Page 3
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others.
Page 4
Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized personnel.
Page 5
Italic text is used in this manual to quote the referenced manuals, chapters, sections and formulas. Bold text is used to indicate the screen texts and names of hard keys. ■ → is used to indicate operational procedures. ■ BeneVision N1 Patient Monitor Operator’s Manual...
3.1 Equipment Preparation Safety Information ............................3 - 1 3.2 Unpacking and Checking .....................................3 - 1 3.3 Environmental Requirements ..................................3 - 2 3.4 Setting Up the Equipment ...................................3 - 2 3.4.1 Connecting the AC Mains ................................3 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Page 7
3.14 The OxyCRG Screen (only available for the independent external display) ................. 3 - 18 3.14.1 Entering the OxyCRG Screen ..............................3 - 18 3.14.2 OxyCRG Events ..................................... 3 - 18 3.14.3 The Display of the ABD Event Area ............................3 - 19 BeneVision N1 Patient Monitor Operator’s Manual...
Page 8
6.2 Network Safety Information ..................................6 - 1 6.3 Connecting the Monitor to the CMS ................................6 - 1 6.3.1 Setting the CMS ....................................6 - 1 6.3.2 Selecting a CMS ....................................6 - 1 6.4 Connecting the eGateway ...................................6 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Do not place the equipment or accessories in any position that might cause it to fall on the patient. • Do not start or operate the equipment unless the setup was verified to be correct. BeneVision N1 Patient Monitor Operator’s Manual 1 - 1...
• The software equipment copyright is solely owned by Mindray. No organization or individual shall resort to modifying, copying, or exchanging it or to any other infringement on it in any form or by any means without due permission.
Page 23
DEFIBRILLATION-PROOF TYPE BF APPLIED PART PART Lock; tighten Zero key Locking Unlocking Direction and angle off rotation Calibration Start Stop Equipotentiality Polarity of d.c. power connector Menu Video output Gas outlet Gas inlet BeneVision N1 Patient Monitor Operator’s Manual 1 - 3...
Page 24
The product bears CE mark indicating its conformity with the provisions of the Council Directive 93/42/EEC concerning medical devices and fulfil the essential requirements of Annex I of this directive. Note: The product complies with the Council Directive 2011/65/EU. 1 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
PAWP is intended for adult and pediatric patients only ■ The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment. It should be used by clinical professionals or under their guidance.
External DC power input connector: connects the N1 to the AC adapter NOTE • Dry the Multi-pin connector of the N1 before connecting the N1 to the Modular Rack or Dock in case of water spray. 2 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
N1 so that you can remove the N1 from the host monitor, Dock or Modular Rack. Modular Rack Modular Rack is used for connecting a N1 and an external parameter module. BeneVision N1 Patient Monitor Operator’s Manual 2 - 3...
Infrared filter: used for communication between the Modular Rack and external parameter module. Contact: power input connector of the external parameter module. Dock Dock is used to connect the N1 or Modular Rack. 2 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
Page 29
Multi-pin connector: power input and communication connector of the N1. Connection status LED: it is on when the N1 is properly connected to the Dock. External power LED: it is on when the external AC power supply is connected. BeneVision N1 Patient Monitor Operator’s Manual 2 - 5...
The monitor can connect the following external parameter modules to perform CO monitoring and IBP monitoring through the Modular Rack. Sidestream CO Mainstream CO Microstream CO IBP module module module module 2 - 6 BeneVision N1 Patient Monitor Operator’s Manual...
Press and hold the latch at the bottom of the N1 or parameter module. If the external module is locked to the Modular Rack, unlock it first. Pull the N1 or parameter module out as indicated. BeneVision N1 Patient Monitor Operator’s Manual 2 - 7...
To prevent N1 from falling off, do not press the release button while transferring N1 with the Modular Rack and Dock. 2.7.5 Installing the N1 to the Dock You can also install N1 directly to the Dock as shown below: 2 - 8 BeneVision N1 Patient Monitor Operator’s Manual...
BeneVision N series and BeneView T series monitor that can be used as N1 host monitor are as follows: ■ BeneVision 22, BeneVision N19, BeneVision N17, BeneVision N15, BeneVision N12, and BeneVision N12C BeneView T5, BeneView T6, BeneView T8, and BeneView T9 ■ BeneVision N1 Patient Monitor Operator’s Manual 2 - 9...
To prevent N1 from falling off, catch it with another hand while pulling it out from the SMR. 2.8.3 Connecting N1 to the Host Monitor through the Dock To connect the N1 to the host monitor through the Dock, follow this procedure: 2 - 10 BeneVision N1 Patient Monitor Operator’s Manual...
Transport Dock. 2.10 Input Devices The monitor allows data entry through touchscreen, keyboard, mouse, and barcode reader. 2.11 Printing Devices You can use Mindray specified printer to output patient information and data. BeneVision N1 Patient Monitor Operator’s Manual 2 - 11...
Use only installation accessories specified by Mindray. • The equipment software copyright is solely owned by Mindray. No organization or individual shall resort to modifying, copying, or exchanging it or to any other infringement on it in any form or by any means without due permission.
• Always use the accompanying power cord delivered with the monitor. • Always use the AC adapter specified by Mindray. • Before connecting the equipment to the AC mains, check that the voltage and frequency ratings of the power line are the same as those indicated on the AC adapter and Dock.
The touchscreen is not used in 60 seconds. ■ ■ Select the Unlock quick key , and swipe the slider up as instructed. BeneVision N1 Patient Monitor Operator’s Manual 3 - 3...
Clearing Old Data Formats (for the Mindray Custom 2D Barcode Reader) If you are using the Mindray custom 2D barcode reader (Model HS-1R or HS-1M), before using the it for the first time, clear old data formats and configure the barcode reader.
1D Scanner: select this option when you are using a 1D scanner or a 2D scanner other than the Mindray custom 2D scanner. ◆ 2D Scanner: select this option when you are using the Mindray custom scanner. NOTE • When you set Scanner Type to 2D Scanner, default settings are applied to Data Encoding Type and DataParseMode.
For more information, see 3.11.3Accessing Parameter Setup Menus. 3.6.1 On-screen Symbols The following table lists the on-screen symbols displayed on the system status information area: Symbol Description Symbol Description Adult, male Adult, female Pediatric, male Pediatric, female 3 - 6 BeneVision N1 Patient Monitor Operator’s Manual...
Battery fault, battery communication fault, or battery charging fault. Contact service personnel for help. 3.6.2 Menus All menus have similar style and structure, see the figure below: Menu heading Submenu tabs BeneVision N1 Patient Monitor Operator’s Manual 3 - 7...
The privacy mode is only available when the patient admitted by the monitor is also monitored by the CMS. The monitor continues monitoring the patient, but patient data is only visible at the CMS. 3 - 10 BeneVision N1 Patient Monitor Operator’s Manual...
Verify the night mode settings before entering the night mode. Pay attention to the potential risk if the setting value is low. 3.7.4.2 Exiting the Night Mode To cancel the night mode, follow this procedure: BeneVision N1 Patient Monitor Operator’s Manual 3 - 11...
