Model 675 Series Submersible Level Transmitter
2.5.3 Lightning Protection Installation Concept
1. Lightning protection devices should be placed as close to the instrument
2. If the distance between the meter and transmitter, or the meter and recorder
3. This or any installation cannot protect against a direct lightning strike, or
4. Proper installation including grounding require detailed installation
Figure 2-6
Lightning Protection Concept
2.6

Cable Termination

Instructions to installing the Ametek Submersible Transmitter using the
supplied desiccant canister:
Remove Power Source During Installation.
This Ametek submersible transmitter is equipped with a sealed termination
area and moisture absorbing desiccant canister. The purpose is to restrict
moisture from entering the transmitter from the cable termination. It
is important that the cable termination not be altered. Both potting and
as possible.
is less than 100 ft., only 1 protector per line may be used.
secondary strikes of sufficient magnitude. AMETEK PMT Products cannot
accept responsibility for damage due to lightning or secondary surges.
drawing. Consult factory.