Specifications; Specifications - Analyser; Specifications - Probe - Novatech 1635 User Manual

Water vapour analyser
Table of Contents

Advertisement

1.2

SPECIFICATIONS

1.2.1 SPECIFICATIONS – ANALYSER
Model:
Range:
Output:
Displays:
Alarms:
Speed of response:
Accuracy:
Power Supply:
Environmental Rating
Degree of Protection
1.2.2 SPECIFICATIONS – PROBE
Model:
Length:
Temperature Range:
Process Connection:
Filter:
Response Time:
Operating Temperature:
Heater Voltage:
Maximum Head Temperature: 150°C with screw terminals / 100°C with weatherproof connector
Reference Gas Connection:
Probe Cable:
Probe Life:
August 2009
1635 Water Vapour Analyser
1635
Keyboard selectable for the following ranges: -
0–100% volume of water vapour
0-100% Relative humidity (RH)
0-10,000 g/Kg
-50-+100 °C Dew point
all with adjustable span and zero.
Isolated 4–20 mA DC linear
Various, Refer Section 3.1
Various, Refer Section 3.3.1 & 3.3.2
Typically 20 seconds
See graph above
110/240 VAC, 50/60 Hz 105VA
Operating Temperature: -25 to 55°C (-15 to 130°F)
Relative Humidity: 5 to 95% (non-condensing)
Vibration: 10 to 150Hz (2g peak)
IP65 without reference air pump
IP54 with reference air pump
1231
250–1000 mm.
0 to 900°C
1.5" BSP or NPT
Sintered titanium alloy particulate filter, removable, 30 or 15 micron
Typically less than 4 seconds
720°C with integral heater
240 (25% duty cycle) / 110 VAC, 100 watts
1/8" NPTF.
Supplied with weatherproof connector to specified length, maximum 100 metres
Typically 2 to 3 years.
7