To enter the outdoor mode, follow this procedure: Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Other tab. 3 - 12 BeneVision N1 Patient Monitor Operator’s Manual...
Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Time tab. Switch on Auto Daylight Saving Time. Adjust daylight saving time settings as necessary. BeneVision N1 Patient Monitor Operator’s Manual 3 - 13...
Switching On or Off a Parameter You can manually switch on or off a parameter when its module is connected. To do so, follow this procedure: Access Parameters On/Off by any of the following ways: 3 - 14 BeneVision N1 Patient Monitor Operator’s Manual...
You can quickly switch the normal screen and the big numeric screen by swiping left or right on the touchscreen with two fingers. You can also select the big numeric screen by proceeding as follows: BeneVision N1 Patient Monitor Operator’s Manual 3 - 15...
NOTE • You cannot change timer settings when a timer is running. • You can set Direct, Run Time, and Reminder Volume only for normal, advanced, and cycled timers. 3 - 16 BeneVision N1 Patient Monitor Operator’s Manual...
• When the N1 is connected to the Dock, the N1 can use the external screen setting of the Dock. For more information, see section 4.5Setting the Dock. BeneVision N1 Patient Monitor Operator’s Manual 3 - 17...
Zero RR, low heart rate, and low SpO2 happen at the same time. Red event NOTE • The monitor records all ABD events for OxyCRG review, but only red events displays in the ABD list of the OxyCRG screen. 3 - 18 BeneVision N1 Patient Monitor Operator’s Manual...
In case of a temporary power failure, if the power is restored within 30 minutes, monitoring will resume with all active settings unchanged; if the monitor is without power for more than 30 minutes, the monitor behaves the same as it is normally turned off. BeneVision N1 Patient Monitor Operator’s Manual 3 - 19...
The technical alarms is reset. The monitor loads the default configuration and goes to the standby mode. ■ Clear Patient Data: all patient data is deleted from the monitor. The monitor still uses the current monitor settings. BeneVision N1 Patient Monitor Operator’s Manual 4 - 1...
Input query criteria. If your monitor is connected with the ADT server, input query criteria from the Discharged Patient page. Select Search. Then a list pops up, including all the patients that meet the query criteria. 4 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Device Location tab. Input monitor name, facility name and department name. BeneVision N1 Patient Monitor Operator’s Manual 4 - 3...
N1 and the host monitor, refer to section 2.8 N1 in Use with the Host Monitor. WARNING • Do not discharge a patient before the patient is successfully transferred. 4 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
Select the Dock Setup tab→ select the Setup tab. Set the Work Mode to Host Mode. Select the IP tab. Set Network Type to LAN1 IP, or WLAN according to your network type. BeneVision N1 Patient Monitor Operator’s Manual 4 - 5...
N1, see section 3.13.2 Setting the External Display. Connecting the CMS You can connect the monitor to the BeneVision CMS. When connected to the CMS, the system provides the following function. 4 - 6 BeneVision N1 Patient Monitor Operator’s Manual...
Page 61
In case of network disconnection, the monitor can transmit the offline data to the CMS when network is reconnected. For more information on the CMS, see BeneVision Central Monitoring System Operator's Manual (PN: 046-007687- 00). BeneVision N1 Patient Monitor Operator’s Manual 4 - 7...
Changing the department will delete all current user configurations. Setting Default Configuration The monitor will load the pre-set default configuration in the following cases: ■ A patient is admitted. A patient is discharged. ■ ■ Patient category is changed. BeneVision N1 Patient Monitor Operator’s Manual 5 - 1...
Select the Main Menu quick key → from the Configuration column select Manage → input the required password → select Select Export Configuration. Select the configurations and User Maintenance Settings to export. Select Export. 5 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Select the Main Menu quick key → from the Configuration column select Manage → input the required password → select Select Modify Password. Respectively input the old password and new password. Select OK. BeneVision N1 Patient Monitor Operator’s Manual 5 - 3...
Selecting a CMS Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Network Setup tab → Central Station Setup tab. BeneVision N1 Patient Monitor Operator’s Manual 6 - 1...
You can also view this monitor from remote devices. This monitor can be viewed by at most 32 remote devices at the same time, in which eight remote devices can watch this monitor’s waveforms. 6 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Cyan The low priority alarm is occurring. The low priority alarm currently is the highest alarm level on the bed. Grey The bed is in the standby mode. BeneVision N1 Patient Monitor Operator’s Manual 6 - 3...
Page 68
To reset remote device alarms, select Alarm Reset. NOTE • You can reset remote device alarms only if the Alarm Reset by Other Bed switch is on at the remote devices. 6 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
Select the Tile Layout tab. Select the numeric area where you want to display the alarm watch tile, and then in the drop-down list, select Alarm Watch → A-W1 or A-W2. BeneVision N1 Patient Monitor Operator’s Manual 6 - 5...
You can delete certifications from the monitor, or import certifications from the USB memory. To do so, follow this procedure: Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Network Setup tab→ WLAN tab. 6 - 6 BeneVision N1 Patient Monitor Operator’s Manual...
Select the Network Setup tab → Encryption tab. Select Encryption Connection Type to set data encryption type when connecting devices: ◆ Only Private Encryption: Mindray private encryption is used to encrypt the transmitted data. You cannot connect devices supporting SSL (secure sockets layer) encryption. ◆...
MLDAP MLDAP refers to Mindray LDAP (Lightweight Directory Access Protocol). It is an independent process which can be installed on eGateway or other application server (Windows). MLDAP provides user identity and authentication. The MLDAP server is connected with the hospital LDAP server. All monitoring devices are connected to the MLDAP server to implement identity and authentication for the following operations: ■...
To do so, follow this procedure: Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Authorization Setup tab. Set Retention Time. BeneVision N1 Patient Monitor Operator’s Manual 6 - 9...
Technical alarm conditions may also be caused when an algorithm cannot classify or interpret the available data. Apart from the physiological and technical alarms, the monitor can also prompt some messages telling the system status or patient status. BeneVision N1 Patient Monitor Operator’s Manual 7 - 1...
• When multiple alarms of the same priority levels occur simultaneously and should be displayed in the same area, all the alarm messages are displayed circularly. 7 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Select the Main Menu quick key → from the Alarm column select Limits.Enter the password if required. Select a parameter tab and set alarm properties as desired. You can also change the alarm properties of individual parameter from corresponding parameter menu. BeneVision N1 Patient Monitor Operator’s Manual 7 - 3...
Setting the Reminder Tones When the alarm volume is set to zero, or the alarm is reset or switched off, the monitor issues a periodical reminder tone. The reminder tone is enabled by default. 7 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
These alarm limits will remain unchanged until you select auto limits again or adjust them manually. The monitor calculates auto limits basing on the following rules: BeneVision N1 Patient Monitor Operator’s Manual 7 - 5...
Page 79
CPP × 0.68 + 8 (CPP-15) or CPP × 0.86 + (CPP+15) or Adult: 20 to 235 (mmHg) Pediatric: 25 to175 (whicherver is (whichever is Neonate: 25 to 100 greater) smaller) 7 - 6 BeneVision N1 Patient Monitor Operator’s Manual...
Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Alarm tab → Other tab. Set ST Alarm Delay. BeneVision N1 Patient Monitor Operator’s Manual 7 - 7...
Desat alarm is permissible or not. To do so, follow this procedure: Select the Main Menu quick key → from the System column select Maintenance → input the required password → select 7 - 8 BeneVision N1 Patient Monitor Operator’s Manual...
For technical alarms, alarm sounds are paused, but alarm lamps and alarm messages remain presented. ■ ■ The remaining alarm pause time is displayed in the physiological alarm information area. ■ The alarm pause symbol is displayed in the system information area. BeneVision N1 Patient Monitor Operator’s Manual 7 - 9...
Page 83
The message Alarm Off with red background is displayed in the physiological alarm information area. ■ The alarm off symbol is displayed in the system status information area. To exit the alarm off status, press the Alarm Pause quick key again. 7 - 10 BeneVision N1 Patient Monitor Operator’s Manual...
NOTE • If a new alarm is triggered after the alarm system is reset, the alarm reset icon will disappear and the alarm light and alarm tone will be reactivated. BeneVision N1 Patient Monitor Operator’s Manual 7 - 11...
To exit the intubation mode, choose either of the following ways: From the bottom of the Resp or CO2 menu, select Exit Intubation Mode. ■ Select the Main Menu quick key →from the Alarm column select Exit Intubation Mode. ■ BeneVision N1 Patient Monitor Operator’s Manual 7 - 13...
All: the monitor sounds if an alarm occurs. ◆ High&Med: the monitor sounds if a high or medium priority alarm occurs. ◆ High Only: the monitor sounds only if a high priority alarm occurs. 7 - 14 BeneVision N1 Patient Monitor Operator’s Manual...
Confirm the alarming parameter or alarm category. Identify the source of the alarm. Take proper action to eliminate the alarm condition. Make sure the alarm condition is corrected. For more information, see D Alarm Messages. BeneVision N1 Patient Monitor Operator’s Manual 7 - 15...
Periodically inspect the electrode application site to ensure skin integrity. If the skin quality changes, replace the electrodes or change the application site. • Interference from ungrounded instrument near the patient and electrosurgery interference can induce noise and artifact into the waveforms. BeneVision N1 Patient Monitor Operator’s Manual 8 - 1...
Store the electrodes at room temperature. • Only open the electrode package immediately prior to use. • Never mix patient electrode types or brands. This may lead to problem due to impedance mismatch. 8 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
RA placement: directly below the clavicle and near the right shoulder. ■ LA placement: directly below the clavicle and near the left shoulder. LL placement: on the left lower abdomen. ■ BeneVision N1 Patient Monitor Operator’s Manual 8 - 3...
Page 92
V7R placement: on posterior chest at the right posterior axillary line in the fifth intercostal space. NOTE • For the 5-leadwire and 6-leadwire placement, place the precordial electrode according to the physician's preference. 8 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
To change the paced status, follow this procedure: Select the ECG numeric area or waveform area to enter the ECG menu. Select the Pacer tab. Set Paced to Yes or No. BeneVision N1 Patient Monitor Operator’s Manual 8 - 5...
From the bottom of the menu, select 12-lead. 8.5.2 Setting ECG Alarm Properties To set ECG alarm properties, follow this procedure: Select the ECG numeric area or waveform area to enter the ECG menu. 8 - 6 BeneVision N1 Patient Monitor Operator’s Manual...
Changing ECG Waveform Speed To change ECG waveform speed, follow this procedure: Select the ECG numeric area or waveform area to enter the ECG menu. Select the Setup tab. Set Speed. BeneVision N1 Patient Monitor Operator’s Manual 8 - 7...
Select the Setup tab. Switch off Smart Lead. 8.5.5 Setting the Priority of the ECG Lead Off Alarm To set the priority of the ECG lead off alarm, follow this procedure: 8 - 8 BeneVision N1 Patient Monitor Operator’s Manual...
Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Module tab → ECG tab. Set ECG Standard to AHA or IEC. Monitoring Arrhythmia Arrhythmia monitoring is intended for adult, pediatric, and neonatal patients. BeneVision N1 Patient Monitor Operator’s Manual 8 - 9...
More than two consecutive PVCs, but lower than the V brady PVCs limit, and the ventricular rate is lower than the V-Tach rate limit. Couplet A Pair of PVCs detected in between normal beats. Multiform PVC Multiform PVCs detected in Multif. PVC's Window (which is adjustable). 8 - 10 BeneVision N1 Patient Monitor Operator’s Manual...
Select the ECG numeric area or waveform area to enter the ECG menu. Select the Arrhythmia tab→ Alarm tab. Enter the password if required. For more information, refer to 6.9.3 Selecting Password for User Authentication. Set alarm properties as desired. BeneVision N1 Patient Monitor Operator’s Manual 8 - 11...
Page 100
Threshold Range Asystole Delay 3 sec to 10 sec Tachy (HR High) 60 bpm to 295 bpm Brady (HR Low) 16 bpm to 120 bpm Extreme Tachy 65 bpm to 300 bpm 8 - 12 BeneVision N1 Patient Monitor Operator’s Manual...
Page 101
6, V-Tach Rate is set to 130, V Brady PVCs is set to 5, and V Brady Rate is set to 40. If both V-Tach PVCs and V-Tach Rate are greater than or equal to the limits, a V-Tach alarm is generated. ■ BeneVision N1 Patient Monitor Operator’s Manual 8 - 13...
To set the arrhythmia alarm timeout period, follow this procedure: Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Alarm tab→ select the Other tab. 8 - 14 BeneVision N1 Patient Monitor Operator’s Manual...
The ST monitoring function is disabled by default. Before you start ST monitoring, enable the ST function. To do so, follow this procedure: Select the ECG numeric area or waveform area to enter the ECG menu. BeneVision N1 Patient Monitor Operator’s Manual 8 - 15...
Access Tile Layout by either of the following ways: ◆ Select the Screen Setup quick key → select the Tile Layout tab. ◆ Select Main Menu quick key → from the Display column select Tile Layout. 8 - 16 BeneVision N1 Patient Monitor Operator’s Manual...
Entering the ST Graphic Window (only available for the independent external display) To display ST Graphic window, follow this procedure: Select the ECG numeric area or waveform area to enter the ECG menu. Select the ST tab. BeneVision N1 Patient Monitor Operator’s Manual 8 - 17...
To set the ST alarm delay time, follow this procedure: Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Alarm tab→ select the Other tab. 8 - 18 BeneVision N1 Patient Monitor Operator’s Manual...
QRS complex. As the J point is a fixed distance away from the ST point, it can be useful to help you correctly position the ST point. (1) ISO point (2) J point (3) ST point (4) ST deviation BeneVision N1 Patient Monitor Operator’s Manual 8 - 19...
QTc is observed you should verify it to ensure that it is not caused by QRS widening. Because normal beats followed by ventricular beats are not included in the analysis, no QT measurement will be generated in the presence of a bigeminy rhythm. 8 - 20 BeneVision N1 Patient Monitor Operator’s Manual...
