OPERATING INSTRUCTIONS
ERROR CODES
Error Code
Problem
E0000 to EE008
Internal
electronics
E0012 to E0014
failures
E0026 to E0027
E0031
Calibration
E0016 to E0019
failures
E0009
Pressure
failures
E0021 to E0022
E0024 to E0025
E0029
Alarm system
failures
58
OPERATING INSTRUCTIONS
Possible Causes and Corrective Actions
•
The unit detected a failure related to the internal electronics
•
Restart the unit by pressing the ON/STANDBY button twice.
•
If problem persists, note error code and the accompanied numeric code (if
any) and contact Delfi
•
The unit detected a failure related to calibration
•
Calibrate the unit. Refer to Calibration under Section 4 Maintenance
•
If problem persists, note error code and the accompanied numeric code (if
any) and contact Delfi
•
The unit detected a pressure failure
•
Pressure failures may be caused by incorrect or drifted pressure
calibration, or by exceeding a safety pressure limit during use (E0024)
•
Calibrate the unit. Refer to Calibration under Section 4 Maintenance
•
If problem persists, note error code and the accompanied numeric code (if
any) and contact Delfi
•
One or more components of the audiovisual alarm system has failed
•
Note error code and the accompanied numeric code (if any) and contact
Delfi
2