Attaching the CanarE to a Patient

With the patient sitting or standing, palpate the knee to locate the patient's left or
right tibial tuberosity according to the Test Plan created with the Clinic Base Station
System (Left Knee or Right Knee selected from the Clinic App Patient Test Plan).
NOTE: The CanarE is an external wearable device intended to mimic a CTE implant
by positioning it in the approximate location of a CTE implant as shown in Figure 8.
Figure 8: CanarE Placement Positioned in the Approximate Location of a CTE Implant
Place the CanarE directly onto the patient's knee with the superior surface of the
CanarE located on the inferior most aspect of the tibial tuberosity as shown in Figure
9.
NOTE: Do NOT place the CanarE over patient clothing. This may cause an
unintended shift of the CanarE during testing, which will affect test result data.
Align the CanarE alignment mark with the axis of the tibia by palpating the patient's
tibia inferior of the CanarE. Attach the CanarE as shown in Figure 9.
Page | 17
Table 4: CanarE LED Charging Status
Definition
Charging: CanarE battery status is between 85% -100% charged
Charging: CanarE battery status is between 40-84% charged
Charging: CanarE battery status is between 20-39% charged
Charging: CanarE battery status is between 0-19% charged
Instructions for Use
Rev. 4
CanarE System