OTS XB--VIKA10 Operation Manual page 7

Vicat softening point temperature tester

Advertisement

Dongguan Xin Bao Instrument Co., Ltd
Click the "Start Temperature Control" button to start temperature control and input the corresponding
calibration temperature at the calibration temperature, such as 50. When the measuring instrument just
displays 50 ° C, click the "Calibration" button twice successively, and the system will indicate that the
calibration is successful and the current point calibration is completed. Then enter "100" and wait for
the calibration of the next point.