Replacement Parts; Bfr Cuffs; Matching Limb Protection Sleeves (Mlps); Matching Limb Protection Sleeves (Mlps) Disposable - Delfi 9-2200-001BFR Operator's Maintenance Manual

Pts for blood flow restriction system
Table of Contents

Advertisement

MAINTENANCE
Malfunction
Battery does not charge when unit
is plugged in to AC power
Unit does not turn OFF (cannot be
set to STANDBY)

REPLACEMENT PARTS

BFR CUFFS

Description
PTS BFR Easi-Fit 18" Tourniquet Cuff, reusable, (4.5" x 18"- RED)
PTS BFR Easi-Fit 24" Tourniquet Cuff, reusable, (4.5" x 24" - GREEN)
PTS BFR Easi-Fit 34" Tourniquet Cuff, reusable, (4.5" x 34" - BLUE)
PTS BFR Easi-Fit 44" Tourniquet Cuff, reusable, (4.5" x 44" – BROWN)

MATCHING LIMB PROTECTION SLEEVES (MLPS)

Description
MLPS (RED) - for PTS BFR Easi-Fit 18" reusable cuff
MLPS (GREEN) - for PTS BFR Easi-Fit 24" reusable cuff
MLPS (BLUE) - for PTS BFR Easi-Fit 34" reusable cuff
MLPS (BROWN) - for PTS BFR Easi-Fit 44" reusable cuff

MATCHING LIMB PROTECTION SLEEVES (MLPS) DISPOSABLE

Description
MLPS Single-Use Disposable (RED) - for PTS BFR Easi-Fit 18" reusable cuff
77
MAINTENANCE
Possible Causes
Faulty or dead battery pack
Pressure in cuff
Internal hardware failure
Corrective Actions
Replace battery pack. Refer to
Battery Testing and Replacement
under Section 4 Maintenance
Deflate cuff and ensure it deflates
fully (disconnect hose if required) to
clear CUFF NOT DEFLATED alarm
Contact Delfi. Unit can be powered
OFF by unplugging the AC power
and removing the battery pack.
Refer to Battery Testing and
Replacement under Section 4
Maintenance
Unit
Part Number
1/ ea
9-7550-018
1/ ea
9-7550-024
1/ ea
9-7550-034
1/ ea
9-7550-044
Unit
Part Number
10 / box
9-7950-018
10 / box
9-7950-024
10 / box
9-7950-034
5 / box
9-7950-044
Unit
Part Number
25 / box
9-7960-018