Datcon DT1310 Series Operating Instructions Manual

Intrinsically safe temperature / resistance transmitter

Advertisement

Quick Links

DT1310 xx xx
Intrinsically Safe Temperature / Resistance Transmitter
Operating Instructions

Advertisement

Table of Contents
loading

Summary of Contents for Datcon DT1310 Series

  • Page 1 DT1310 xx xx Intrinsically Safe Temperature / Resistance Transmitter Operating Instructions...
  • Page 2: Table Of Contents

    DT1310 xx xx Contents 1. About this document .............4 1.1. Function ................... 4 1.2. Target group................4 1.3. Symbolism used............... 4 2. For your safety ...............5 2.1. Authorized personnel ............... 5 2.2. Appropriate use................ 5 2.3. Warning about misuse ............. 5 2.4.
  • Page 3 DT1310 xx xx 7. Setting-up ..............19 7.1. First steps................19 7.2. Connecting wire setting............20 7.3. Input mode setting..............21 7.4. Analogue output setting ............22 7.5. Error current / voltage setting..........22 7.6. Setting averaging ..............23 7.7. Output refresh time setting............. 25 7.8.
  • Page 4: About This Document

    DT1310 xx xx 1. About this document 1.1. Function This operating instructions manual has all the information you need for quick set-up and safe operation of DT1310 xx Please read this manual before you start setup 1.2. Target group This operating instructions manual is directed to trained personnel.
  • Page 5: For Your Safety

    For safety and warranty reasons, any internal work on the instruments must be carried out only by DATCON personnel. 2.2. Appropriate use The DT1310 xx xx is an Intrinsically Safe Temperature / Resistance Transmitter.
  • Page 6: Safety Information For Ex Areas Environmental Instructions

    DT1310 xx xx 2.6. Safety information for Ex areas Environmental instructions Please note the Ex-specific safety information for installation and operation in Ex areas. 2.7. Environmental instructions Protection of the environment is one of our most important duties. Please take note of the instructions written in the following chapters: •...
  • Page 7: Product Description

    DT1310 xx xx 3. Product description 3.1. Delivery configuration Delivered items The scope of delivery encompasses: • DT1310 xx xx • documentation: this operating instructions certification warranty 3.2. Type designation 20220128-V0...
  • Page 8: Principle Of Operation

    DT1310 xx xx 3.3. Principle of operation Area of application The DT1310 xx xx Intrinsically Safe Temperature / Resistance Transmitter provide signal transmission and conversion between Pt100 / Pt500 / Pt1000 sensor or resistance / potentiometer and signal processing unit. The transmitter feature complete 3-way isolation: the input to output and both to the power supply circuits are isolated.
  • Page 9: Adjustment

    DT1310 xx xx Power supply The instrument works from a 19-35 VDC supply voltage. The power consumption is 1.4 W. 3.4. Adjustment After connecting the DT1310 xx xx is ready to work with the factory default parameters are written in Chapter 6.1. For working with the default parameters there is no need any adjustment.
  • Page 10: Indicators, Infrared Communicator And The Configuration Start Button

    DT1310 xx xx 3.5. Indicators, infrared communicator and the configuration start button The following figure shows the indicators, the infrared communicator and the configuration start button on the instruments front: 1. The „power” green indicator has two function: - continuous light indicates that the instrument is in measuring mode - a short flash indicates that a successful communication has granted with the configuration software.
  • Page 11: Storage And Transport

    DT1310 xx xx 3.6. Storage and transport This instrument should be stored and transport in places whose climatic conditions are in accordance with chapter 10.1. Technical specifications, as described under the title: Environmental conditions. The packaging of DT1310 xx xx consist of environment- friendly, recyclable cardboard is used to protect the instrument against the impacts of normal stresses occurring during transportation.
  • Page 12: Mounting

    DT1310 xx xx 4. Mounting 4.1. General instructions The instrument should be installed in a cabinet with sufficient IP protection, where the operating conditions are in accordance with chapter 10.1. Technical specification, as described under the title: Operating conditions. Mounting position The instruments are designed in a housing for mounting on TS-35 rail.
  • Page 13: Mounting

    DT1310 xx xx 4.3. Mounting The following figure shows the mounting procedures (fixing on the rail): Mounting on the rail The mounting doesn’t need any tool. 1. Tilt the instrument according to the figure; put the instrument’s mounting hole onto the upper edge of the rail (figure step 1.).
  • Page 14: Connecting

    DT1310 xx xx 5. Connecting 5.1. Preparing the connection Always observe the following safety instructions: • The connection must be carried out by trained and authorized personnel only! • Connect only in the complete absence of supply voltage • Use only a screwdriver with appropriate head Take note the suitability of the connecting cable Select and prepare (wire cross-section, insulation, etc.).
  • Page 15: Connecting Pt / Resistor / Potentiometer To The Input

