Note: If the Set-up menu does not match the functions in your instrument, check the software version number -
On Page 2 of this manual and
On the EPROM in the socket labelled IC12 (In the window, top centre of the door PCB)
4.2
SET-UP & RUN MODES
For the SET-UP mode keyboard to operate, press the SET-UP/RUN button. The SET-UP light will come on when the
SET-UP mode has been entered.
SET-UP mode cannot be entered if the keyboard lock switch on the inside of the analyser is in the UP position. The
keyboard lock switch can be found on the door PCB (1630-2), on the lock side, at the top. If access is attempted while
the keyboard is locked, the message 'Illegal Access' will be displayed.
The temperature of a heated probe may fall if the set-up mode is used for more than 2 minutes.
While the analyser is in SET-UP mode the outputs will be frozen. Key functionality while in setup mode is indicated
on the keypad by the BLUE labels. If no buttons are pressed for 1 minute, the analyser will automatically revert to
RUN mode.
If an auto-calibration check is scheduled to occur while the analyser is in SET-UP mode, it will be delayed until the
analyser returnes to RUN mode.
To cancel an automatic calibration check cycle, press AUTO CAL button while in RUN mode.
4.3

FUNCTION SELECT

When returning to SET-UP mode, the analyser will return to the last selected set-up set-up function.
To select other functions, operate the 'FUNCTION
decrement to the previous function.
4.4

ENTER OPTION OR VALUE

A. Options.
To step through the available options for each function press the 'OPTION
When the required option is selected press the 'ENTER' button. An asterisk will then appear alongside the option
selected. When stepping through the set-up functions, the display will always first indicate the last options entered.
The 'Lower Line Select' and 'Data To Print' Set-up 40 and 58 are multiple options. One or more options may be
selected for these functions.
B. Values
To set a value for a particular function press the 'OPTION
to decrease the value. A momentary press will change the value one digit. Holding the button will change the value
more quickly. Once the correct option or value is displayed it can be entered into the analyser's memory by pressing
the 'ENTER' button. When a value has been entered an asterisk will appear at the R.H.S. of the lower line.
August 2009
1635 Water Vapour Analyser
' button to increment to the next function, or 'FUNCTION
' button to increase the value and the 'OPTION
' or 'OPTION
' buttons.
' to
' button
33