2.2.1 Take out the instrument. The host machine should be placed on a dry, ventilated, non-vibrating
horizontal ground to make the instrument stable.
2.2.2 The computer and printer should be placed on the workbench, and the signal cable and power
cord should be connected.
2.2.3 Connect the inlet of the cooling water pipe with the inlet water pipe, and the outlet water pipe
with the sewer.
2.2.4 Pour methyl silicone oil (or transformer oil) into the oil groove and check whether there is oil
leakage in the oil discharge nozzle and other parts.It is strictly prohibited to power on the instrument
before adding heating medium in the medium box.
2.2.5 Add about 10 liters of methyl silicone oil (or transformer oil), and the oil level is about 50 ~
60mm from the upper plane of the tank.
2.2.6 Put the three sample racks into the oil groove and connect the connection of the deformation
sensor. There is only one temperature sensor on the testing machine, which has been installed at the
factory without user installation. Check whether the load rod movement up and down smoothly.
2.2.7 Check whether the connection is correct and connect the mainframe power supply. At this time,
the stirring motor should run and whether it is sensitive and reliable.
III 、 Calibration
3.1. After the thermal deformation and VEKA softening point temperature detection equipment is
powered on, the test system enters the following figure after completing the self-test:
Click on "About..."Button to enter the system information interface, as shown in the figure below