CORRECTING COMMON PROBLEMS

If a solution is not readily apparent, use the table below to help identify and correct specific malfunctions. After you
correct the malfunction, verify that the generator successfully completes the self-test.
Situation
Generator does not
respond when
turned on
Generator is on, but did
not complete the
self-test.
Activation and / or
alarm tones do
not sound; speaker is
malfunctioning.
52
Possible cause
Disconnected power cord, faulty wall
receptacle, or faulty power cord
Fuse drawer is open or fuses blown
Loose or disconnected internal cables
Faulty power entry module or
Connections
Faulty power switch
An alarm condition exists.
Loose or disconnected internal cables
Faulty power switch
Main board malfunction
Display board malfunction
Loose or damaged connection between
speaker board and main board
Loose or disconnected cable between
main board and display board
Main board malfunction
Display board malfunction
Recommended action
1. Check power cord connections (generator and wall
receptacle).
2. Connect the power cord to a functional wall receptacle. If
necessary, replace the power cord.
1. Close the fuse drawer. If necessary, replace
the fuse(s).
2. If a problem persists, use a backup
generator.
Check all internal connections.
Check the power entry module and its cable
connections.
Replace the power switch.
Check the display for an error code. Note
the number and refer to Error Code list.
Check and correct all internal connections.
Replace the power switch.
Replace the main board.
Replace the display board.
Check / connect all connections from the
speaker board to the main board.
Check / connect ribbon cable from the main
board to the display board.
Replace the main board.
Replace the display board.