Advertisement

3. Allow to remain visibly wet for 2 minutes, per CaviWipes manufacturer recommendation.
4. Allow to air dry at ambient temperature.
The following solutions have been tested for safely cleaning the CanarE, CanarE Charger, and
Clinic Cart. Wipe the devices after each use with a soft cloth dampened with one of these
solutions:
1.
1:10 Bleach Solution
2.
1:10 Mild Detergent Solution
3.
Ammonia Solution (RTU)
4.
70% Isopropyl Alcohol
5.
0.5% Hydrogen Peroxide
6.
Phenolic germicidal detergent solution (RTU)
7.
Iodophor germicidal detergent solution (RTU)
WARNING: Do not immerse the CanarE, CanarE Charger, or accessories in water or cleaning
agents as this could cause electrocution.
NOTE: The CanarE and accessories are resilient to dust, lint, and light (including sunlight).
However, lint or dust may get into the USB ports or the ports may become degraded and
interfere with power to or communications with the CanarE or accessories. Do not attempt to
clean or repair the USB ports. Instead, contact Canary Medical for a replacement.

10.3. Clinic Cart Maintenance

Before first use and after periods of non-use, check the Clinic Cart for proper operation. If the
Cart does not operate correctly, contact Canary Medical.
Check the overall appearance of the Cart for any obvious damage or wear.
Periodically inspect the casters and brakes. Remove all foreign material picked up by the
casters and brakes.