Never load more than the max. capacity to a measuring cell. Overloaded measuring cells no longer have the required accuracy. Overloaded or deformed measuring cells are no longer allowed to use and must be changed quickly. CT P1 / CT P2 -BA-e-1810...
MΩ ≥5000 ( at 50VDC ) Insulation resistance Compensated temperature °C -10 ~+40 Operating temperature °C -35 ~ +65 Storage temperature °C -40 ~ +80 Recommended torque on M12: 75 M18:500 M24:750 fixation bolts CT P1 / CT P2 -BA-e-1810...
Page 5
7.1 4-wire diagram Input(+) Output(-) White Input (-) Black Output(+) Green 7.2 6-wire diagram (option on request) Input(+) Sense(+) Blue Output(-) White Sense(-) Yellow Black Input(-) Output(+) Green 8 Dimensions in mm Dimension ØD Capacity CT P1 / CT P2 -BA-e-1810...
38.1 18.8 17.5 18.8 4.8 19.8 5000kg x 1.5 10000kg 222.3 25.4 50.8 120.7 101.6 50.8 52.4 50.8 25.4 24 x 2 9 Certificate of Compliance The Certificate of Compliance can be downloaded from the website www.sauter.eu CT P1 / CT P2 -BA-e-1810...
This manual is also suitable for:
Ct p2Ct 500-3p1Ct 1000-3p1Ct 1500-3p1Ct 2500-3p1Ct 3000-3p1...