Cirris Touch1 Ultra Performance Verification Manual

Hide thumbs Also See for Touch1 Ultra:

Advertisement

Quick Links

Touch1 Ultra
Performance Verification Manual
Version 2010.1.4
August 9, 2010
(ISO 9001 Certified)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Touch1 Ultra and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Cirris Touch1 Ultra

  • Page 1 Touch1 Ultra Performance Verification Manual Version 2010.1.4 August 9, 2010 (ISO 9001 Certified)
  • Page 2 Touch1 Ultra Performance Verification Manual Version 2010.1.4 Copyright 2006 by Cirris Systems Corporation All Rights Reserved Cirris Systems Corporation 1991 W. Parkway Boulevard Salt Lake City, Utah 84119...
  • Page 3: Table Of Contents

    Self Test Touch Pad Test Probe Test Low and High Voltage Threshold Tests High Voltage Tests Capacitance and 4-Wire Tests Scanner Tests Tester Checkout Summary ......................14 Appendix ............................. 16 Calibration Certificate ........................18 Touch1 Ultra Checkout Manual – Page 1...
  • Page 4: Introduction

    National Institute of Standards and Technology (NIST). Calibration should be performed on your Touch1 Ultra tester at least once a year. You should also check the calibration whenever you suspect the tester may not be operating properly.
  • Page 5: Set Up

    Depending on your calibration system requirements, you may need to prepare a form to record the results of the calibration. At the end of this manual is a Touch1 Ultra Certificate of Calibration. If you use this form, make a photocopy so you can maintain the master for the next time you perform the calibration tests.
  • Page 6: Tester Checkout Tests

    If during testing you want to cancel all remaining tests and see a summary of the test results so far, press Cancel • To print the results on a printer, press . Some tests have individual summary results Print you can also print. Touch1 Ultra Checkout Manual – Page 4...
  • Page 7: Self Test

    Self Test When this test is selected, the Touch1 Ultra will perform ten passes of a series of major tests. These are the same tests performed automatically on power up. They test the basic functionality of the tester to see if it can perform reasonable measurements within specification.
  • Page 8: Touch Pad Test

    To pass the test, all five buttons must be pressed before the timeout. To check the calibration, press all five buttons making sure a star is drawn where the screen is touched. Touch1 Ultra Checkout Manual – Page 6...
  • Page 9: Probe Test

    If no pop-up window is displayed when the test point(s) are probed, press The test will fail and a Failed Probe Test screen will display. No Response. Touch1 Ultra Checkout Manual – Page 7...
  • Page 10 The LV/HV Measurement Test checks the low voltage and high voltage measurements. To pass the test, the Touch1 Ultra must measure the resistor’s value to +- 1% +-0.1 ohm and the high voltage leakage at 5M ohm +- 10%. The Resistance Adapter included in the Performance Verification Kit is needed to perform this test.
  • Page 11 3. The Touch1 Ultra will perform the tests and display either a Pass or Fail result for each test. Scroll down to view the results for all the tests. HV Voltages (R + V) Test The High Voltage Test checks that all the voltages in the high voltage settings of the Touch1 Ultra are within range.
  • Page 12 4. The Touch1 will apply high voltage during each test so keep your hands clear from all connections during the test. Press to begin the test. START Press START to begin the series of high voltage tests. Touch1 Ultra Checkout Manual – Page 10...
  • Page 13 6. Repeat steps 4 and 5 until all the voltages have been tested and recorded. After 50 volts, the tests will increment by 100 volts until 1500V. The Touch1 Ultra will perform the tests and display an overall summary with either a Pass or Fail result for each test.
  • Page 14: Scanner Tests

    The Touch1 Ultra will perform the tests and display either a Pass or Fail result for each test. Results will be displayed for both the capacitance and fourwire tests. Scanner Test The Scanner Test verifies the analog circuitry and relays inside the scanners are working properly all the way to each test point.
  • Page 15 3. The Touch1 Ultra will perform the test and display either a Pass or Fail result for each portion of the test. A screen will then prompt you to remove the adapter and install it in the next “J” position on the scanner. Two tests are done for each scanner.
  • Page 16: Tester Checkout Summary

    The summary will display FAILED for any failed tests. 2. To save the Test Summary results, you must either print them or copy them to paper by hand. If you have a printer connected to your Touch1 Ultra, you can print the Test Summary by pressing PRINT Touch1 Ultra Checkout Manual –...
  • Page 17 An overall summary is given for each test along with the individual test results. Scroll down to view the individual results for all the tests. Touch1 Ultra Checkout Manual – Page 15...
  • Page 18: Appendix

    In other words, the insturment being used to measure the calibrated instrument must be at least four times as accurate as the calibrated instrument. Touch1 Ultra Checkout Manual – Page 16...
  • Page 19 How does everyone know you had an instrument calibrated? The calibration certificate is the record of who, when, and by what equipment the instrument was calibrated. A Touch1 Ultra Certificate of Calibration, which you can photocopy for your calibration, is provided following this section.
  • Page 20: Calibration Certificate

    Touch1 Ultra PCHK Version 1.3 Temperature: Relative Humidity: Tester Serial Number: Instruments used: Serial Number Cal. Date Due Date Zero Ohm Adapter Resistor Leak Adapter Capacitance/Fourwire Adapter HV Block PCHK Adapter Voltmeter Statement of Traceablility: Certified by: Touch1 Ultra Checkout Manual – Page 18...