Select the QT numeric area, ECG numeric area or ECG waveform area to enter the ECG menu. Select the QT tab. From the bottom of the menu, select QT View . BeneVision N1 Patient Monitor Operator’s Manual 8 - 21...
You can select one lead or all leads for QT calculation. To do so, follow this procedure: Select the QT numeric area, ECG numeric area or ECG waveform area to enter the ECG menu. 8 - 22 BeneVision N1 Patient Monitor Operator’s Manual...
The ECG signal may be inaccurate due to hardware or software problems. As a result, the ECG waveform amplitude becomes greater or smaller. In that case, you need to calibrate the ECG module. To do so, follow this procedure: BeneVision N1 Patient Monitor Operator’s Manual 8 - 23...
2. Check that electrodes are not detached or dry. Perform skin preparation again as described in 8.4.1 Preparing the Patient Skin and apply fresh and moist electrodes. 3. Check that the patient cable or leadwires are not damaged. Change them if necessary. 8 - 24 BeneVision N1 Patient Monitor Operator’s Manual...
Page 113
Patient Skin and 8.4.2 Applying Electrodes . Monitor 3. Check for ECG filter setting. Set ECG Filter mode to to reduce baseline wander on the display. This page intentionally left blank. BeneVision N1 Patient Monitor Operator’s Manual 8 - 25...
Resting 12-Lead ECG Analysis Resting 12-Lead ECG Analysis Introduction The monitor can be configured with either Glasgow 12-lead ECG analysis algorithm or Mindray 12-lead ECG analysis algorithm. The Glasgow algorithm is intended for adult, pediatric, and neonatal patients. The Mindray algorithm is intended for adult patients only.
We recommend using pediatric lead placement V4R, V1, V2, V4 - V6 if the patient is under 16 years of age. Please record V4R using the V3 electrode. Also set V3 Electrode to V4R. This is a normal practice for a patient of this age. 9 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
At the completion of 12-lead ECG interpretation, select Print to output the report via the printer. Exiting the ECG 12-Lead Screen To exit the ECG 12-Lead screen, select Exit on the ECG 12-Lead screen. BeneVision N1 Patient Monitor Operator’s Manual 9 - 3...
CAUTION • Only use parts and accessories specified in this manual. • Respiration monitoring is not for use on the patients who are very active, as this will cause false alarms. BeneVision N1 Patient Monitor Operator’s Manual 10 - 1...
Dry the skin completely before applying the electrodes. CAUTION • Proper skin preparation is necessary for good signal quality at the electrode site, as the skin is a poor conductor of electricity. 10 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Store the electrodes at room temperature. Open the electrode package immediately prior to use. • Check that the electrode packages are intact and not expired. Make sure the electrode gel is moist. BeneVision N1 Patient Monitor Operator’s Manual 10 - 3...
Setting the Resp Waveform Speed To set the Resp waveform speed, follow this procedure: Select the Resp numeric area or waveform area to enter the Resp menu. Select the Setup tab. Set Speed . 10 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
Once set, the detection level will not adapt automatically to different respiration depths. It is important to remember that if the depth of breathing changes, you may need to change the detection level. 10.6 Resp Troubleshooting For more information, see D Alarm Messages . BeneVision N1 Patient Monitor Operator’s Manual 10 - 5...
If this is a consideration, do not set the high alarm limit to 100%, which is equivalent to switching off the alarm. BeneVision N1 Patient Monitor Operator’s Manual 11 - 1...
If low perfusion persists, choose another method to measure oxygen saturation if possible. NOTE • PI is only available for Mindray SpO and Masimo SpO 11.5 Preparing for SpO Monitoring To prepare to monitor SpO , follow this procedure: Select an appropriate sensor according to the module type, patient category and weight.
Saturation levels may fluctuate rather than remaining steady for a period of several seconds. Often, the patient may fluctuate above and below an alarm limit, re-entering the non-alarm range several times. During such 11 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
The reliability of FastSAT is dependent on the setting for the averaging time and the input signal. FastSAT is disabled by default. To enable FastSAT, follow this procedure: Select the SpO numeric area or waveform area to enter the SpO2 menu. Select the Setup tab. Switch on FastSAT . BeneVision N1 Patient Monitor Operator’s Manual 11 - 5...
Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Select the Alarm tab → select the Other tab. 11 - 6 BeneVision N1 Patient Monitor Operator’s Manual...
Auto , the system will automatically select the first option as the PR source. If the current PR source is unavailable, the system will automatically switch PR Source to Auto . When you select IBP , the system will automatically select the first pressure label as the PR source. BeneVision N1 Patient Monitor Operator’s Manual 11 - 7...
Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized replacement parts which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device. 11 - 8 BeneVision N1 Patient Monitor Operator’s Manual...
Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device. BeneVision N1 Patient Monitor Operator’s Manual 11 - 9...
Plug the probe or temperature cable to the temperature connector. If you are using a disposable probe, connect the probe to the temperature cable. Follow the probe manufacturer’s instructions to connect the probe to the patient. BeneVision N1 Patient Monitor Operator’s Manual 12 - 1...
2. Try using a known good probe in case the sensor is damaged. tympanic thermometer display is frozen. Install or remove the probe cover to activate the thermometer. 12 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Check more frequently when making automatic or STAT measurements. Auto NIBP measurements with one and two minute intervals are not recommended for extended periods of time. • NIBP diagnostic significance must be decided by the physician. BeneVision N1 Patient Monitor Operator’s Manual 13 - 1...
The NIBP display shows only numerics. (10) (1) Systolic pressure alarm limits (2) NIBP unit: mmHg or kPa (3)The last NIBP measurement time (4) Time to the next measurement (for Auto mode and Sequence mode) 13 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Middle of the cuff should be at the level of the right atrium of the heart. If it is not, you must use the measurement correction formula to correct the measurement. For more information, see 13.8.9 Correcting the NIBP Measurements . BeneVision N1 Patient Monitor Operator’s Manual 13 - 3...
Set alarm properties as desired. 13.8.2 Setting the Initial Cuff Inflation Pressure To set initial cuff inflation pressure, follow this procedure: Select the NIBP numeric area to enter the NIBP menu. 13 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
To display multiple sets of the most recent NIBP measurements, follow this procedure: Access Tile Layout by either of the following ways: ◆ Select the Screen Setup quick key → select the Tile Layout tab. BeneVision N1 Patient Monitor Operator’s Manual 13 - 5...
Mindray-qualified service personnel only. 13.10.2 NIBP Accuracy Test The NIBP accuracy test should be performed once every two years or when you doubt the NIBP measurements. The NIBP accuracy test should be performed by Mindray-qualified service personnel only. 13.11 NIBP Troubleshooting For more information, see D Alarm Messages .
All invasive procedures involve risks to the patient. Use aseptic technique. Follow catheter manufacturer's instructions. • Mechanical shock to the invasive blood pressure transducer may cause severe shifts in zero balance and calibration, and cause erroneous readings. BeneVision N1 Patient Monitor Operator’s Manual 14 - 1...
