Presys DCY-2057 Technical Manual page 62

Pressure controller
Table of Contents

Advertisement

PRESYS
|
Instruments
• 500 psi Sensor
• 1000 psi Sensor
Voltage Input Calibration (0 to 55mV)
In order to calibrate the voltage input at 0 to 55mV connect an accurate DC
voltage source to channel 2 (terminals 5(+) and 6(-)). It is required 2 voltage
references listed in table 5.
Calibration of voltage input (0 to 5V)
In order to calibrate the voltage input at 0 to 5V connect an accurate DC
voltage source to channel 2 (terminals 5(+) and 6(-)). It is required 2 voltage
references listed in table 6.
DCY - 2057 Pressure Controller
Reference
0.00 psi
100.00 psi
200.00 psi
300.00 psi
400.00 psi
500.00 psi
Reference
0.0 psi
200.0 psi
400.0 psi
600.0 psi
800.0 psi
1000.0 psi
Reference
0.000 mV
50.000 mV
Table 5 - Required voltages for the calibration
of voltage input from 0 to 55mV
Mnemonic
0psi
100psi
200psi
300psi
400psi
500psi
Mnemonic
0psi
200psi
400psi
600psi
800psi
1000psi
Mnemonic
C. 0nV
C.50nV
Maintenance