OTS XB--VIKA10 Operation Manual page 11

Vicat softening point temperature tester

Advertisement

Dongguan Xin Bao Instrument Co., Ltd
diameter of 10mm; the thickness of the molded sample should be (3~6.5)mm; the thickness of the
plate sample should be the original thickness of the plate, But when the thickness exceeds 6.5mm, the
thickness of the sample is reduced to (3~6.5)mm through single-sided machining. If the thickness of
the plate is less than 3mm, directly stack up to three specimens together so that the total thickness is
between 3mm and 6.5mm, and the thickness of the top piece is at least 1.5mm.
The supporting surface and side of the sample should be parallel, and the surface should be flat and
smooth, without bubbles, no sawtooth marks, dents or flashing and other defects. The stacking surface
of the sample should be kept flat and there should be no gaps between the stacks. There are at least 2
samples in each group.
Before measuring the Vicat softening point temperature, please prepare the sample, perform
pretreatment, measure the size of the sample, then install the sample, click the "Vicat determination"
button to enter the interface as shown in the figure below
11
Add: The 3rd Industrial Zone, Chashan Town, Dongguan. Tel: 0769-33365786 www.ots-testequipment.net