Liquid Leakage Sensor Amplifier
K7L-AT50
Ultra-miniature Sensor Amplifier Reliably
Detects a Wide Variety of Liquids Ranging
from Water to Liquid Chemicals with Low
Conductivity.
• Detects liquids with impedance as high as 50 MΩ using inter-
electrode resistance detection. Detection of IPA and pure water
is possible.
• Four selectable sensing ranges ensure detection suited to the
characteristics of the liquid.
• Incorporates a noise canceller circuit connected to a 3-conduc-
tor cable, ensuring a high level of noise immunity and reliable
operation (patent pending).
• The power supply block and Sensing Band are isolated, allow-
ing the installation of more than one device in the same place.
Ordering Information
Product name
Liquid Leakage Sensor Amplifier
Sensors
Sensing Band
Point Sensor
Mounting Brackets
Sensing Band Stick-
and Stickers
ers
Point Sensor Mount-
ing Brackets
Terminal Blocks
Track-mounted Socket
Note: 1. Compared with the standard model, F03-16PE.
2. One F03-20 Terminal Block is included as an accessory with the K7L-AT50.
3. The minimum order for the F03-25, F03-26PES, or F03-26PEN Sensing Band Stickers is one set (contains 30 Stickers).
4. The minimum order for F03-20 Terminal Blocks, F03-26PTN Sensing Band Stickers, or F03-26PS Point Sensor Mounting Brackets is one
set (contains 10 Terminal Blocks, Stickers, or Mounting Brackets).
■ Available Sensing Band Lengths
1 m
2 m
F03-16PE
OK
OK
F03-15
OK
OK
F03-16PT
OK
OK
F03-16SF(C)
Note: 1. To place an order for 1 m of the F03-16PE for example, specify F03-16PE-1M.
2. If you cannot find the length you need, please order the nearest larger length, then cut it to the required size.
Model
K7L-AT50
F03-16PE
F03-15
F03-16PT
F03-16SF
F03-16SFC
F03-16PS
F03-16PS-F
F03-25
F03-26PES
F03-26PEN
F03-26PTN
F03-26PS
F03-20
P2RF-08-E
P2RF-08
5 m
10 m
15 m
20 m
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
---
Standard model
Greater durability and condensation resistance. (See note 1.)
Greater temperature and chemical resistance. (See note 1.)
Greater flexibility and superior workability. (See note 1.)
Greater flexibility and enables visual confirmation when the inner color ap-
pears.
Easier to wipe off than the band type.
Electrodes have fluororplastic coating to resist chemicals.
Used for F03-15 or F03-16SF(C).
Used for F03-16PE (adhesive tape).
Used for F03-16PE (screws).
Used for F03-16PT (screws).
Used for F03-16PS.
---
---
---
25 m
30 m
40 m
50 m
OK
OK
OK
OK
OK
OK
OK
Liquid Leakage Sensor Amplifier
Characteristics
60 m
70 m
75 m
80 m
OK
OK
OK
OK
K7L-AT50
90 m
100 m
OK
OK
OK
3

Advertisement

loading

Summary of Contents for Omron K7L-AT50

  • Page 1 Note: 1. Compared with the standard model, F03-16PE. 2. One F03-20 Terminal Block is included as an accessory with the K7L-AT50. 3. The minimum order for the F03-25, F03-26PES, or F03-26PEN Sensing Band Stickers is one set (contains 30 Stickers).
  • Page 2 Note: These values are possible on condition that a completely insulated 3-conductor VCT cable with a thickness of 0.75 mm and a dielectric strength of 600 V is used together with a Liquid Sensing Band specified by OMRON. (A 0.2-mm cable can also be used.) Accessories F03-20 Terminal Block (for connecting the connecting cable and Sensing Band) Screwdriver for ADJUST setting.
  • Page 3 K7L signal transmission circuit Signal transmission line Sensing Band Subtraction circuit Line 1 Line 2 Connecting cable Resistance measurement circuit K7L-AT50 Liquid Leakage Sensor Amplifier...
  • Page 4 ■ Connections K7L-AT50 F03-20 Terminal Block Signal OUT+ Open-collector output transmission Level input 1 OUT- 100 mA at 30 VDC max. Connecting cable Sensing Band Level input 2 (0.2 mm min.) 12 to 24 VDC ■ Connection Examples NPN Output...