the CO background. Send: set cal co2 bkg Receive: set cal co2 bkg ok set cal co2 coef B-22 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 265 The “high coef 0”, high coef 1”, and high coef 2” commands report the coefficients of the curve developed from the Hi Multi-Point Calibration in dual or auto range mode. The “low coef 0”, low coef 1”, and low coef 2” Thermo Electron Corporation Model 410i Instruction Manual B-23...
  • Page 266 Send: sp conc Receive: sp conc 1000 set sp conc value set high sp conc value set low sp conc value B-24 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 267 It is sent as a type '5' binary c_link response with no checksum. The RLE encoding consists of a 0 followed by an 8-bit count of consecutive 0xFF bytes. The following 'c' code will expand the incoming data. Thermo Electron Corporation Model 410i Instruction Manual B-25...
  • Page 268 This command is meant for backward compatibility on the C series. Screen information is reported using the “iScreen” command above. Send: screen Receive: screen This is an I series Instrument. Screen Information not available B-26 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 269 The example below sets the CO full-scale range to 50 ppm. Send: set range 5 Receive: set range 5 ok Table B-6. Standard Ranges mg/m Selection 1,000 1,000 2,000 2,000 5,000 5,000 10,000 10,000 custom range Thermo Electron Corporation Model 410i Instruction Manual B-27...
  • Page 270 This command sets the zero/span valves to the sample mode. The example below sets the instrument to sample mode, that is, the instrument is reading the sample gas. Send: set sample Receive: set sample ok B-28 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 271 These commands turn the pressure compensation on or off. The example below turns pressure compensation off. Send: set pres comp off Receive: set pres comp off ok temp comp Thermo Electron Corporation Model 410i Instruction Manual B-29...
  • Page 272 This command sets the screen’s level of contrast, according to Table B-7. The example below sets the contrast level to 50%. Send: set contrast 5 Receive: set contrast 5 ok Table B-7. Contrast Levels Level Contrast Level 100% date B-30 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 273 This command reports the current time (24-hour time). The example below reports that the internal time is 2:15:30 pm. Send: time Receive: time 14:15:30 set time hh:mm:ss hh = hours mm = minutes ss = seconds Thermo Electron Corporation Model 410i Instruction Manual B-31...
  • Page 274 This command sets the analyzer’s IP address, where address consists of four numbers ranging from 0-255 inclusive, separated by “.”. Send: set addr ip 192.168.1.200 Receive: set addr ip 192.168.1.200 ok B-32 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 275 This command reports the current state of use of DHCP on or off. DHCP is used to assign an IP address to the analyzer automatically. The example below shows that DHCP is on. Send: dhcp Receive: dhcp on set dhcp onoff Thermo Electron Corporation Model 410i Instruction Manual B-33...
  • Page 276 This command reports the host name string. Send: host name Receive: host name analyzer01 set host name string This command sets the host name string, where string is 1-13 alphanumeric characters. B-34 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 277 Send: set mode local Receive: set mode local ok program no This command reports the analyzer’s model information and program version number, which will be dependant on the current version. Thermo Electron Corporation Model 410i Instruction Manual B-35...
  • Page 278 C-Link Protocol Commands I/O Configuration Send: program no Receive: program no iSeries 410i 01.01.10.003 set layout ack This command disables the stale layout/layout change indicator (‘*’) that is attached to each response if the layout has changed. Send: set layout ack...
  • Page 279 This command assigns digital input channel (1-16) to activate the action indicated by index (1-35), when the input transitions to the designated state (high or low). Use “list din var” command to obtain the list of supported index values and corresponding actions. Thermo Electron Corporation Model 410i Instruction Manual B-37...
  • Page 280 Voltage Output Not Assigned Not Assigned Not Assigned Current Output Low CO Low CO High/Low CO Current Output Low CO High CO Range Status Current Output Not Assigned Not Assigned Not Assigned B-38 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 281 “closed,” if all the relays are set to same state, that is all open or all closed. The example below shows that the status when all the relays logic is set to normally “open”. Send: relay stat Receive: relay stat open Thermo Electron Corporation Model 410i Instruction Manual B-39...
  • Page 282 (e.g., all of the floating point inputs will be displayed with an 'f' output specifier, and all of the integer inputs will be displayed with a 'd' specifier). B-40 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 283: Format Specifier For Ascii Responses

