3. MANTENANCE
3.3 Replacing the Purifi cation pack
Follow the instructions below.
When a purifi cation pack is fi tted into the system it will record the serial number of the
Purifi cation Pack. After 8760 hours (12 months) a reminder is activated to replace the Purifi cation
Pack. Status light will be blinking white showing system needs attention.
1. Process button (OFF). Rotate the handle
to the unlock position, remove Purifi cation Pack.
3. Remove transit plugs and insert a new
Purifi cation pack.
5. Press for 10 seconds to enter menu.
2. Pull down and tilt towards to remove.
4. Process button (ON), normal operation resumes.
.
6. Press to scroll, press and hold to confi rm
the reset pack reminder. Normal operation resumes.
Chorus 2 Operator Manual MANU40003 Version 03 3/19 Page 5