Download Print this page
NIVELCO NIVOTRACK MT 2 Series User Manual

NIVELCO NIVOTRACK MT 2 Series User Manual

Advertisement

Quick Links

Thank you for choosing a NIVELCO instrument
We are sure that you will be satisfied with it throughout its use
1. APPLICATION
NIVOTRACK MT -2 series float type magnetosrictive level transmitter is a two-wire precision measuring
instrument for clean liquids both in normal and hazardous location. These instruments re quire minimum
maintenance since the float is the only moving part.
2. TECHNICAL DATA
Insertion length L
Range
Zero offset
Process pressure (maximum)
Medium density
Medium temperature
Ambient temperature
Float size / material
Material of wetted parts
Output
Accuracy
Linearity
Repeatability
Maximum load resistance
Maximum output / Fault indication
value
Damping
Supply voltage
Ex protection
Ex power supply
Electric protection
Ingress Protection
Process connection
Electric connection
Enclosure
Mass
*
2.1
ACCESSORIES
User's Manual
Warranty sheet
2 x Pg 16 cable gland,
Sealing ring for connection
with BSP thread only (klingerit
oilit)
2.3 DIMENSIONS
89
2xPg16
2x1/2"NPT
**
4 mA
**
UP
20 mA
**
0.5 m ... 3.7 m according to the Order Codes
maximum: up to the bumping point of the float
minimum: 0,6 x Insertion length
maximum: 0,4 x Insertion length
2,5 MPa (25 bar) for steel floats ; 0,3 MPa (3 bar) (20°C) for plastic floats
min. 0.7 g/cm
with float
52 mm ; min. 0.5 g/cm
3
*
- 40 °C...+130 °C
PP: - 20 °C...+80 °C
Ex : - 40 °C... +70 °C
- 40 °C ... +70 °C
52x59 mm / 1.4571
95 mm / 1.4571
Stainless steel : 1.4571(DIN)
PP or PVDF float: see Order Codes
Normal operation
4 mA bottom... 20 mA top
Reverse operation 20 mA bottom... 4 mA top (on special request)
±0.04% or 1 mm
0.035% or 0.8 mm
0,01% or 0,4 mm
R
= ( U
-10.5 V) / 0.02 A
t
t
1 sec, 5 sec adjustable
10.5 ... 36 V DC
EEx ia IIB T5
U
= 28.4 V DC
I
max
max
III. Protection Class
According to the order codes
2 x Pg 16 and 2 x
/
NPT thread, shielded cable: 0,75 ... 1.5 mm
1
2 "
Powder paint coated aluminium
2.2 kg + guiding tube 0.35 kg/m
See derating diagram under 3.!
2.2 O
RDER CODES
NIVOTRACK
Model
Code
Process connection
Transmit
1" BSP
1.4571(DIN)
T
1
1
/
" BSP1.4571
2
2" BSP
1.4571
1" NPT
1.4571
1
/
" NPT1.4571
1
2
2" NPT
1.4571
1" BSP
sliding sleev
1
1
/
" BSP sleeve
2
2" BSP sleeve
1" NPT sleeve
1
/
" NPT sleeve
1
2
2" NPT sleeve
DN 80
DN 100
DN 125
89
2xPg16
2x1/2"NPT
**
4 mA
**
UP
20 mA
According to the order code
95 mm
3
*
PVDF: - 20 °C...+130 °C
*
76x87 mm / PP or PVDF
U
= power supply voltage
t
24mA
= 99 mA
P
= 0.67 Watt
max
IP 67
2
M
- 2
Insertion
Code
Code
lenrth
A
0 m
0
B
1 m
1
C
2 m
2
D
3 m
3
E
G
H
J
K
L
M
N
P
R
Note:
1. Reverse operation on request (20 ... 4 mA)
S
2. Not all code variations can be selected
3. INSTALLATION
The NIVOTRACK MT -2 should be
located in an area, which allows easy
access for service, calibration and
monitoring.
Accuracy
of
measurement will be affected by
waves on the surface or excessive
vibration. Therefore the instrument
should be located away from the inlet
and outlet as far as possible. The
process fluid should be free of any
suspended solids or material that could stick between the float and the guide rod. If used at
temperatures in exess of 70 °C the maximum ambient temperature allowed will be in
accordance with the derating diagram.
USER'S MANUAL
*
Nivelco Process Control Co.
H - 1043 Budapest, Dugonics u. 11.
Telephone: (36-1) 369-7575 Fax: (36-1) 369-8585
E-mail: sales@nivelco.com http://www.nivelco.com
-
Insertion
Code
Version / Ex
lenrth
0 m
0
Steel ball float
0.1 m
1
Steel cylindrical float
0.2 m
2
Steel ball float
0.3 m
3
Steel cylindrical float
0.4 m
4
PP cylindrical float
0.5 m
5
PVDF cylindrical float
0.6 m
6
PVDF cylindrical float
0.7 m
7
PP cylindrical float
0.8 m
8
PVDF cylindrical float
0.9 m
9
PVDF Cylindr float
°C
70
the
45
Medium temperature
Manufacturer:
Code
1
95 mm
2
52 mm
5
95 mm Ex
6
52 mm Ex
A
76mm , flange PP
B
76mm, flange PP
C
76mm, fl. A38+PVDF
E
76mm , flange PP Ex
F
76mm, flange PP Ex
G
76mm, fl. A38+PVDF Ex
70
130
°C

Advertisement

loading

Summary of Contents for NIVELCO NIVOTRACK MT 2 Series

  • Page 1 Thank you for choosing a NIVELCO instrument We are sure that you will be satisfied with it throughout its use 1. APPLICATION NIVOTRACK MT -2 series float type magnetosrictive level transmitter is a two-wire precision measuring instrument for clean liquids both in normal and hazardous location. These instruments re quire minimum maintenance since the float is the only moving part.
  • Page 2 Local current meter (Jumper JP2). Jumper JP2 allows for the connection of an auxiliary current meter: Nivelco provide a warranty for the period of 2 (two) years. Move Jumper JP2 from pins 1-2 to pins 2-3. Repairs under warranty are performed at the Manufacturer's premises; costs of dismantling, reinstalling or shipping are borne by the Customer.