If measuring intracranial pressure (ICP) with a sitting patient, level the transducer with the top of the patient’s ear. Incorrect leveling may give incorrect values(not applicable if measuring ICP with the Codman ICP transducer). 14 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Codman ICP transducer. For more information, see 14.4.1 Zeroing the Codman ICP transducer . If the target monitor does not support the Codman ICP transducer, do not use it for ICP monitoring. Follow this procedure to transfer the patient: BeneVision N1 Patient Monitor Operator’s Manual 14 - 3...
If monitors of different brands are used to zero the Codman ICP transducer, the zero reference values can be different. Use a Mindray monitor to Zero the Codman ICP transducer if you will take ICP measurement using a Mindray monitor. Otherwise the ICP measurement can be inaccurate.
Select the Setup tab. Set the following properties of the IBP waveform: ◆ Speed ◆ Scale : if Auto is selected, the size of the pressure’s waveform will be adjusted automatically. BeneVision N1 Patient Monitor Operator’s Manual 14 - 5...
Repeat step 2 in another waveform area if needed. Select to save the setting and exit the window. The main screen will display the overlapped IBP waves. 14 - 6 BeneVision N1 Patient Monitor Operator’s Manual...
The most accurate PAWP values are obtained at the end of the respiration cycle when the intrathoracic pressure is fairly constant and the artifact caused by respiration is minimal. WARNING • PAWP monitoring is not intended for neonatal patients. BeneVision N1 Patient Monitor Operator’s Manual 14 - 7...
Select the PA numeric area or waveform area to enter the PA menu, and then select PAWP . Wedge the flotation catheter into the pulmonary artery by observing the PA waveform changes on the screen, referring to the following figure. 14 - 8 BeneVision N1 Patient Monitor Operator’s Manual...
■ Select Speed to set a sweep speed for the displayed waveforms on the PAWP screen. ■ Select Scale to set the size of the PA waveform on the PAWP screen. BeneVision N1 Patient Monitor Operator’s Manual 14 - 9...
2. Perform zero calibration again. Do not sway the IBP transducer and tubing during zero calibration. For more information, see 14.3.3 Zeroing the IBP transducer . 3. If zero calibration still fails, replace the transducer. 14 - 10 BeneVision N1 Patient Monitor Operator’s Manual...
2.7.1Installing the N1 or External Parameter Module into the Modular Rack . In sequence, the follows are sidestream CO module, microstream CO module, mainstream CO module and N1 monitor. BeneVision N1 Patient Monitor Operator’s Manual 15 - 1...
(I/E) ratio. Measurement accuracy of the microstream CO2 module may be affected by the breath rate.For more information, refer to A.14.8CO Specifications. 15 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Connect the one end of the gas sample line. ◆ If you’re using the sidestream CO module, connect the watertrap to the CO module, and connect the gas sample line to the watertrap. BeneVision N1 Patient Monitor Operator’s Manual 15 - 3...
Page 152
Y-piece. (1) Sample line (2) Connect to the ventilator (3) Airway adapter (4) Connect to the patient ◆ For non-intubated patients, place the nasal cannula onto the patient. 15 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
15.5.2 Preparing to Measure CO Using Microstream CO Module To prepare the CO module for measurement, follow this procedure: Connect one end of the sample line to the microstream CO module. BeneVision N1 Patient Monitor Operator’s Manual 15 - 5...
Using Mainstream CO Module 15.6.1 Preparing to Measure CO Using Mainstream CO Module To prepare the CO module for measurement, follow this procedure: Connect the airway adapter to the sensor head. 15 - 6 BeneVision N1 Patient Monitor Operator’s Manual...
If a blockage is detected, clear or replace the adapter. To zero the sensor, follow this procedure: Connect the sensor to the module. In the CO2 menu, select Setup tab. BeneVision N1 Patient Monitor Operator’s Manual 15 - 7...
Entering the Standby Mode You can set the CO module to one of the following modes according to the module status: Select Measure mode when you use the CO modue for monitoring. ■ 15 - 8 BeneVision N1 Patient Monitor Operator’s Manual...
Module Without O When you are using the sidestream CO module without the O monitoring function to monitor a neonatal patient, you can select flow rate. To do so, follow this procedure: BeneVision N1 Patient Monitor Operator’s Manual 15 - 9...
Select the CO numeric area or waveform area to enter the CO2 menu. Select the Setup tab. Set the following compensation according to the actual condition. ■ Balance Gas 15 - 10 BeneVision N1 Patient Monitor Operator’s Manual...
Wait until the startup finishes. Completely block the gas inlet on the sidestream CO2 module or on the N1. Then the alarm message “ CO2 Airway Occluded ” will appear on the screen. Block the gas inlet for another one minute. BeneVision N1 Patient Monitor Operator’s Manual 15 - 11...
Solution EtCO measurements too low 1. Ventilate the room if the environmental CO concentration is too high. 2. Check the sample line and connectors for leakage. 3. Check the patient status. 15 - 12 BeneVision N1 Patient Monitor Operator’s Manual...
CO sampling consumables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device and/or CO sampling consumable. BeneVision N1 Patient Monitor Operator’s Manual 15 - 13...
■ Current window time line: indicates the time length of the current window. Waveform area: displays trend curves. The color of trend curves is consistent with the color of parameter labels. BeneVision N1 Patient Monitor Operator’s Manual 16 - 1...
Events are displayed in a chronological order. The most recent event is displayed at the top. The number of asterisk symbols before and event matches alarm priorities as follows: 16 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
The graphic trends review page displays trend data in a graphic form. To enter the graphic trends review page, select the Main Menu quick key → from the Review column select Graphic Trends . BeneVision N1 Patient Monitor Operator’s Manual 16 - 3...
Alarms are saved as events and will be maintained if the equipment is powered down. The time of equipment power down is not recorded as an event and cannot be reviewed. • A total loss of power does not affect the events already stored. 16 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
Page 166
■ ■ I = Inoperative (for example, Lead Off ) M = Missed beat ■ 16.2.7.5 Printing Event Reports To print event reports, follow this procedure: Enter the events review page. BeneVision N1 Patient Monitor Operator’s Manual 16 - 5...
Select Overview to switch to the compressed waveform page. 16.2.8.4 Printing the Full Disclosure Waveform Report To print a compressed waveform report, follow this procedure: Enter the full disclosure review page. 16 - 6 BeneVision N1 Patient Monitor Operator’s Manual...
Event statistics area: displays the total number of ABD events and the numbers of each event within the Zoom period. 16.2.9.3 Changing the Resolution of Trend Curves To set the resolution of trend curves, follow this procedure: Enter the OxyCRG review page. Set Zoom . BeneVision N1 Patient Monitor Operator’s Manual 16 - 7...
(only available for the independent external display) The minitrends screen shows graphic minitrends of parameters. 16.3.1 Entering the Minitrends Screen Choose one of the following methods to enter the minitrends screen: 16 - 8 BeneVision N1 Patient Monitor Operator’s Manual...
