1.1.2 Output Ranges
Measurement Range
Thermocouple Output
Current Output (switch selectable)
3LT
3P7
3G5
3HT
Spectral Response
Accuracy
Repeatability
1-4
2LT
-18 to 760°C (0 to 1400°F), Type J
0 to 870°C (32 to 1600°F), Type K
2P7
10 to 360°C (50 to 650°F), Type J
2G5
260 to 1650°C (500 to 3000°F), Type R
2HT
500 to 1700°C (1000 to 3000°F), Type R
Range Setting
1
2
3
4
Range Setting
1
2
Range Setting
1
2
Range Setting
1
2
LT Versions
P7 Versions
G5 Versions
HT Versions
± 1% of reading or 1.4°C (2.5°F) @ 25°C (77°F)
whichever is greater
± 0.5% of Reading, ± 1 digit °C or °F
Thermalert ET Series Operator's Manual
°F Version
°C Version
0 to 200
0 to 400
0 to 800
0 to 1400
°F Version
°C Version
50 to 350
50 to 650
°F Version
°C Version
500 to 1500
500 to 3000
°F Version
°C Version
1000 to 2000
1000 to 3000
8 to 14 microns
7.9 microns (nominal)
5.0 microns (nominal)
2.2 microns (nominal)
0 to 100
0 to 200
0 to 400
0 to 800
10 to 110
10 to 360
260 to 815
260 to 1650
500 to 1000
500 to 1700