Construction And Function - ICS Schneider Messtechnik induQ IVMZ.2 Series Operating Manual

Magnetic inductive flow sensor
Table of Contents

Advertisement

— IVMZ.2
3

Construction and function

Components:
Housing:
The Housing consists of plastic and has the
IP65 degree of protection.
Electrical connection:
The electrical connection is made via 4-pin
plug M12x1.
Process connection:
The process connections are available in
different sizes.
Operation / flow indicator LED.
Construction:
The measuring tube with its earthing sleeves and electrodes passes
through the housing and forms the external process connection of
the IVMZ.2.
A magnetic field for the measurement process is generated inside the
housing, which also contains the sensor and signal conditioning cir-
cuitry.
The two stainless steel electrodes are located in the middle of the
measuring tube between the earthing sleeves.
The IVMZ.2 does not need any moving parts to make measurements. The inside of the
meas-uring tube is completely open, allowing the fluid to flow unhindered through the
measuring tube.
Function:
The magnetic inductive flow sensor functions according to the induction principle. A DC volt-
age is generated by the movement of a conductor in a magnetic field:
The measuring tube of the IVMZ.2 is located in a
mag-netic field (B).
An electrically conductive medium (V) flows through
the measuring tube. The positive and negative charge
carriers are oppositely deflected.
A voltage (U) is generated at right angles to the mag-
netic field, which is picked up by the two electrodes.
Thereby, the induced voltage is proportional to the av-
erage flow velocity of the liquid.
The electronics of the IVMZ.2 converts the induced
voltage into a flow-proportional frequency signal.
Technical changes reserved
Construction and function
- 7 -

Advertisement

Table of Contents
loading

Table of Contents