Select the Main Menu quick key → from the System column select Maintenance → input the required password → select Set the Review tab. From the Tabs page, Event , or Arrh Mark deselect those you do not need to review. BeneVision N1 Patient Monitor Operator’s Manual 16 - 9...
MEWS and NEWS and NEWS2 are intended for adult patients only. The patient category of the Custom Score is defined by the Mindray Clinical Score Configuration Tool. For more information, see Mindray Clinical Scoring Config Tool Instruction for Use (P/N: 046-012986-00).
Swipe right across the touchscreen with a single finger ■ ■ Select the button Take NEWS2 as an example, the EWS screen is shown as follows. Your screen may be slightly different due to the configuration. 17 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Scale 1 : for patient without hypercapnic respiratory failure. ◆ or patients with a prescribed oxygen saturation requirement of 88–92% (for example, in Scale 2 : f patients with hypercapnic respiratory failure). BeneVision N1 Patient Monitor Operator’s Manual 17 - 3...
NIBP : the monitor automatically starts scoring at the completion of each NIBP measurement. ◆ Alarm : the monitor automatically starts scoring when an alarm occurs to the parameter for scoring. 17 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
Select the parameter area where you want to display the GCS score, and then from the popup list select GCS . The following figure shows the GCS parameter area. Your display may be configured to look slightly different. BeneVision N1 Patient Monitor Operator’s Manual 17 - 5...
From the Eye Opening area, Verbal Response area, and Motor Response area, respectively select an item that represents the patient’s status. Select OK to accept the total score. The following table lists the default score range and color of relevant consciousness level. 17 - 6 BeneVision N1 Patient Monitor Operator’s Manual...
The Surviving Sepsis Campaign recommends to incorporate goals of care into treatment and end-of- life care planning, utilizing palliative care principles where appropriate. • The SepsisSight goals of care are intended to be applied in a time defined by the SSC guidelines. BeneVision N1 Patient Monitor Operator’s Manual 17 - 7...
■ Rescue start time and end time The use of drugs, for example adenaline, amiodarone and other drugs ■ The administrated treatments, for example, CPR, defibrillation, and other treatments ■ BeneVision N1 Patient Monitor Operator’s Manual 17 - 9...
To exit the rescue mode, choose either of the following ways: ■ Select the Exit Rescue Mode quick key. Select the Main Menu quick key → from the Alarm column select Exit Rescue Mode . ■ 17 - 10 BeneVision N1 Patient Monitor Operator’s Manual...
You can change the patient category and weight. This will not change the patient category and weight stored in the patient demographic information. BeneVision N1 Patient Monitor Operator’s Manual 18 - 1...
Page 184
Select Oxycont Unit , Hb Unit , and Pressure Unit . Then corresponding parameter values will be ■ automatically converted and updated accordingly. Select Range to show the normal range of each parameter. ■ 18 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
For more details about the printer, see the document accompanying the printer. With product upgrades, the monitor may support more printers and no prior notice will be given. If you have any doubt about the printer you have purchased, contact Mindray. 19.2...
Access alarm related tabs such as the Alarm tab for a parameter in one of the following ways: ◆ Select the parameter or waveform area of the desired parameter → select the Alarm tab. 19 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
3 lines and 4 columns with 2.5 seconds of ECG data generated by auto for each column. measurement Simultaneous : Record simultaneous 12-lead ECG data. NOTE • When ECG Lead Set is set to 3-Lead, ECG report cannot be printed. BeneVision N1 Patient Monitor Operator’s Manual 19 - 3...
19.9 Viewing Printer Statuses You can view the status of the most recent ten printing tasks in the Print Queue window. To view the status of printing tasks, follow this procedure: 19 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
In these cases, you need to install paper and then re-send the print request. Restart the printer if necessary. Therefore, you’d better ensure that there is enough paper in the printer before sending a print request. BeneVision N1 Patient Monitor Operator’s Manual 19 - 5...
20.3 Installing the Battery The battery must only be installed by service personnel trained and authorized by Mindray. To install the battery, contact your service personnel. The battery is installed when the monitor leaves the factory. BeneVision N1 Patient Monitor Operator’s Manual...
Battery-related Alarms The capacity of the battery is limited. When the battery is low, the monitor presents the Low Battery alarm, the alarm lamp flashes, and the monitor produces an alarm sound. 20 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
The performance of a rechargeable battery deteriorates over time. You should check the battery performance every two months or if you doubt that the battery may fail. BeneVision N1 Patient Monitor Operator’s Manual 20 - 3...
• Do not open the battery, heat the battery above 60 °C, incinerate the battery, or short the battery terminals. They may ignite, explode, leak or heat up, causing personal injury. 20 - 4 BeneVision N1 Patient Monitor Operator’s Manual...
21.2 Care and Cleaning Safety Information WARNING • Use only Mindray approved cleaners and disinfectants and methods listed in this chapter to clean or disinfect your equipment. Warranty does not cover damage caused by unapproved substances or methods. • We make no claims regarding the efficacy of the listed chemicals or methods as a means for controlling infection.
21.6 Sterilization Sterilization is not recommended for this monitor, related products, accessories, or supplies unless otherwise indicated in the Instructions for Use that accompany the products, accessories or supplies. 21 - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Product discoloration ■ Metal part corrosion Brittle and breaking wires, connectors, and equipment housing ■ ■ Reduced cable and leadwire life Overall system performance degradation ■ ■ Equipment malfunction or failure BeneVision N1 Patient Monitor Operator’s Manual 21 - 3...
The service personnel must be properly qualified and thoroughly familiar with the operation of the equipment. CAUTION • If you discover a problem with any of the equipment, contact your service personnel or Mindray. • Use and store the equipment within the specified temperature, humidity, and altitude ranges. •...
You can check system software version, module hardware and software version, and firmware version. 22.5 Testing Methods and Procedures Except the following maintenance tasks, all other test and maintenance tasks should be performed by Mindray- qualified service personnel only. Regular check, including visual inspection and power-on test ■...
Dispose of the monitor and its accessories when its service life is reached. Follow local regulations regarding the disposal of such product. WARNING • For disposal of parts and accessories, where not otherwise specified, follow local regulations regarding disposal of hospital waste. BeneVision N1 Patient Monitor Operator’s Manual 22 - 3...
375 to 805.5 Mainstream CO Temperature (ºC) 0 to 40 -20 to 60 module Relative humidity 10 to 90 10 to 90 (noncondensing) (%) Barometric (mmHg) 427.5 to 805.5 400 to 805.5 A - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Temp, IBP, ECG and cable connected, and auto NIBP measurements at an interval of 15 minutes. Shutdown delay: at least 15 minutes after the low battery alarm first occurs. BeneVision N1 Patient Monitor Operator’s Manual A - 3...
External power LED 1 (green) Battery LED 1 (two color-coded: yellow and green) A.7.2 Dock Connection status LED 1 (green) External power supply LED 1 (green) A.7.3 Transport Dock Power-on LED 1 (green) A - 4 BeneVision N1 Patient Monitor Operator’s Manual...
