Thank you for choosing a NIVELCO instrument.
We are sure that you will be satisfied throughout its use!
1. APPLICATION
The NIVOCONT vibrating rod level switches are suitable for low and high level indication of granules and powders with a
3
min. 0.05 kg/dm
density such as cement, lime, sand, grain, feed, sugar, etc. Dust Ex versions are available for using the
instrument in explosion-proof environment.
2. TECHNICAL DATA
2.1 G
ENERAL SPECIFICATION
ERSION
V
Probe length
Material of wetted parts
Housing material
Process connection
Temperature ranges
Max. pressure (absolute)
(1)
Minimum medium density
Response time
Not vibrating (covered)
(selectable)
Vibrating (free)
Supply voltage (universal)
Power consumption
Electrical connections
Ingress protection
Electrical protection
Ex protection
ATEX
mark
IECEx
plastic housing
Mass
aluminium housing
(1)
Depend on friction and granular size of the medium
2.2 S
PECIAL DATA
ABLE EXTENDED VERSION
C
T
EMPERATURE DATA
R_K-5_ _-5Ex
R_C-5_ _-5Ex
Medium temperature
+60°C
(3)
min.: -30°C ....max.:
Ambient temperature range
+60°C
(3)
min.: -30°C ....max.:
Max. surface temperature of
+85°C
process connection
Max. surface temperature
+85°C
Temperature class
T90°C
(2)
Medium temperature for max.1 hour + 95 °C
(3)
To operate the level switch with the maximum values of the related temperature data
the applied cable should permanently withstand up to +90 °C temperature.
UTPUT
ATA
O
D
Output type
Output rating
Output protection
Voltage drop (switched on)
Residual current (switched off)
T
EMPERATURE DIAGRAM
Ambient temperature (T
2.4 O
RDER CODE
VERSION
CODE
PROCESS.
CONN.
Standard
K
Standard polished
S
1 ½ " BSP
(4)
High temp.
H
1 ½ " NPT
(4)
High temp. polished
T
(4)
only for standard and pipe extended versions
TANDARD
IPE EXTENDED
S
P
207 mm
0.3 ... 3 m
1.4571
Aluminium: Powder paint coated (R-500 series)
Plastic: PBT fibre-glass reinforced, flame-retardant (R-600 series)
R_H; R_R.; R_K: 1½" BSP
R_N; R_L; R_C: 1½" NPT
See table 2.2 and Temperature diagram
25 bar (2.5 MPa)
3
0.05 kg/dm
(max. granular size: 10 mm)
< 1.8 sec or 5 ±1.5 sec
< 2 sec or 5 ±1.5 sec
normal type: 20...255 V AC/DC
Ex type: 20...250 V AC (50/60Hz) or 20...50V DC
≤ 2.5 VA / 2 W
2 pcs. M20x1.5 cable glands with Ex ta IIIC protection type or
for normal types M20x1.5 plastic glands for cable Ø 6 to 12 mm,
2 pcs. plug-in type terminal blocks for 0.25 to 1.5 mm
internal thread for 2x ½" NPT cable protective pipe
Housing: IP67
Class I. (to be grounded!)
II 1/2D Ex ta/tb IIIC T90 °C...T170 °C Da/Db
Ex t IIIC T* Da/Db IP67
1.5 kg
1.5 kg (+ 1.4 kg/m)
1.88 kg
1.88 kg (+ 1.4 kg/m)
TANDARD AND PIPE EXTENDED VERSION
S
RK_-5_ _-5Ex, RS_-5_ _-5Ex
with the exception of the cable extended version
(2)
+70°C
+80°C
+60°C
+70°C
+50°C
+60°C
+60°C
+50°C
+85°C
+95°C
85°C
85°C
+85°C
+95°C
85°C
85°C
T100°C
T90°C
ELAY
R
R_ _- 5_ _-1
R_ _- 5_ _- 5Ex
SPDT (potential free)
250 V AC, 8A, AC 1
T
[°C]
A
60
50
35
20
0
0
-30
50
75
100 110
-30
Figure 2.
) versus medium temperature (T
A
N I V O C O N T R
CODE
HOUSING
Standard
Pipe
Cable
Alu cast
H
R
K
Plastic
N
L
C
ABLE EXTENDED
C
1 ... 20 m
Probe: 1.4571
Cable: PE coated
6 bar (0.6 MPa)
2
wire cross section
-30°C≤Tamb≤+60°C *(see table 2.2)
1.5 kg (+ 0.6 kg/m)
1.88 kg (+ 0.6 kg/m)
IGH TEMP
H
.
RH_-5_ _-5Ex,
RT_-5_ _-5Ex
+95°C
+110°C
+160°C
+60°C
+50°C
+35°C
+95°C
+95°C
+135°C
+95°C
+110°C
+160°C
T100°C
T115°C
T170°C
OLID STATE
S
R_ _- 5_ _-3
SPST (electronic)
50 V, 350 mA peak
Overvoltage, overcurrent and overload
< 2.7 V @ 350 mA
< 10 μA
160 T
[°C]
M
)
M
-
-
*
CODE
PROBE
CODE
LENGTH
5
Standard
Pipe
6
207 mm
02
0.3 ... 3 m
03...30
1 ... 20 m
IECEx BKI 13.0001X ♦ BKI16ATEX0005 ♦ rkh5021a0600h_06 ♦ 1/2
NIVOCONT
NIVELCO Process Control Co.
H-
1043 Budapest, Dugonics u. 11.
Telefon: (36-1) 889-0100
E-mail: sales@nivelco.com
R_H
R_R
R_L
F = 500 N
M = 100 Nm
M = 100 Nm
Figure 1.
Maximum allowed torque and force
2.3 A
CCESSORIES
User's manual
Warranty Card
Declaration of Conformity
2 pcs. 3-pole terminal blocks
1 ½ " sealing, for BSP only
2 pcs. M20x1.5 cable glands
SUPPLY / OUTPUT / Ex
Cable
20-255 V AC/DC / Relay
20-255 V AC/DC / Electronic
20-250 V AC or 20-50 V DC / Relay / Ex
01...20
* The order code of an Ex version should end in „Ex"
Manufacturer:
Fax:
(36-1)
889-0200
www.nivelco.com
R_K
R_C
F = 45 kN
CODE
1
3
5

Advertisement

loading

Summary of Contents for NIVELCO NIVOCONT

  • Page 1 We are sure that you will be satisfied throughout its use! 1. APPLICATION The NIVOCONT vibrating rod level switches are suitable for low and high level indication of granules and powders with a min. 0.05 kg/dm density such as cement, lime, sand, grain, feed, sugar, etc. Dust Ex versions are available for using the instrument in explosion-proof environment.