9. Technical data

9.
Technical data
Instrument
Characteristic
Value
Parameters
Temperature (°C / °F)
Measuring range
Pt100 probe: -50...+300°C / -58...+572 °F
NTC probe: -50...+120°C / -58...+248 °F
Resolution
0.1 °C / 0.1 °F
System accuracy
Accuracy
See
ë
Probe
1x mini DIN socket for Pt100 or NTC temperature probe
Measuring rate
2/s
Operating temperature range
Storage temperature
-30...+70 °C / -22...+158 °F
Voltage supply
1x 9 V monobloc battery/rech. battery
Battery life
approx. 70 h
Protection class
with TopSafe (accessory part) and probe connected: IP 65
EC Directive
2014/30/EC
Warranty
2 years
System accuracy
Measuring range
Instrument
Measuring instrument + NTC temperature probe
-50.0°C...-25.1°C
±1% of reading ±0.7% of reading
-25.0...+40.0°C
±0.2°C
+40.1...+80.00°C
±0.3°C
+80.1...+120.0°C
±0.5°C
Measuring instrument + Pt100 temperature probe
-50.0...-25.1°C
±0.2°C
-25.0...+40.0°C
±0.2°C
+40.1...+140.0°C
±0.2°C
+140.1...+200.0°C
±0.2°C
+200.1...+300.0°C
±0.3°C
-20...+50 °C / -4...+122 °F
Probe
System
±1. 8 % of readi n g
±0.2°C
±0.5°C
±0.4°C
±0.8°C
±0.6°C
±1.2°C
±0.3°C
±0,6°C
±0.2°C
±0.5°C
±0.4°C
±0.7°C
±0.6°C
±0.9°C
±0.8°C
±1.2°C
Declaration of Conformity