Interface Specifications of the Modular Rack Multi-pin connector A.9.3 Interface Specifications of the Dock Network connector Equipotential grounding terminal AC power input connector VGA connector Host monitor connector USB connector Multi-pin connector BeneVision N1 Patient Monitor Operator’s Manual A - 5...
48 hours at maximum. The specific storage time depends on the waveforms stored and the number of stored waveforms. OxyCRG view 48 hours. Trend data is stored one dot per second and the waveform stored is a compressed waveform. A - 6 BeneVision N1 Patient Monitor Operator’s Manual...
Magnetic Field emission according to MIL STD 461F, Chapter RE101: Radiated emissions, magnetic field, 30Hz to 100KHz. Limit class: Army. Magnetic Field susceptibility: Radiated susceptibility, magnetic field, 50 and 60 Hz, 30 A/m. ■ BeneVision N1 Patient Monitor Operator’s Manual A - 7...
Two monitors are used to view other monitors. ■ Only one monitor can transmit history data. ■ The weakest strength of the AP signal where the monitor is located cannot be less than -65 dBm. A - 8 BeneVision N1 Patient Monitor Operator’s Manual...
Notch filter 50/60 Hz Monitor, surgical, and ST mode: notch filter turns on automatically Diagnostic mode and High Freq Cut-off: notch filter is turned on/off manually Differential input impedance ≥5 MΩ BeneVision N1 Patient Monitor Operator’s Manual A - 9...
Page 222
12 RR intervals and then averaging them. The HR value displayed on the monitor screen is updated no more than one second. A - 10 BeneVision N1 Patient Monitor Operator’s Manual...
Page 223
12-lead ECG Interpretation Sampling rate 1000 samples/s (waveform) 500 samples/s (algorithm) Amplitude quantisation 24 bits PVCs/min Measurement range Adult: 0 to 300 Pediatric, neonate: 0 to 350 Alarm limit Range Step BeneVision N1 Patient Monitor Operator’s Manual A - 11...
20 to 254 bpm Resolution 1 bpm Response time <30 s (normal perfusion, no disturbance, PR value sudden changes from 25 bpm, to 220 bpm) Accuracy ±3 bpm Refreshing rate ≤1 s A - 14 BeneVision N1 Patient Monitor Operator’s Manual...
Page 227
1 bpm Accuracy ±3 bpm or ±3%, whichever is greater PR from IBP Module Measurement range 25 to 350 bpm Resolution 1 bpm Accuracy ±1 bpm or ±1%, whichever is greater BeneVision N1 Patient Monitor Operator’s Manual A - 15...
(depending on the configuration) in a typical patient population. A.14.7 IBP Specifications Standard Meet the standard of IEC 60601-2-34 2011. Technique Direct invasive measurement Measurement range -50 to 300 mmHg Resolution 1 mmHg BeneVision N1 Patient Monitor Operator’s Manual A - 17...
High 1 to 99 mmHg High (low limit + 2%) to 100% 18% to (high limit - 2)% High (low limit + 2%) to 100% 18% to (high limit - 2)% A - 18 BeneVision N1 Patient Monitor Operator’s Manual...
Page 231
Using the CO module without O sensor: Adult, pediatric: 120 ml/min Neonatal: 70 ml/min, 90 ml/min Sample flowrate tolerance ±15% or ±15 ml/min, whichever is greater. Start-up time 20s (typical), 90 s (maximum) BeneVision N1 Patient Monitor Operator’s Manual A - 19...
Page 232
≤750 ms@120 ml/min awRR measurement range 0 to 150 rpm awRR measurement precision <60 rpm: ±1 rpm 60 to 150 rpm: ±2 rpm awRR resolution 1 rpm Data sample rate 50 Hz A - 20 BeneVision N1 Patient Monitor Operator’s Manual...
Page 233
0 to 70 rpm: ±1 rpm 71 to 120 rpm: ±2 rpm 121 to 150 rpm: ±3 rpm awRR resolution 1 rpm Mainstream CO Module Standard Meet the standard of ISO 80601-2-55 2011 BeneVision N1 Patient Monitor Operator’s Manual A - 21...
Page 234
Meet the requirement for measurement accuracy within 6 hours Resolution 1 mmHg Rise time <60 ms Data sample rate 100 Hz awRR measurement range 0 to 150 rpm awRR measurement accuracy ±1 rpm awRR resolution 1 rpm A - 22 BeneVision N1 Patient Monitor Operator’s Manual...
The device needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided below. • Other devices may affect this device even though they meet the requirements of CISPR. BeneVision N1 Patient Monitor Operator’s Manual B - 1...
Page 236
50 Hz / 60 Hz 50 Hz / 60 Hz frequency typical commercial or hospital environment. magnetic fields IEC 61000-4-8 Note: U is the AC mains voltage prior to application of the test level. B - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Page 237
Note 1 : At 80 MHz and 800 MHz, the higher frequency range applies. Note 2 : These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. BeneVision N1 Patient Monitor Operator’s Manual B - 3...
This device complies with Part 15 of the FCC Rules and RSS-210 of Industry Canada. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. B - 4 BeneVision N1 Patient Monitor Operator’s Manual...
Page 239
Any changes or modifications to this equipment not expressly approved by Mindray may cause harmful radio frequency interference and void your authority to operate this equipment. The maximum antenna gain permitted complies with the e.i.r.p. limits as stated in RSS-210.
Nonsus V-Tach CCU: On Other departments: Off Vent Rhythm CCU: On Other departments: Off Pause Irr. Rhythm Promt A-Fib Promt PVCs/min CCU: On Other departments: Off Pauses/min CCU: On Other departments: Off C - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Relative ST Dual (ST Alarm Mode set to High limit 0.1 mV Low limit -0.1 mV Priority Alarm Outputs ST Analysis ST Segment Auto Show Marker ST Point J+60 Auto Adjust BeneVision N1 Patient Monitor Operator’s Manual C - 3...
Art-S/Art-D/Art-M High The Art value is lower than the Art Extreme alarm low limit. Check the patient’s Extremely Low condition and check if the alarm limit settings are correct. D - 2 BeneVision N1 Patient Monitor Operator’s Manual...
Check the connections of the electrodes and leadwires. ECG Signal Invalid Patient skin impedance is too high. Check ECG electrode application. ECG Learning Prompt ECG learning is manually or automatically triggered. BeneVision N1 Patient Monitor Operator’s Manual D - 3...
Low Perfusion Prompt The SpO sensor is not properly placed or the patient’s perfusion index is too low. 1. Check the sensor and sensor position. 2. Reposition the sensor if necessary. D - 4 BeneVision N1 Patient Monitor Operator’s Manual...
The IBP sensor fails. Replace the sensor. XX No Sensor High, Med, or The IBP patient cable and/or corresponding IBP Low, sensor is not connected or detached. Check the configurable cable and sensor connection. BeneVision N1 Patient Monitor Operator’s Manual D - 5...
1. Make sure that the ambient pressure meets the specifications, and check for sources that affect the ambient pressure. 2. Replug the module. If the alarm persists, contact your service personnel. D - 6 BeneVision N1 Patient Monitor Operator’s Manual...
