JFE Advantech MK-750ST-E Instruction Manual page 42

Gas leak viewer
Table of Contents

Advertisement

Symptom
Assumed reason
Threshold is too high
for leak judgement.
Ultrasonic sensor is
dirty.
The status bar
Receiving sound is
turned red.
too strong/loud.
Malfunction
Sound pressure
There are several
peak spot is
sound sources.
shifted aside.
Abnormal flow
The unit configured
amount.
for flow amount is
incorrect.
Incorrect distance to
the leak source.
Incorrect
measurement
conditions.
Unable to save
Not enough memory
image.
capacity.
Internal memory
error.
(E003) and (E004)
Unable to display
Data not saved.
saved image.
(E001)
Countermeasure
Decrease threshold coefficient (K) for
leak judgement.
The instrument's sensitivity decreases if
there is any debris attached to ultrasonic
sensor. Use air-duster or soft cloth and
clean it off.
Decrease the Gain in case the peak
value is too high.
Your instrument may be malfunctioning
in case it shows the peak value as "0".
Combined sound source peak spot may
be displayed in case there are several
sound sources. Try to measure from
different angle.
Change the unit.
Reconfigure the distance to the leak
source to a correct value.
Flow amount changes depending on the
leak spot area size, shape, pressure,
temperature, humidity, and etc. Only
use the leak amount only as your
reference.
Delete unnecessary files.
Reboot the instrument. If rebooting
does not solve the problem, request our
representative for inspection and repair
if necessary.
The instrument cannot display image
unless there are photo image file, sound
pressure map superposition image file,
and sound pressure map data file. .
31
Refer
p.20
p.36
p.18
p.20
p.14
p.19
p.23
p.23