:FRF?
Harmonics
To return harmonics, first the number of harmonic and the scope need to be selected
and then they need to be added to the list of results on the display.
:HMX:VLT:SEQ 0
:HMX:VLT:RNG 9
:SEL:VHM
Now, assuming :SEL:CLR has not be issued after example 1, then the following results
would be returned by FRD?
Vrms, Arms, Freq, Watt, VA, Var, PF, Vpk+,Apk+, Vh1 Mag, Vh1 phase, Vh2 Mag, Vh2
phase, .... Vh9 Mag, Vh9 phase.
Standby power
First, select standby power mode
:MOD:SDY:PER 60
:MOD:SDY
:SEL:CLR
:SEL:VLT
:SEL:WAT
:SEL:FRQ
:SEL:VCF
Page 5-6
Returns the results selected for confirmation using the
label that appears on the display. In this case will return,
"Vrms, Arms, Freq, Watt, VA, Var, PF, Vpk+,Apk+
Select odd and even harmonics (use 1 to select odd
harmonics only)
Return all harmonic from 1 to 9.
Add Voltage harmonics to the list.
Set the standby power mode period to 60 seconds.
Clears selection of results
Selects Vrms
Selects Watts
Selects Frequency
Selects Volts crest factor
© Voltech 2007
PM1000+ Power Analyzer User Manual