Saluki SAV2040X User Manual page 22

Electronic calibration kit module
Table of Contents

Advertisement

Tel: 886.909 602 109 Email: sales@salukitec.com
www.salukitec.com
Perform Calibration Tasks Module
After successfully connecting with VNA, it will enter the calibration task module. The normal initial interface of this module is shown in
the figure:
The leftmost list box allows you to select the type of calibration. For example, if you want to perform dual-port calibration, select "2 Port
Cal" in the left list, and the right side will display the default vector network port and the calibration port corresponding situation. (1, 2)
is the port number of the vector network analyzer, (A, B) corresponds to the two ports A and B of the calibration part, here you need to
set according to the specific connection, for example, the vector network port 1 is connected to the calibration part B port, B should be
selected in the second list box of the first row.
You can check [Auto Check Port] to specify automatic port connection detection. The software will combine specific algorithms to
determine whether the port pair specified by the user is correct. If not, it will be modified for measurement. If it is not checked, the
software will directly perform measurement regardless of port correspondence. At this time, make sure that the vector network port
corresponds to the calibration unit port correctly.
Click the [Measure] button to perform calibration measurement. The measurement completion interface is shown in the figure below.
Click the [Done] button, and the entire calibration process is completed.
Note: After the calibration is completed, be sure to click the [Done] button to end the calibration. Otherwise, the calibration status of
VNA cannot be saved.