following:
The interface shows the current deformation of the dial indicator on each shelf. If the difference
between the dial gauge data and the standard displacement measuring instrument exceeds the standard
regulations, the dial gauge should be replaced.
3.5 Heating rate measurement
The measurement of heating rate needs to start a new test, such as: Vicat measurement, select 50℃/h
or 120℃/h rate, and then keep the displacement stable to ensure that the test does not end due to the
arrival of deformation, click the "start test" button to start heating control. After 12 minutes, record the
temperature difference at every 6-minute interval, and the error is within ±0.5℃ to be qualified.
9