NOTE:
Changing Recirculating Solution (Hydrogen Peroxide)
HOT LINE
Switch off the HOTLINE® and dispose of the Disposable Set
■
in accordance with the usual procedures for your
establishment. Drain the Fluid from the HOTLINE® and
discard.
Refill the HOTLINE® with the second of the Isopropyl Alcohol
■
solution batches you made earlier.
If the HOTLINE
routine recirculating solution change, make sure you always use the
correct mix of Isopropyl Alcohol and Distilled water.
At the annual routine service interval the Hydrogen Peroxide
solution must be replaced as follows:
■
Prepare two batches of 1.4 litres of recirculating solution
according to formula 1 above for 0.3% Hydrogen Peroxide
maintenance solution.
■
Drain the HOTLINE®, and refill with one of the prepared
batches.
Install an L-70 or L-70 NI Disposable Administration Set in
■
the HOTLINE®'s Disposable Set connector.
■
Turn on the HOTLINE® and allow the fresh fluid to circulate
for 30 minutes.
■
Switch off the HOTLINE® and dispose of the Disposable Set
in accordance with the usual procedures for your
®
Blood and Fluid Warmer
S E C T I O N
5
®
requires topping up at any time before the next
Technical Service Pack
M a i n t e n a n c e
S o l u t i o n s
57