Safety Measures for Focus and Stage

 
The focus and/or the stage position can
be excluded when saving. In this way, you
can prevent the objective from colliding with a
specimen when you call up a memory location
that moves the objective far downwards. This
also prevents the motorized mechanical stage
from changing its position when a scenario is
called up.
Leica SmartTouch™
Excluding the focus and/or stage
1. Touch the "Config" tab.
2. Touch the symbol for assigning the memory
locations.
3. Touch the assigned memory location.
4. Touch "Focus recall" and/or "Stage recall" so
that these two components are no longer
controlled by the memory.