Problems and Solutions

The status indicator fails to light when the software is started, or the front-
end commands and so on do not appear where they should
2
The I
C cable or one or more BNC-to-BNC cables from the front-end to the
PowerLab are not connected, have been connected incorrectly (to the wrong
input, for instance), or are loose.
Turn everything off. Check to see that all cables are firmly seated and
screwed in. BNC cables from the Bio Amp must be connected to a positive
input on the PowerLab. Make sure the input is the same channel from
which you expect to use the front-end in the software. Start up again to see
if this has fixed the problem.
You are using an early version of LabChart or Scope.
Upgrade to the latest version of the software. Contact your
ADInstruments representative for information.
2
The BNC or I
C cable is faulty.
Replace the cable and try again. Immediately label all cables proved faulty
so that you don't use them again by accident.
The Bio Amp is faulty
This is the least likely event. If the front-end will not work properly after
the previous measures, then try using it on another PowerLab. If the same
problems recur with a second PowerLab, the front-end may be faulty.
Contact your ADInstruments representative to arrange for repairs.
On starting up the software, an alert indicates that there is a problem with
the front-end or driver
The correct Bio Amp driver is not installed on your computer (it should be in
the Essential Files folder in the LabChart or Scope folder).
Reinstall the software.
You are using an early version of LabChart or Scope.
Upgrade to the latest version of the software. Contact your
ADInstruments representative for information.