Sterile Needle Guide UA1336

10
The puncture line pattern is shown in Figure 6 The distance between the guide
channel of the needle guide to the first dot on the scan image puncture line is 12-
15mm depending on which angle is chosen. The distance between the dots is 5 mm.
Figure 6. Illustration showing the puncture line and blind spot (See 4th "WARNING"on
page 12).
Needle guide UA1336 and the needle guide insert palettes are supplied sterile in peel
packs and are for single-use only. Contents are only sterile if the package is intact.
The needle guide, the inserts and both palettes must be discarded after use.
WARNING
Disposable components are packaged sterile and are intended for single-use only.
Do not use if:
Integrity of packaging is violated.
Expiration date has passed.
Package label is missing.
The sterile-packed needle guides must be stored at a temperature range from +15°C
(+57°F) to +25°C (+77°F) and at a storage humidity of 30% to 80%.
WARNING
Sterile-packed components must be stored in a safe environment and kept out of direct
sunlight. Large temperature changes during storage may cause condensation and violate
the integrity of the packaging.
Please refer to Care, Cleaning and Safety for an example of how to open a sterile-
packed product.
WARNING
For contaminated disposals such as transducer covers or needle guides, follow disposal
control policies established for your office, department or hospital.
blind spot
April-2023
8815 User Guide (BB1312-N)