10. Technical Specifications

.
Measurement limits:
(however, optional values within the ranges are also possible, by worse accuracy).
Overloadability: with a pressure up to 150% of the measuring limit. See the Quality Certification
and the Data Table for cases different from the above.
Output, transmitter:
Maximum serial resistance of two-line transmitters:
Accuracy at the current output:
- nominal value:
- linearity:
- hysteresis, repeatability together:
- temperature dependency:
- supply voltage dependency:
Supply voltage:
Protection:
Mechanical data:
- weight:
Electric shock protection:
- insulation breakdown value:
- electric connection:
Electromagnetic compatibility:
Materials in contact with the medium to be measured:
- seal:
0 –
100
mbar
0 –
160
mbar
0 –
250
mbar
0 –
400
mbar
0 –
600
mbar
0 –
1.00
bar
0 –
1.60
bar
0 –
2.50
bar
0 –
4.00
bar
0 –
6.00
bar
0 –
10.0
bar
0 –
16.0
bar
0 –
25.0
bar (as recommended)
from 4-20 mA, with a two-line system
Rmax = (Ut-12) / 0.02 [Ohm]
±0.25%, referring to the end-value
(at special cases see on the data label)
0.5%
nominally 0.5%, referring to the end-value
±0.5% / 10 °C (between 0 and 60 °C)
0.1% / 10 V
12–28 VDC
IP 65, in accordance with MSZ EN 60529
DT7200 L ~0.3 kg
DT7200 M ~0.35 kg
DT7200 H ~0.35 kg
The cover is groundable.
500 V
IP 65
Hirschmann GDM 2009 type connector (IEC 60529)
The device is compatible with the relevant prescriptions
of the standard MSZ EN IEC 61326-1:2021 in terms of
electromagnetic compatibility and noise emission.
KO36 steel
Teflon (PTFE) or Buna EP (EPDM) or Viton (FKM) or
Isolast
(FFKM)
9503
9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dt7000Dt7100Dt7100hDt7200lDt7200mDt7200h