Electronic calibration kit module
Table of Contents

Advertisement

Tel: 886.909 602 109 Email: sales@salukitec.com
www.salukitec.com
Identification and Connection of VNA
Open the electronic calibration software. This software first initializes the system parameters, reads the relevant files, and then
automatically scans the GPIB bus or the vector network analyzer available on the network.
The connection interface is shown below:
If the VNA is connected to the industrial computer through the network, please fill in the IP address and port number. If the software is
used in the VNA itself, please fill in the default IP address (127.0.0.1), the port number is provided by the instrument manufacturer,
and Agilent is generally "5025". If you connect via a GPIB cable, please fill in the GPIB address set at the vector terminal, and click the
[Connect] button. If the connection is successful, go directly to the next step. If the connection is not successful, the display interface is
as shown in the figure below.
At this time, it is necessary to confirm whether the instrument is connected correctly, whether the port connector is tightened, and
whether the corresponding address is correct. After the connection is successful, if the driver supports this instrument model, click
[Next] to directly enter the next module, if not, the display interface is as follows:
If the connected instrument belongs to a certain product series listed on the right, please select [Add Support], and the [Add] button on
the right will become available. After selecting the corresponding product series and clicking [Add], the software can support this
model. After the software is restarted, the software will support this model by default.