    Each line is composed of three major components: (1) a text field, (2) a value field, and (3) a button. None of these three components is required. The text field contains statically displayed text. Thermo Electron Corporation Model 410i Instruction Manual B-41...
  • Page 284 All alarm information is presumed to be two bits long (low and high). The bitfield extraction is performed on the integer part of the source. Typical alarm information would appear as '@6.4'. B-42 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 285 This is not currently used. Examples Some examples ('\n' is the C syntax for an end-of-line character): Concentrations\n Thermo Electron Corporation Model 410i Instruction Manual B-43...
  • Page 286 -> remote, etc.). The selection list shows that only the first two values, in this case, are to be shown to the user when the button is pressed. The 'T' button indicates full translation, input code to string, and user selection number to output string. B-44 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 287 The background value is taken from the 7th element, and the precision specifier is taken from the 8th. If the asterisk were not present, it would indicate instead that 8 digits after the decimal point should be displayed. Thermo Electron Corporation Model 410i Instruction Manual B-45...
  • Page 288 C-Link Protocol Commands Record Layout Definition B-46 Model 410i Instruction Manual Thermo Electron Corporation...
  • Page 289: Appendix C Modbus Protocol

    This is achieved by using the supported MODBUS commands listed below. For details of the Model 410i MODBUS Protocol specification, see the following topics: •...
  • Page 290 0x0000 to 0xFFFF (Passed back in response) Protocol Identifier 2 Bytes 0x00 (MODBUS protocol) Length 2 Bytes 0x0000 to 0xFFFF (Number of following bytes) Unit Identifier 1 Byte 0x00 to 0xFF (Passed back in response) 410i Instruction Manual Thermo Electron Corporation...
  • Page 291 These requests specify the starting address, i.e. the address of the first output specified, and the number of outputs. The outputs are addressed starting at zero. Therefore, outputs numbered 1–16 are addressed as 0–15. Thermo Electron Corporation 410i Instruction Manual...
  • Page 292 Here is an example of a request and response to read outputs 2–15: Request Field Name (Hex) Function 0x01 Starting Address Hi 0x00 Starting Address Lo 0x02 Quantity of Outputs Hi 0x00 Quantity of Outputs Lo 0x0D 410i Instruction Manual Thermo Electron Corporation...
  • Page 293 Request Function code 1 Byte 0x03 or 0x04 Starting Address 2 Bytes 0x0000 to maximum allowed by instrument Quantity of Registers 2 Bytes 1 to maximum allowed by instrument Thermo Electron Corporation 410i Instruction Manual...
  • Page 294 Register value Lo (13) 0x64 The contents of register 10 are shown as the two byte values of 0x02 0x2B. The contents of registers 11–13 are 0x00 0x00, 0x00 0x64 and 0x00 0x64, respectively. 410i Instruction Manual Thermo Electron Corporation...
  • Page 295 03=Illegal Data, 04=Slave Device Failure Here is an example of a request to write Coil 5 ON: Request Field Name (Hex) Function Output Address Hi Output Address Lo Output Value Hi Output Value Lo Thermo Electron Corporation 410i Instruction Manual...
  • Page 296 Output Address Lo Output Value Hi Output Value Lo MODBUS Commands The following Tables 1–3 lists the MODBUS commands supported for the Supported Model 410i. Table C-1. Read Registers for 410i Register Number Variable 40001&40002 NOT USED 40003&40004 NOT USED 40005&40006...
  • Page 297 MODBUS Protocol MODBUS Commands Supported Table C-1. Read Registers for 410i Register Number Variable 40045&40046 ANALOG IN 7 40047&40048 ANALOG IN 8 40049&40050 40051&40052 LO CO2 40053&40054 HI CO2 Table C-2. Write Coils for 410i Coil Number Action Triggered ZERO MODE...
  • Page 298 MODBUS Protocol MODBUS Commands Supported Table C-3. Read Coils for 410i Coil Number Status INTENSITY ALARM MOTOR SPEED ALARM BIAS VOLTAGE ALARM MB STATUS ALARM INTERFACE BD STATUS ALARM I/O EXP BD STATUS ALARM CONC ALARM C-10 410i Instruction Manual...

Table of Contents