    DT1310 xx xx 5.2. Connecting Pt / resistor / potentiometer to the input The following figure shows the wiring plan, connecting the Pt / resistor / potentiometer to the input: Wiring plan, connecting the Pt / resistor / potentiometer to the input (see also “Application example”)
  • Page 16: Connecting The Signal Processing Unit And The Power Supply

    DT1310 xx xx 5.3. Connecting the signal processing unit and the power supply The following figure shows the wiring plan, connecting the signal processing unit and the power supply: Wiring plan, connecting the signal processing unit and the power supply (see also “Application example”) Be careful the polarity of...
  • Page 17: The First Start-Up, Indicators

    DT1310 xx xx 6. The first start-up, indicators 6.1. The first start-up After connecting the DT1310 xx xx is ready to work with the factory default parameters. When you are going to use other parameters you should set-up the instrument according to Chapter 7. Factory default settings: •...
  • Page 18: Indicators

    DT1310 xx xx 6.2. Indicators (1) power (green): mode an communication indicator (2) st. (red): mode and error indicator The two indicator gives information in conjunction as follows: • Measurement mode: power: light, st.: dark. • Error state: power: light, st.: blinking. The number of blinkings gives the error information while the output current is forced into 3.2 mA or 20.8 mA (depends on setting).
  • Page 19: Setting-Up

    DT1310 xx xx 7. Setting-up 7.1. First steps Necessary tools For setting-up you need: • mini USB A (5 pin)-USB A cable for infrared interface • the infrared interface (DT10) • USB driver installer.exe on the CD • DT1010_DT1310 V2012-08-31.exe (or latter) configuration software on the CD •...
  • Page 20: Connecting Wire Setting

    DT1310 xx xx 7.2. Connecting wire setting Function The Pt / resistance / potentiometer can be connected to the input either with 4 / 3 / 2 wire. The 4 wire connection gives the best result: the instrument is able to compensate the connecting cable resistances. The 3 wire mode is a compromise it may give a good result when all the 3 wire have same resistance.
  • Page 21: Input Mode Setting

    DT1310 xx xx 7.3. Input mode setting Function The instrument has four measuring modes: • Pt temperature • potentiometer % • potentiometer scaled • resistance [Factory default: Pt temperature measurement.] Sequence of operations 1. Select the appropriate input mode by clicking the button. The figure shows the factory default setting.
  • Page 22: Analogue Output Setting

    DT1310 xx xx 7.4. Analogue output setting Function The output signal can be current: 0 / 4-20 mA, or voltage 0 / 2-10 V. (should be specified at order) [Factory default: 4-20 mA] Sequence of operations 1. Select the appropriate output signal by clicking the button.
  • Page 23: Setting Averaging

    DT1310 xx xx 7.6. Setting averaging Function The instrument performs cca. 12 measurements in each seconds. (In 3 wire mode only 5 measurements). The measurement result is generated as the average of several measurements. Here you can define the number of measurements that should be used for calculating the averaged numerical value.
  • Page 24 DT1310 xx xx Settling time after stabilizing the input signal (3 wire mode): Averaging number: Settling time (sec): 0.25 13.6 20220128-V0...
  • Page 25: Output Refresh Time Setting

    DT1310 xx xx 7.7. Output refresh time setting Function The output current / voltage refresh time can be set independently from the measuring time. There are 6 different refresh times (sec): 0.1; 0.3; 0.5; 1.0; 1.5; 2.0 Sequence of operations 1.
  • Page 26: Analogue Output Scaling

    DT1310 xx xx 7.8. Analogue output scaling Function There are two output versions of the instrument: DT1310 xx Ix xx: 0 / 4-20 mA current output or DT1310 xx U xx: 0 / 2-10 V voltage output It can be assign any temperature or resistance range or potentiometer scale to the output.
  • Page 27: Two Wire Zeroing

    DT1310 xx xx 7.9. Two wire zeroing Function When the Pt / resistance / potentiometer is connected to the input with two wire, the wire resistance will cause a measurement error. With this function can be compensated this error. (The compensation will work only properly at the same ambient temperature.) [Factory default: wire resistance = 0 Ohm.] Sequence of operations...
  • Page 28: Pt Correction Values

    DT1310 xx xx 7.10. Pt correction values Function The real Pt characteristics may differ from the theoretical R-°C curve. The difference will cause a measurement error on temperature measurement. When you are capable to calibrate the Pt sensor on minimum 3 points you will know the resistance difference of the theoretical and real characteristics.
  • Page 29 DT1310 xx xx Sequence of operations The figure shows the factory default settings. 2. Type-in the 3 temperature value: 3. Type-in the bold resistance difference values: The setting is ready. 20220128-V0...
  • Page 30: Potentiometer Calibrating

