LASER MAINTENANCE
Laser Maintenance
If the ventus is operated in a smoky or dirty environment, occasional cleaning of the laser window may be
necessary. To perform this procedure, the laser must be turned off and, using optical cloth dampened with research
grade methanol, the laser window must be gently wiped.
•
Always follow the instructions given in this Operating Manual
•
Never touch the connector on the laser head with anything other than the Umbilical cable provided and
always follow the connection instructions in this Operating Manual
•
Do not open the laser head or PSU; this will immediately invalidate the warranty
•
Do not subject the laser head to mechanical shock; if severe this can cause mis-alignment of the laser cavity
•
Do not allow the output window of the laser to be touched as this may damage the precision optical coatings
used. Avoid very dirty atmospheres where dirt may settle on the window
•
Do not operate or store this laser system in very humid or damp environments
Liability
Laser Quantum accepts no liability for damage to persons or property caused by incorrect or unsafe use of any of its
products; this is the sole responsibility of the user. Proper safety regulations for the use of these products must be
observed at all times.