November 2022
3

Set jumpers and switches

3.1
Set alarm and write protection on the circuit board
If alarm and security jumpers are not set, the transmitter operates with the
default alarm condition HIGH and Security OFF.
Prerequisites
Write protection must be set after configuration.
Procedure
1. Remove the cover on the circuit side (see label marked circuit side).
2. To set the 4-20 mA alarm output to LOW, move the alarm switch to
the LOW position.
3. To enable the security write protection feature, move the write
protect switch to the ON position.
4. Replace the cover and tighten securely.
Figure 3-1: Circuit Board
3.2
Set alarm and write protection on the LCD display
Prerequisites
To have the LCD display override the circuit board settings, the write
protection switch on the circuit board needs to be in the OFF position and
the alarm switch on the circuit board needs to be in the HIGH position.
Procedure
1. To set the 4-20 mA alarm output to LOW, place jumper between the
right and center hole position.
2. To enable the security write protection feature, place jumper
between the left and center hole position - ON.