PROBLEM SOLVER CHART

Always follow the instructions below in case a problem occurs.
Using methods other than those described may cause the pump to function incorrectly.
Cause
• The set is not fitted or wrongly fitted in the
pump.
• The pressure sensor area is dirty ( Fig 1,
position 5 & 6)
• The pump has been untouched in hold
mode for 3 minutes or more.
• The pump administered the installed dose
= volume.
• No flow rate is installed.
Flow rate = 0 ml/h
• The door is not correctly closed.
• The pump detected air in the pump set.
• The air sensor area is dirty( Fig 1, position
8).
• The feeding set is not inserted correctly.
• The battery capacity is too low.
The pump is not able to deliver
highly viscous fluids at high flow rates with
the present charge level
of the battery.
• Battery failure.
• The pump is priming the set.
• The pump detected an upstream occlu-
sion between the pump and the feeding
bag.
• The pressure sensor area is dirty ( Fig 1,
position 5 & 6).
• The pump detected a
downstream occlusion between the pump
and the patient.
• The pressure sensor area is dirty
(Fig 1, position 5 & 6).
Correction
• Stop the alarm by pressing "START/STOP"
• Insert the Flocare
Infinity
feeding set into the pump as indicated on
®
the blister packaging (or refer to Fig 5a - 5d) and close the door.
• Restart the pump.
• Clean the sensors, reinsert the feeding set in the pump and restart
the pump.
• Stop the alarm and prolong the hold mode with another 3 minutes
by pressing "START/STOP" key.
• Program the pump and start it by pressing the "START/STOP".
• The programmed dose/volume has been administered
• Press start/stop to switch off the alarm
• Press INFO and then CLR to reset the administered volume (to reset
the flow rate and dose as well, press the CLR button for 3 seconds)
• Make sure the correct flow rate is programmed.
• Make sure the door is properly closed prior to
starting a feeding program.
• Make sure there is no excessive foam in the reservoir and ensure
there is enough feed in the reservoir for the set therapy.
• If the feeding reservoir is empty, replace the reservoir and continue
feeding. If necessary, prime the set (FILL SET).
• Make sure the air sensor is clean.
• Make sure the feeding set is properly inserted in the pump. (Fig 5a -
5d)
• Connect the charger to the pump and mains and charge the pump
for approx. 6 hours. During charging the pump can be used.
• Turn the pump "off", contact your healthcare provider or refer the
pump to service.
• Press the "FILL SET" key another time to stop the pump and bring it
back in the hold mode.
• Stop the alarm by pressing the "START/STOP" key.
• Remove the feeding set out of the pump and check the permeability
by flushing the line.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the feeding set in the pump and restart
the pump.
• Stop the alarm by pressing the "START/STOP" key.
• Remove the feeding set out of the pump and check the permeability
by flushing the line.
• Check the permeability of the feeding tube.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the feeding set in the pump and restart
the pump.