JFE Advantech MK-750ST-E Instruction Manual page 43

Gas leak viewer
Table of Contents

Advertisement

Symptom
Assumed reason
Unable to display
File corrupted.
saved image.
(E002)
Incorrect calendar
Calendar
information
information inside
shown with saved
the instrument is
file
incorrect.
Unable to charge a
Battery pack is
battery pack
deteriorated
Charging out of
charger's
operational
temperature range
Continuous
The temperature
operation time is
inside the instrument
short even with
is high
fully charged
battery pack
Battery pack is
deteriorated
Countermeasure
Unable to restore corrupted file. Please
delete the corrupted file.
Clock may off sets if you do not use the
instrument for a while. Reconfigure the
clock.
The battery pack is deteriorated if it does
not get charged fully. The battery pack is
expendable accessory. Please purchase
another designated battery pack.
The battery charger's operational
temperature is between 0℃ and 40℃.
Please charge in the above mentioned
temperature range.
Turn the power OFF or bring it over to
an environment in low temperature to
bring the temperature of battery pack
itself. Once the temperature of the
battery pack itself becomes low, its
operation returns normal.
Continuous use time length becomes
shorter as the temperature inside the
instrument becomes higher. This is due
to battery pack's protection circuit
becomes active, and not malfunction of
the battery pack.
The battery pack is deteriorated if its
operation time length is short under
normal temperature range. The battery
pack is expendable accessory. Please
purchase another designated battery
pack.
32
Refer
p.23
p.16