ICS Schneider Messtechnik induQ IVMZ.2 Series Operating Manual

ICS Schneider Messtechnik induQ IVMZ.2 Series Operating Manual

Magnetic inductive flow sensor

Advertisement

Quick Links

Operating manual
Operating manual .................................................. page 1 - 20
Magnetic Inductive Flow Sensor
Series IVMZ.2
Types VMZ03 / VMZ06 / VMZ08 / VMZ15 / VMZ20 / VMZ25

Advertisement

Table of Contents
loading

Summary of Contents for ICS Schneider Messtechnik induQ IVMZ.2 Series

  • Page 1 Operating manual Operating manual ..........page 1 - 20 Magnetic Inductive Flow Sensor Series IVMZ.2 Types VMZ03 / VMZ06 / VMZ08 / VMZ15 / VMZ20 / VMZ25...
  • Page 2: Table Of Contents

    — IVMZ.2 Table of contents page About this operating manual ..................3 Device description ..................... 4 1.1 Delivery, unpacking and accessories ..............4 1.2 Intended use ......................5 1.3 Exclusion of liability ....................5 Safety instructions ..................... 6 Construction and function ..................7 Installation of IVMZ.2 ....................
  • Page 3: About This Operating Manual

    — IVMZ.2 About this operating manual About this operating manual The operating manual is aimed at specialists and semi-skilled personnel. • • Before each step, read through the relevant advice carefully and keep to the specified order. Thoroughly read and understand the information in the section “Safety instructions”. •...
  • Page 4: Device Description

    Device description — IVMZ.2 Device description of the IVMZ.2 series is a flow sensor without moving parts. The meas- urement is performed using magnetic induction. The IVMZ.2 is used for measuring or metering water and electrically conductive fluids. The compact design and independence from the intake and discharge sections allows the IVMZ.2 to be used under a variety of conditions.
  • Page 5: Intended Use

    — IVMZ.2 Device description Accessories:  Connection cable with moulded M12x1 coupling socket.  M12x1 coupling socket as component. 1.2 Intended use The magnetic inductive flow sensor IVMZ.2 must only be used for measuring and metering liquids with a minimum conductivity of 20 μS/cm. WARNING! No safety component! The magnetic inductive flow sensor of the series IVMZ.2 is no safety component in ac- cordance with Directive 2006/42/EC (Machine Directive).
  • Page 6: Safety Instructions

    Safety instructions — IVMZ.2 Safety instructions Before you install the IVMZ.2, read through this operating manual carefully. If the instructions con-tained within it are not followed, in particular the safety guidelines, this could result in danger for people, the environment, and the device and the system it is connected to. The IVMZ.2 corresponds to the state-of-the-art technology.
  • Page 7: Construction And Function

    — IVMZ.2 Construction and function Construction and function Components:  Housing: The Housing consists of plastic and has the IP65 degree of protection.  Electrical connection: The electrical connection is made via 4-pin plug M12x1.  Process connection: The process connections are available in different sizes.
  • Page 8: Installation Of Ivmz.2

    Installation of IVMZ.2 — IVMZ.2 Installation of IVMZ.2 Before installing, check that  the wetted materials of the device are suitable for the media being used ( § 9.2 “Materials table”).  the equipment is switched off and is in a safe and de-energised state. ...
  • Page 9: Installation Instructions

    — IVMZ.2 Installation of IVMZ.2 4.1 Installation instructions CAUTION! Risk of malfunction due to external mag- netic fields! Magnetic fields close to the device can cause malfunctions and should be avoided.  Ensure that no external magnetic fields are present at the instal-lation site of the IVMZ.2.
  • Page 10: Assembly

    Installation of IVMZ.2 — IVMZ.2 4.2 Assembly The IVMZ.2 is installed directly into the pipeline. The compact design and lightweight of the unit make wall mounting unnecessary. IMPORTANT NOTICES: • Only use suitable gaskets for installation. • Observe the flow direction indicated on the type plate. •...
  • Page 11: Electrical Connection

    — IVMZ.2 Electrical connection Electrical connection The electrical connection of the IVMZ.2 is made via the 4-pin plug M12x1 at the top of the housing. The wiring of the IVMZ.2 depends on the ordered version. A distinction is made between fre-quency output or analogue output and frequency output.
  • Page 12: Wirings

    Electrical connection — IVMZ.2 5.1 Wirings Pinout: The pinout differs according to the chosen configuration of the device. Pinout: Possible pinout: Pin 1: Pin 2: n. c. (not connected) / Analogue U/I Pin 3: Pin 4: Frequency M12x1  Connect the connection cable according to your version and the pinout on the type plate. Supply voltage: IVMZ.2 with frequency output: Push-Pull:...
  • Page 13: Commissioning And Measuring Operation

    — IVMZ.2 Commissioning and measuring operation Commissioning and measuring operation Before switching on the IVMZ.2 for the first time, please follow the instructions in the following section. 6.1 Commissioning Check that  the IVMZ.2 has been installed correctly and that all screw connections are sealed.
  • Page 14: Maintenance And Cleaning

    Maintenance and cleaning — IVMZ.2 Maintenance and cleaning Maintenance: The IVMZ.2 is maintenance-free and cannot be repaired by the user. In case of a defect, the device must be replaced or sent back the manufacturer for repair. CAUTION! Material damage! When opening the device, critical parts or components can be damage.
  • Page 15: Disassembly And Disposal

    — IVMZ.2 Disassembly and disposal Disassembly and disposal CAUTION! Risk of injury! Never remove the device from a plant in operation.  Make sure that the plant is shut down professionally. Before disassembly: Prior to disassembly, ensure that  the equipment is switched off and is in a safe and de-energised state. ...
  • Page 16: Technical Data

    Technical data — IVMZ.2 Technical data The technical data of customised versions may differ from the data in these instructions. Please observe the information specified on the type plate. 9.1 Characteristics IVMZ.2 Type IVMZ03 IVMZ06 IVMZ08 IVMZ15 IVMZ20 IVMZ25 Measurement device characteristics Flow range [l/min] 0.1…2 0.25..5...
  • Page 17: Materials Table

    — IVMZ.2 Technical data Type IVMZ03 IVMZ06 IVMZ08 IVMZ15 IVMZ20 IVMZ25 Process variables Medium to measure Water and other conductive liquids - Conductivity > 20 μS/cm - Temperature -15…80 °C (non-freezing) (PVDF measuring tube) - Temperature -15…60 °C (non-freezing) (POM measuring tube) Ambient temperature -15…60 °C Storage temperature...
  • Page 18: Dimensions

    Technical data — IVMZ.2 9.3 Dimensions IVMZ.2 DN 3 / DN 6 /DN 8 / DN 15 / DN 20: VMZ06 M12x1 Type VMZ03 G⅜ B Ø 3 VMZ06 G½ B Ø 8 x 2 VMZ08 G½ B Ø 8 VMZ15 G¾...
  • Page 19 — IVMZ.2 For your notes Technical changes reserved - 19 -...