3
C H A P T E R
created a unique twisted-pair cable with braided outer shield. The Model CC305S
cable has been designed to provide the best possible rejection of most noise
signals. Although every situation is different, we feel that this cable provides the
best shielding over the widest range of laboratory conditions.
The mount chassis is grounded through pin 3 of the Channel 1 laser connector.
This pin must be grounded to ground the mount chassis. Standard ILX Lightwave
cables and laser current sources ground pin 3 automatically. The pin-out
configurations of the 9-pin D-sub connectors on the LDM-4616 mount are shown
below in Figures 3.1 and 3.2.
10
LDM-4616
O P E R A T I O N
Connecting the LDM-4616 Mount to a Laser Diode Controller
Figure 3.1 Back Panel TEC Connector
Figure 3.2 Back Panel Laser Connector