    DT1310 xx xx 7.11. Potentiometer calibrating Function Before selecting potentiometer measuring mode it should be calibrated the potentiometer. At calibrating must define the start and the end position of potentiometer’s wiper. So you can fit the measuring range of the instrument to the potentiometer scale.
  • Page 31 DT1310 xx xx 3. Set the potentiometer wiper to the end position, than click on the right button. You will see on the right button the resistance value of the potentiometer in the end position. Note: Any wiper position can be assign to the 0% and 100% so you can realize reverse operation too.
  • Page 32: Potentiometer Scaling

    DT1310 xx xx 7.12. Potentiometer scaling Function When you are going to use the potentiometer with better resolution than 1% you can assign resistance values with 1, 2, 3, 4 decimal resolution to the start and end position of the potentiometer. Sequence of operations The figure shows the factory default settings.
  • Page 33: New Password Setting

    DT1310 xx xx 7.13. New password setting Function Here you can exchange the factory password for a new one. The range is: 0-9999. [Factory default: 1000] Sequence of operations 1. Type-in the new password into the upper field. (0-9999) 2. Type-in one more time the new password into the lower field.
  • Page 34: Resetting Default Settings

    DT1310 xx xx 7.14. Resetting default settings Function In this case all the settings are deleted, and the default settings are restored. Using this function makes sense in that case, when the settings of the instrument have changed so much, that it is easier to start the setting-up process from the default factory setting.
  • Page 35: Analogue Output Limiting

    DT1310 xx xx 7.15. Analogue output limiting Function The analogue output works correspond with NAMUR standard. For example the 4-20 mA type output has 3.8-20.5 mA current range. You can limit this range. For example: let the low output limit is 12 mA and the high output limit is 17 mA. In this case if calculated current is lower than 12 mA or higher than 17 mA, the output current will indicate the error state.
  • Page 36: Fault Rectification

    8.2. Repairing There is no user repairable part inside the instrument. In accordance with Point 2.1.: For safety and warranty reasons, any internal work on the instrument must be carried out by DATCON personnel. 20220128-V0...
  • Page 37: Dismounting

    DT1310 xx xx 9. Dismounting 9.1. Dismounting procedure The following figure shows the dismounting procedures: Dismounting from the rail The dismounting procedure needs a screwdriver for slotted screws. 1. Before dismounting disconnect all wires. 2. Put the screwdriver end into the fixing assembly’s hole (figure step 1.).
  • Page 38: Disposal

    DT1310 xx xx 9.2. Disposal According with the concerning EU directive, the manufacturer undertakes the disposal of the instrument that are manufactured by it and intended to be destroyed. Please deliver it in contamination-free condition to the site of the Manufacturer or to a specialized recycling company. 20220128-V0...
  • Page 39: Appendix

    DT1310 xx xx 10. Appendix 10.1. Technical specification Robbanásbiztonsági adatok Certification: BKI18ATEX0004 X, BKI18ATEX0004 X/1 Marking: II (1) G [Ex ia Ga] IIC II (1) D [Ex ia Da] IIIC -20 °C ≤ Ta ≤ +50 °C Maximum voltage: 250 Vrms (power suply and signal output) Ex safety data Pt / resistor / potentiometer input 8.75 V 3.6 mA...
  • Page 40 DT1310 xx xx Configuration Configuration: DT10 infrared interface via USB port with DT1310.exe software Connector via PC: USB 2.0 A Maximum distance the DT10 interface 30 cm and the instrument optical sensor: Galvanic isolation Test voltage: 2500 VDC (between input-output, input- power supply terminals) 500 VDC (between output-power supply terminals)
  • Page 41: Application Example

    DT1310 xx xx 10.2. Application example 20220128-V0...
  • Page 42: Atex Tanúsítvány

    DT1310 xx xx 10.3. ATEX Tanúsítvány 20220128-V0...
  • Page 43 DT1310 xx xx 20220128-V0...
  • Page 44 DT1310 xx xx 20220128-V0...
  • Page 45 DT1310 xx xx 20220128-V0...
  • Page 46 DT1310 xx xx 20220128-V0...
  • Page 47 DT1310 xx xx 20220128-V0...
  • Page 48 DT1310 xx xx 20220128-V0...
  • Page 49 DT1310 xx xx 20220128-V0...
  • Page 50 DT1310 xx xx 20220128-V0...
  • Page 51 DT1310 xx xx 20220128-V0...

Table of Contents