Repeat the above operations until the corresponding temperature point calibration is completed. The
temperature correction points are: 50, 100, 150, 200, 250 ℃; it must be noted that the calibration
temperature cannot exceed the flash point temperature of the medium oil.
Click the "Stop Temperature Control" button to stop the heating control and complete the calibration,
and click the "Back" button to return to the main interface.
Note: The upper limit protection temperature should be set before heating up.
3.4 Deformation measurement
The measurement of the deformation measurement system requires a micrometer. Before
measurement, please make sure that the dial indicator and the standard measurement instrument for
measurement have been correctly installed on the instrument stand. Please refer to the instrument
manual for the installation method. Click the "Deformation Measurement" button to enter the