Chapter 8
Counters

Frequency Measurement

NI USB-621x User Manual
You can use the counters to measure frequency in several different ways.
You can choose one of the following methods depending on your
application:
Method 1: Measure Low Frequency with One Counter—In this
method, you measure one period of your signal using a known
timebase. This method is good for low frequency signals.
You can route the signal to measure (F1) to the Gate of a counter. You
can route a known timebase (Ft) to the Source of the counter. The
known timebase can be 80MHzTimebase. For signals that might be
slower than 0.02 Hz, use a slower known timebase.
You can configure the counter to measure one period of the gate signal.
The frequency of F1 is the inverse of the period. Figure 8-10 illustrates
method 1.
F1
Gate
Ft
Source
Single Period
Measurement
Interval Measured
F1
1
2
3
Ft
Period of F1 =
Frequency of F1 =
Figure 8-10. Method 1
8-10
...
...
N
N
Ft
Ft
N
ni.com