Battery Off High Restart the monitor. If the alarm persists, contact your service personnel. RT Clock Need Reset High Contact your service personnel. RT Clock Not Exist High Contact your service personnel. BeneVision N1 Patient Monitor Operator’s Manual D - 7...
Cause and solution alarm reset Storage Card Error High The storage card fails or files are damaged. Restart the monitor to format the storage card. If the alarm persists, contact your service personnel. D - 8 BeneVision N1 Patient Monitor Operator’s Manual...
Page 261
Select a desired area to display the parameter numerics and (XX refers to the waveforms. For more information, see parameter label) 3.11.2 Displaying Parameter Numerics and Waveforms . BeneVision N1 Patient Monitor Operator’s Manual D - 9...
No physical damage to the enclosure and accessories. No physical damage to meters, switches, connectors, etc. No residue of fluid spillage (e.g., water, coffee, chemicals, etc.). No loose or missing parts (e.g., knobs, dials, terminals, etc.). BeneVision N1 Patient Monitor Operator’s Manual E - 1...
300 μ A in Normal Condition 1000 μ A in Single Fault Condition ◆ For IEC60601-1, ◆ 500 μ A in Normal Condition ◆ 1000 μ A in Single Fault Condition E - 2 BeneVision N1 Patient Monitor Operator’s Manual...
The following outlet conditions apply when performing the Mains on Applied Part test. Normal Polarity; ■ ■ Reversed Polarity LIMITS ◆ For CF applied parts: 50 μA ◆ For BF applied parts: 5000 μA BeneVision N1 Patient Monitor Operator’s Manual E - 3...
100 μA in Normal Condition ◆ 500 μA in Single Fault Condition NOTE • Make sure the safety analyzer is authorized comply with requirement of IEC60601-1. • Follow the instructions of the analyzer manufacturer. E - 4 BeneVision N1 Patient Monitor Operator’s Manual...
Page 267
Sensor Measurement range(%) Accuracy 518B, 518C 70~100 1.8380 90~100 1.2487 80~90 1.6938 70~80 2.4207 512FLH 90~100 1.4192 80~90 1.5238 70~80 2.3300 The Linear Matching Graph of 520A SpO Sensor Matching results of SpO2 measured by the EUT with 520A SpO2 sensor and the reference equipment The Linear Matching Graph of 520I SpO Sensor Matching results of SpO2 measured by the EUT with 520I SpO2 sensor and the reference equipment...
Page 268
The Linear Matching Graph of 520P SpO Sensor Matching results of SpO2 measured by the EUT with 520P SpO2 sensor and the reference equipment The Linear Matching Graph of 520N SpO Sensor Matching results of SpO2 measured by the EUT with 520N SpO2 sensor and the reference equipment F - 3...
Page 269
The Linear Matching Graph of 512H SpO Sensor Matching results of SpO2 measured by the EUT with 512H SpO2 sensor and the reference equipment The Linear Matching Graph of 512E SpO Sensor Matching results of SpO2 measured by the EUT with 512E SpO2 sensor and the reference equipment F - 4...
Page 270
The Linear Matching Graph of 512F SpO Sensor Matching results of SpO2 measured by the EUT with 512F SpO2 sensor and the reference equipment The Linear Matching Graph of 512G SpO Sensor Matching results of SpO2 measured by the EUT with 512G SpO2 sensor and the reference equipment F - 5...
Page 271
The Linear Matching Graph of 518B/518C SpO Sensor Matching results of SpO2 measured by the EUT with 518B SpO2 sensor and the reference equipment The Linear Matching Graph of 512FLH SpO Sensor Matching results of SpO2 measured by the EUT with 512FLH SpO2 sensor and the reference equipment F - 6...
The Accuracy of Masimo SpO Sensors Table information for the plots below show A values measured with Masimo Set Oximetry Technology in a clinical study. Adtx/Pdtx/4000/4001 Inf/Neo/NeoPt/4002/4003/4004/4005 F - 7...
The Accuracy of Nellcor SpO Sensors Sensor Motion # of Data Points Acceptance Pass/Fail Criteria 70% - 100% MAX-A, MAX-P 1.54 ≤2.0 Pass MAX-I, MAX-N 1.41 ≤2.0 Pass DS-100A 2.16 ≤3.0 Pass D-YS, OXI-P/I, OXI-A/N 1.96 ≤3.5 Pass Modified Bland-Altman for SpO - MAX-A, MAX-P Sensors (No Motion): SaO vs.
Page 275
Modified Bland-Altman for SpO - MAX-I, MAX-N Sensors (No Motion): SaO vs. (SpO 1. SpO - SaO 2. SaO 3. Upper 95% LoA 4. Mean Bias 5. Lower 95% LoA 2 (%) Modified Bland-Altman for SpO2 - DS-100A Sensor (No Motion): SaO vs.
Page 276
Modified Bland-Altman for SpO2 - D-YS, OXI-A/N, OXI-P/I Sensors (No Motion): SaO (SpO - SaO 1. SpO - SaO 2. SaO 3. Upper 95% LoA 4. Mean Bias 5. Lower 95% LoA 2 (%) F - 11...
International Special Committee on Radio Interference CMOS complementary metal oxide semiconductor central monitoring system C.O. cardiac output carbon dioxide COHb carboxyhemoglobin Compl compliance cardiac power index cardiac power output central venous pressure direct current BeneVision N1 Patient Monitor Operator’s Manual G - 3...
Page 280
ELWI extravascular lung water index Early Warning Score femoral arterial pressure Federal Communication Commission Food and Drug Administration FeCO Mixed Expired CO2 Concentration fraction of inspired FiAA inspired anesthetic agent G - 4 BeneVision N1 Patient Monitor Operator’s Manual...
Page 281
ITBI intrathoracic blood volume index ITBV intrathoracic blood volume left arm left atrial pressure liquid crystal display left cardiac work LCWI left cardiac work index light emitting diode BeneVision N1 Patient Monitor Operator’s Manual G - 5...
Page 282
PAWP pulmonary artery wedge pressure pCVP central venous pressure pediatric PEEP positive end expiratory pressure peak expiratory flow pre-ejection period peak inspiratory flow G - 6 BeneVision N1 Patient Monitor Operator’s Manual...
Page 283
Sepsis-Related Organ Failure Assessment arterial oxygen saturation from pulse oximetry signal quality index suppression ratio Surviving Sepsis Campaign signal strength index systolic time ratio stroke volume stroke volume index systemic vascular resistance BeneVision N1 Patient Monitor Operator’s Manual G - 7...
Page 284
VEPT volume of electrically participating tissue velocity index consumption for one breath oxygen consumption index G - 8 BeneVision N1 Patient Monitor Operator’s Manual...
Page 285
Caution: This device complies with Part 15 of the FCC rules and Industry Canada license‐exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. The manufacturer is not responsible for any radio or TV interference caused by unauthorized modifications ...
Need help?
Do you have a question about the BeneVision N1 and is the answer not in the manual?
Questions and answers